No CrossRef data available.
Published online by Cambridge University Press: 06 March 2019
Using a high purity germanium solid state detector with sufficiently high energy resolution to discriminate between copper Kα and Kβ radiation, the superlattice reflections of the ordered Cu-Zn β′ brass structure have be detected by the anomalous dispersion effect. By coupling the high purity germanium solid state detector to a multichannel analyser, the superlattice reflections of the β′ brass structure were also detected by energy dispersive x-ray diffraction. Other applications for this combination of high resolution detector, electronic energy discriminator and multichannel analyser are indicated.