Innovative Techniques for X-ray Calorimetry

Published online by Cambridge University Press:  12 April 2016

S. Labov
Affiliation:
Laboratory for Experimental Astrophysics, Lawrence Livermore National Laboratory, Livermore, CA, USA
E. Silver
Affiliation:
Laboratory for Experimental Astrophysics, Lawrence Livermore National Laboratory, Livermore, CA, USA
D. Landis
Affiliation:
Lawrence Berkeley Laboratory, Berkeley, CA, USA
N. Madden
Affiliation:
Lawrence Berkeley Laboratory, Berkeley, CA, USA
F. Goulding
Affiliation:
Lawrence Berkeley Laboratory, Berkeley, CA, USA
J. Beeman
Affiliation:
Lawrence Berkeley Laboratory, Berkeley, CA, USA
E. Haller
Affiliation:
Lawrence Berkeley Laboratory, Berkeley, CA, USA
J. Rutledge
Affiliation:
University of California, Irvine, CA, USA
G. Bernstein
Affiliation:
University of California, Berkeley, CA, USA
P. Timbie
Affiliation:
University of California, Berkeley, CA, USA

Abstract

Core share and HTML view are not available for this content. However, as you have access to this content, a full PDF is available via the ‘Save PDF’ action button.

In our x-ray calorimetry effort, we have developed several techniques which may be helpful to other groups working in this field. We are studying several different monolithic and composite calorimeter designs. In our readout configuration, the preamplifier circuit employs negative voltage feedback which allows us to accurately measure the temporal profile of the thermal pulse produced by an x-ray absorbed in a micro-calorimeter. Rise times of less than two microseconds have been observed in monolithic devices operating at .3 K. Furthermore, the feedback preamplifier can be configured for either positive or negative electro-thermal feedback. This preamplifier system is followed by an analog pulse shaping amplifier with a frequency response that can be adjusted to yield the maximum signal to noise ratio for a given thermal response of the calorimeter. In addition, we have developed several diagnostic procedures which have been useful in determining the operating and noise characteristics of our devices. These include an infrared light-emitting diode which flashes a discrete amount of energy on to the calorimeter, and a capacitively coupled test input to the preamplifier which allows us to directly determine the total noise in the thermal detection system. Finally, we are developing an adiabatic demagnetization refrigerator with a temperature control system that is designed to stabilize the 0.1 K cold stage to better than 8 μK. This is required for a resistive thermal detector with resolving power of 1000.

Type
9. Future X-ray Observatories, Detectors and Instrumentation
Copyright
Copyright © Cambridge University Press 1990

References

Coron, N., Artzner, G., Dambier, G., Jegoudez, G., Leblanc, J., Lepeltier, J. P., Deschamps, J. Y., Rocchia, R., Tarrius, A., Testard, O., Hansen, P. G., Jonson, B., Ravn, H. L., Stroke, H. H., and Turlot, E. 1985, Proc. of SPIE, 567, 389.Google Scholar
Fraser, G. W. 1987, Nucl. Inst. Meth., A256, 553.CrossRefGoogle Scholar
Goulding, F. S., Landis, D. A., and Pehl, R. H. 1969, in Semiconductor Nuclear-Particle Detectors and Circuits, ed. Brown, W., Higinbotham, W., Miller, G., and Chase, R. (Washington, D.C.: National Academy of Sciences), p. 455.Google Scholar
Goulding, F. S. and Landis, D. A. 1982, IEEE Trans. Nucl. Sci., NS-29, 1125.CrossRefGoogle Scholar
McCammon, D., Moseley, S. H., Mather, J. C., Mushotzky, R. F. 1984, J. Appl. Phys., 56, 1263.CrossRefGoogle Scholar
Moseley, S. H., Mather, J. C., and McCammon, D. 1984, J. Appl. Phys., 56, 1257.CrossRefGoogle Scholar
Moseley, S. H., Kelly, R. L., Schoelkopf, R. J., Szymkowiak, A. E., McCammon, D., and Zhang, J. 1988, IEEE Trans. Nucl. Sci., 35, 59.CrossRefGoogle Scholar
Silver, E., Labov, S., Landis, D., Madden, N., Goulding, F., Beeman, J., Haller, E., Rutledge, J., Bernstein, G., and Timbie, P. 1988, Proc. IAU 115, this volume.Google Scholar
Silver, E., Labov, S., Goulding, F., Madden, N., Landis, D., and Beeman, J. 1989, Nucl. Instr. Meth., submitted.Google Scholar
Zhang, J. and McCammon, D. 1988, Proc. IAU 115, this volume.Google Scholar