Hostname: page-component-cd9895bd7-dzt6s Total loading time: 0 Render date: 2024-12-27T05:51:39.387Z Has data issue: false hasContentIssue false

Movement response to skin incision: analgesia vs. bispectral index and 95% spectral edge frequency

Published online by Cambridge University Press:  16 August 2006

H. Singh
Affiliation:
Department of Anaesthesiology, University of Hirosaki School of Medicine, 5-Zaifu-cho, Hirosaki, 036-8562, Japan
T. Sakai
Affiliation:
Department of Anaesthesiology, University of Hirosaki School of Medicine, 5-Zaifu-cho, Hirosaki, 036-8562, Japan
A. Matsuki
Affiliation:
Department of Anaesthesiology, University of Hirosaki School of Medicine, 5-Zaifu-cho, Hirosaki, 036-8562, Japan
Get access

Abstract

We studied the movement response to skin incision in 68 adult (males/females) ASA I–II patients receiving propofol±fentanyl intravenous anaesthesia using the bispectral index and 95% spectral edge frequency monitoring with an A-1050 EEG monitor. Following Ethics Committee approval, patients were randomly assigned to one of the following four treatments: Group P (n=17): propofol infusion, 1 mg kg −1 min−1 intravenous for 2 min, followed by propofol infusion, 200 μg kg−1 min−1, until skin incision; Group PF1 (n=17): fentanyl bolus, 1 μg kg−1 intravenous+ propofol infusion as in Group P; Group PF2 (n=17): fentanyl bolus, 2 μg kg−1 intravenous+ propofol infusion as in Group P; and Group PF3 (n=17): fentanyl bolus, 3 μg kg−1 intravenous+ propofol infusion as in Group P. The bispectral index and 95% spectral edge frequency were monitored continuously and recorded prior to induction of anaesthesia (base-line) and at skin incision. Twelve, 10, 4 and 4 patients responded to skin incision in Groups P, PF1, PF2 and PF3, respectively (P and PF1 vs. PF2 or PF3; P=0.0001, and 0.006). The bispectral index and 95% spectral edge frequency were significantly lower at skin incision compared with the base-line values in all the four treatment groups. However, only bispectral index values were significantly lower in the nonmovement as compared with the movement (M) category (32.6±8.9 vs. 37.4±10.3; P=0.04). Though deeper levels of hypnosis to lower bispectral index and 95% spectral edge frequency values may be effective in preventing the movement response to skin incision, provision of adequate analgesia rather than lower bispectral index and 95% spectral edge frequency (clinical maintenance) values may be more reliable for preventing the response to skin incision as bispectral index and 95% spectral edge frequency measure the hypnotic component of the anaesthetic effect. Lower bispectral index values may be more discriminatory as compared with 95% spectral edge frequency values for preventing the movement response to skin-incision.

Type
Original Article
Copyright
1999 European Society of Anaesthesiology

Access options

Get access to the full version of this content by using one of the access options below. (Log in options will check for institutional or personal access. Content may require purchase if you do not have access.)