Improvements in SDD Efficiency – From X-ray Counts to Data

Published online by Cambridge University Press:  06 March 2017

Tara Nylese*
Affiliation:
EDAX Inc., Materials Analysis Division, AMETEK, 91 McKee Drive, Mahwah, NJ 07430
Jens Rafaelsen
Affiliation:
EDAX Inc., Materials Analysis Division, AMETEK, 91 McKee Drive, Mahwah, NJ 07430

Abstract

Type
X-ray Microanalysis
Copyright
Copyright © Microscopy Society of America 2017 

Introduction

The Si(Li) energy-dispersive spectroscopy (EDS) X-ray detector was the industry standard from the inception of its use on electron microscopes [Reference Fitzgerald1]. Since then silicon drift detectors (SDDs) have been introduced [Reference Strìder2], and these are currently the X-ray detector of choice across most fields and industries. The original advantage of SDDs was an increase in the speed of the sensor. This “speed gain” was a loosely defined benefit because it referred only to the X-ray charge collection at the chip and the increased rate of electron transfer to the collecting anode due to a voltage-biased drift field. In the earliest generations of SDDs these speed improvements were not fully realized because of limitations in electronic processing and other physical considerations. More recently there have been tremendous improvements in performance of leading-edge SDD technologies at all steps of the signal chain. This has led some companies to take a whole system approach that provides enhancements in collection efficiency as well as pulse processing speed. In general terms, the final number of useful X-rays analyzed by a given EDS system depends on the number of X-rays generated in the sample, the fraction of the generated X-rays that reaches the detector chip, the detector chip response, and the efficiency of the electronics processing the signal.

Materials and Methods

Input X-ray count rate

The input counts per second (ICPS) seen by a given EDS detector is highly dependent on the sample, electron accelerating voltage used, and the beam current at the sample. Keeping those parameters the same, the number that is usually considered to have the biggest impact on the ICPS is detector area. While detector area is an easy and intuitive number to understand, one has to be careful when trying to estimate the number of detected X-rays based on this number because the solid angle of the detector can vary significantly depending on which microscope or microscope port the detector is mounted on and how far the detector can be inserted. Recent reports provide solid-angle calculations for different detector geometries [Reference Zaluzec3] and a helpful web-based calculation tool [Reference Zaluzec4].

Window support grid area

A noteworthy parameter in the above-mentioned web-based calculation tool is the areal fraction covered by the window support grid. Most EDS detectors will have a thin window to protect the detector itself from the exterior environment. These windows will have a support grid to prevent the structure from collapsing. Figure 1 shows that the support grid can obscure a significant part of the active area of the EDS detector. Thus, one should consider both the open area of the support grid as well as the active area of the detector when comparing different detector types.

Figure 1 SEM images showing the support grid structure for two X-ray windows. (a) Polymer window. Image width = 2.5 mm. (b) Silicon nitride window. Image width= 1.3 mm. The hexagonal pattern of the silicon nitride window support structure covers 18% of the window area, whereas the venetian blind support structure for the polymer window covers 23% of the window area.

Window material

The composition of the window is another parameter to consider. Earlier systems used Be windows that had very limited transmission in the low-energy range. Current EDS detectors typically have a 300 nm thick polymer window or a <100 nm thick silicon nitride window. Assuming all other parameters are the same (detector-to-sample distance, detector area, support grid coverage), an analysis of a nitrogen-containing sample would result in roughly twice the number of counts in the nitrogen peak for a detector with a silicon nitride window compared to a polymer window. Figure 2 shows transmission curves for the two window materials and their effects on the nitrogen peak in spectra of CrN.

Figure 2 A selection of X-ray element maps acquired from a steel L profile with an average input count rate of 884 kcps at 0.12 µs amplifier processing time. (a) Phase map showing steel (yellow) and a Ni-rich inclusion (blue). Inside the Ni-rich inclusion, a Mo-rich phase is shown in bright purple and yellow, and a Si-rich phase is shown in cyan. The epoxy embedding material (green) is carbon-rich. (b) Ni map. (c) Mo map. (d) C map. Image width = 800 µm.

Output X-ray count rate

With an understanding of the geometric and material aspects affecting the input count rate to the SDD, we can now turn to the relationship of input to output count rate that leads to throughput. The X-ray count throughput is of primary importance to the microanalyst because this is the measure of counts that are actually used in the generation of spectra, linescans, and element maps. The series of processes that describes the ICPS to OCPS conversion is referred to as the “charge-to-voltage conversion.” This encompasses (1) the X-ray photon energy conversion to a number of electron-hole pair events in the SDD sensor and (2) the charge-to-voltage converter and pulse-shaping amplifier in the electronics [Reference Goldstein5]. While Si(Li) detectors commonly had the entire preamplifier external to the detector crystal itself, SDDs often integrate a JFET (junction gate field-effect transistor) on the detector chip to reduce capacitance and noise [Reference Lechner6]. Some recent systems employ an external CMOS (complementary metal-oxide semiconductor) preamplifier, which results in further reductions in capacitance and noise, allowing better performance at high count rates and better low-energy response [Reference Bombelli7]. Following the preamplifier is the pulse processor, which generates a voltage output pulse typically consisting of a triangular or trapezoidal curve with a flat at the top (Figure 3). The total pulse processing time is the amount of time it takes to analyze a single X-ray event and place it in the spectrum; consequently, this factor determines the number of X-rays that the pulse processor can analyze in a given amount of time at different pulse processor settings. Unfortunately, the terminology is not as rigorous as could be desired and often the terms amplification time, processing time, time constant, and peaking time are used interchangeably to either describe the rise/fall time or the entire pulse time. In Si(Li) detectors the charge cloud had to travel through the bulk of the Si crystal, while the SDD detector utilizes a drift field closer to the surface of the detector, resulting in faster charge collection and thus higher throughput. For SDDs the holding time is typically on the order of a few hundred nanoseconds while the rise/fall time varies from hundreds of ns to tens of μs, resulting in total pulse processing times ranging from roughly 0.5 μs to several tens ofμs. The lower limit for the total pulse processing time a given detector can run with, and consequently the upper limit for the throughput, is defined by the amount of time it takes for an electron cloud generated in the detector chip by an X-ray event to move to the detector anode for collection. In order to measure the generated charge correctly, the entire charge cloud has to be collected within the pulse time. If the pulse processing time is shorter than the charge collection time, only part of the energy will be collected within a pulse, resulting in a low-energy tail on the primary peak.

Figure 3 Illustration of pulse processor ramp signal and timings.

Processing time

An important factor related to the final OCPS is the detector busy time, when the detector system is doing the following: (a) analyzing an incoming pulse, (b) resetting the accumulated charge (reset times are typically less than 1 μs), and transferring data (on newer systems this is done continuously and the detector does not need to stop acquisition to transfer data). When the detector is busy, it should not accept a second X-ray pulse because the system can only process one X-ray event at a time. However, when two X-ray pulses fall within the same processing time window, the energies of the individual X-rays cannot be resolved and both pulses must be ignored. The time elapsed for measuring the discarded pulses is called “dead time.” As the ICPS increases, the probability of these near-coincident double pulses also increases and so does the dead time. For a given amp time, a low ICPS registers a low dead time, and a high ICPS yields a high dead time. Thus, the dead time can be used as a relative rate meter.

Throughput versus energy resolution

Many electronics processing architectures provide user selectable “Processing Time” or “Amp Time” settings: short times for high throughput or long times for improved energy resolution to resolve peaks with energies that are closely spaced. Thus, traditionally there was a known trade-off in deciding settings for speed and counting statistics versus energy resolution. The energy resolution of a peak (peak width) is dependent on two factors: the statistical distribution of the charge carrier conversion in the sensor and the thermal noise of the amplification process, primarily from the FET, the first stage of amplification [Reference Goldstein5]. At short amplifier times there would typically be more noise, leading to peak broadening and a low-energy cut-off as the noise became comparable to the charge contribution of low-energy X-rays. However, recent processing routines using the external CMOS “CUBE” preamplifier significantly reduces the noise at short pulse processing times compared to that of JFET devices [Reference Bombelli7]. Thus, the trade-off between count throughput versus energy resolution is no longer a factor for EDS systems using CUBE technology. Figure 4 shows representative Mn Kα energy resolution versus amp time for a selection of detectors, and the improvement at short times for the Octane Elite detectors is evident. Figure 5 shows the improvements in energy resolution and low-energy response for the Octane Elite detector at a short processing time.

Figure 4 Energy resolution (Mn Kα at FWHM) as a function of processing time for two generations and sizes of SDD detectors (Super being roughly twice the area of Plus) where the newer (Octane Elite) employs CUBE technology and shows significant improvements in resolution at fast processing times. The same digital pulse processor was used for all measurements.

Figure 5 Spectra from a carbon-coated SiO2 sample illustrating energy resolution and low-energy performance at fast processing times for two detectors. The spectra are normalized to the Si K peak. Although these detectors have comparable performance at the slow pulse processing times, the reduced overall noise of the CUBE pre-amplifier results in significantly better energy resolution and low-energy cut-off at fast pulse processing times. Note that C is not detectable at the 0.48 μs processing time in the older generation Octane detector. The increased O signal for the Octane Elite detector is attributable to the silicon nitride window.

Short amplifier processing time is the key to the SDD system’s ability to efficiently collect and analyze X-rays at high counts rates, rates that were previously impossible. The ultimate throughput of X-ray counts in a system is dependent on the time to process an X-ray pulse (Figure 6)—the shorter the amp processing time, the higher the throughput. Each processing time will have a maximum throughput (the top of each curve in the figure). These high-throughput count rates now permit high-speed X-ray mapping with input count rates over 750 kcps. This means the acquisition of a spectrum image that traditionally took hours can now be acquired in minutes.

Figure 6 Theoretical SDD output count rate as a function of input count rate for different amplifier processing times. Straight blue lines show that an input count rate of 884 kcps yields 585 kcps output count rate into the spectrum.

Instrumentation

The benefits and the improved performance of the new SDD systems can clearly be seen when the advances are applied to typical analytical examples. The data presented here were collected using a JEOL 7000F SEM with an EDAX Octane Elite Super detector with a silicon nitride window and a CUBE amplifier. To illustrate high-speed mapping, we used an epoxy-embedded, cross-sectioned and polished, cast-steel L profile with compositional variations at the bend of the L. To show quantification reproducibility and light-element sensitivity at high count rates, we present data from a Ti-6Al-4V alloy and a tool steel with a low C concentration.

Results

High-speed X-ray mapping

In the past handling high count rates typically meant sacrificing spectral resolution and light-element performance, resulting in data of limited use. However, with modern detectors capable of very high count rates, acceptable qualitative and quantitative data can be acquired in extremely short times. Figure 7 shows a group of X-ray maps acquired with average ICPS of 844 kcps going up to roughly 1M cps on parts of the sample. With a short 0.12 μs amp time, the average dead time was 37%. For this acquisition a pixel dwell time of 1 ms was used, which is roughly equivalent to a slow SEM image scan. Even phase maps [Reference Nylese and Anderhalt8] can be generated with these short dwell and processing times (Figure 7a), while still maintaining sensitivity to low-energy C X-rays. Obviously fast processing times are needed when collecting maps at high count rates, but most of the time, data acquisition is carried out at lower count rates and beam currents, often because of sample limitations or imaging needs. This is where improved detector efficiency can be helpful. For a fixed-input count rate, a better detector efficiency would yield a higher number of counts in the spectrum.

Figure 7 A selection of X-ray element maps acquired from a steel L profile with an average input count rate of 884 kcps at 0.12 µs amplifier processing time. (a) Phase map showing steel (yellow) and a Ni-rich inclusion (blue). Inside the Ni-rich inclusion, a Mo-rich phase is shown in bright purple and yellow phase, and a Si-rich phase is shown in cyan. The epoxy embedding material (green) is carbon-rich. (b) Fe map. (c) Mo map. (d) C map. Image width = 800 µm.

Ti-6Al-4V alloy

The precision and accuracy of quantitative analysis and the software’s ability to deal with potential spectral artifacts are concerns with going to higher count rates and faster processing times. To address these concerns, a titanium alloy known as Ti-6Al-4V, a common aerospace alloy known for its superior tensile strength and high-temperature toughness, was used to demonstrate quantitative analysis accuracy at high speeds. This alloy has a Ti Kβ–V Kα peak overlap (Figure 8), making quantitative analysis of the minor concentration of vanadium a challenge. The overlapped peak separation method in this analysis is based on an iterative Bayesian deconvolution. For more details see[Reference Eggert and Scholz9].

Figure 8 Overlay of normalized spectra collected from a Ti-6Al-4V alloy for 10 seconds at 20 kV and 20,000 cps incoming count rate with a processing times of 3.84 µs (red solid curve) and 0.24 μs (cyan outline) under identical microscope conditions. The Ti Kβ–V Kα peak overlap was separated with an iterative Bayesian deconvolution.

The analysis was performed at traditional collection rates: 20 kV, ICPS=20,000 cps, and 30 sec live time. Using an amplifier processing time of 3.84 µs resulted in a 124 eV energy resolution, and the standard-less quantification of the spectrum resulted in weight percent concentrations of 89.39% Ni, 6.17% Al, and 4.44% V (Table 1). We will consider these the baseline or “target” quantitative values for subsequent faster collections. Three additional measurements were taken with the same ICPS: one with 10 live seconds collection time at the same processing time and two at 0.24 µs processing time with 10 s and 1s collection times, respectively. The quantitative data remained good even for the difficult V overlap. This shows that the small loss in detector resolution at the short processing times did not affect the quantification reliability, and good quantification is possible even with limited statistics at short collection times. However, to reap the benefits of the faster processing times, one should increase the count rates significantly. Table 1 shows that even at 100,000 ICPS the quantification numbers show no significant changes. An ICPS of 100,000 cps was chosen because this is a value that can be obtained on most microscopes and samples with a reasonable beam current.

Table 1 Results of standard-less quantitative analysis of a Ti-6Al-4V alloy. The Ti Kβ –V Kα  peak overlap was separated with an iterative Bayesian deconvolution. Spectrum acquisition was with an Octane Elite Super detector at 20 kV.

Carbon in steel

A second application focuses on the accurate quantification of a known trace amount of carbon in a tool steel standard. Low-energy X-ray analysis of carbon is challenging because of X-ray absorption within the sample and in the various components of the detector. The low carbon concentration of 0.50 wt% creates a further challenge to detection and quantification. In this example, spectra were collected for 30 seconds at count rates ranging from 15,000 ICPS through to 200,000 ICPS, and the amplifier processing time was adjusted to keep the dead time below 30%. Figure 9a shows an overlay of the zoomed carbon peak region of the spectrum, comparing the two extreme collection rates. Figure 9b shows the stability of the results as the count rate was increased, with a C concentration ranging from 0.46 wt% to 0.52 wt% compared with the given known value of 0.50 wt%.

Figure 9 Quantitative results for trace carbon in a steel standard at 20 kV and 30% dead time. (a) Carbon peak acquired at 15,000 input cps with 7.86 µs processing time (red) and 200,000 input cps with 0.96 µs processing time (blue line). (b) Carbon concentration versus input count rate.

Summary and Conclusions

In order to really assess the efficiency of an EDS system, one needs to look beyond the simple numbers of detector area or collection solid angle and consider the entire processing chain. The input count rate under a given set of conditions is dependent on geometry, collection solid angle, window material, window support structure, and specific SDD construction. Once the X-rays are detected, they also need to be processed properly to be of any use. With new developments in amplifier circuits and pulse processors, the quality and speed of spectrum acquisition are constantly improving. With improved energy-resolution stability and light-element detection at fast processing times, higher throughput can be achieved at modest count rates by using a short pulse processing time with only a slight impact on spectral quality. Fast processing speeds can confidently be used for elemental mapping and for challenging applications with elemental overlaps.

Acknowledgements

The authors thank Russ Smith of EDAX for his continued focus on the importance of the quality of quantitative results even at high collection speeds. We also thank Kylie Simpson and Robert Rosenthal, our summer interns, who carried out the carbon measurements at high count rates.

References

[1] Fitzgerald, R et al., Science 159(3814) (1968) 528530.CrossRefGoogle Scholar
[2] Strìder, L et al., Mikrochim Acta (Suppl. 15) (1998) 1119.Google Scholar
[3] Zaluzec, NJ, Microsc Microanal (2014) 13181326.Google Scholar
[4] Zaluzec, NJ, XEDS Tools SolidAngle Calculator V-2014091607 (October 2016) http://tpm.amc.anl.gov/NJZTools/XEDSSolidAngle.html.Google Scholar
[5] Goldstein, JI et al., Scanning Electron Microscopy and X-Ray Microanalysis, Second edition. Plenum Press, New York, 1992.Google Scholar
[6] Lechner, P et al., Adv X Ray Anal 47 (2004) 5358.Google Scholar
[7] Bombelli, L et al., IEEE Nuclear Science Symposium Conference Record (2011) 1972–75.Google Scholar
[8] Nylese, T and Anderhalt, R, Microscopy Today 22(2) (2014) 1822.CrossRefGoogle Scholar
[9] Eggert, F and Scholz, W, Phys Status Solidi A 97 (1986) K9.Google Scholar
[10] Törmä, PT et al., IEEE T Nucl Sci 61 (2014) 695699.CrossRefGoogle Scholar
Figure 0

Figure 1 SEM images showing the support grid structure for two X-ray windows. (a) Polymer window. Image width = 2.5 mm. (b) Silicon nitride window. Image width= 1.3 mm. The hexagonal pattern of the silicon nitride window support structure covers 18% of the window area, whereas the venetian blind support structure for the polymer window covers 23% of the window area.

Figure 1

Figure 2 A selection of X-ray element maps acquired from a steel L profile with an average input count rate of 884 kcps at 0.12 µs amplifier processing time. (a) Phase map showing steel (yellow) and a Ni-rich inclusion (blue). Inside the Ni-rich inclusion, a Mo-rich phase is shown in bright purple and yellow, and a Si-rich phase is shown in cyan. The epoxy embedding material (green) is carbon-rich. (b) Ni map. (c) Mo map. (d) C map. Image width = 800 µm.

Figure 2

Figure 3 Illustration of pulse processor ramp signal and timings.

Figure 3

Figure 4 Energy resolution (Mn Kα at FWHM) as a function of processing time for two generations and sizes of SDD detectors (Super being roughly twice the area of Plus) where the newer (Octane Elite) employs CUBE technology and shows significant improvements in resolution at fast processing times. The same digital pulse processor was used for all measurements.

Figure 4

Figure 5 Spectra from a carbon-coated SiO2 sample illustrating energy resolution and low-energy performance at fast processing times for two detectors. The spectra are normalized to the Si K peak. Although these detectors have comparable performance at the slow pulse processing times, the reduced overall noise of the CUBE pre-amplifier results in significantly better energy resolution and low-energy cut-off at fast pulse processing times. Note that C is not detectable at the 0.48 μs processing time in the older generation Octane detector. The increased O signal for the Octane Elite detector is attributable to the silicon nitride window.

Figure 5

Figure 6 Theoretical SDD output count rate as a function of input count rate for different amplifier processing times. Straight blue lines show that an input count rate of 884 kcps yields 585 kcps output count rate into the spectrum.

Figure 6

Figure 7 A selection of X-ray element maps acquired from a steel L profile with an average input count rate of 884 kcps at 0.12 µs amplifier processing time. (a) Phase map showing steel (yellow) and a Ni-rich inclusion (blue). Inside the Ni-rich inclusion, a Mo-rich phase is shown in bright purple and yellow phase, and a Si-rich phase is shown in cyan. The epoxy embedding material (green) is carbon-rich. (b) Fe map. (c) Mo map. (d) C map. Image width = 800 µm.

Figure 7

Figure 8 Overlay of normalized spectra collected from a Ti-6Al-4V alloy for 10 seconds at 20 kV and 20,000 cps incoming count rate with a processing times of 3.84 µs (red solid curve) and 0.24 μs (cyan outline) under identical microscope conditions. The Ti Kβ–V Kα peak overlap was separated with an iterative Bayesian deconvolution.

Figure 8

Table 1 Results of standard-less quantitative analysis of a Ti-6Al-4V alloy. The Ti Kβ –V Kα  peak overlap was separated with an iterative Bayesian deconvolution. Spectrum acquisition was with an Octane Elite Super detector at 20 kV.

Figure 9

Figure 9 Quantitative results for trace carbon in a steel standard at 20 kV and 30% dead time. (a) Carbon peak acquired at 15,000 input cps with 7.86 µs processing time (red) and 200,000 input cps with 0.96 µs processing time (blue line). (b) Carbon concentration versus input count rate.