Hostname: page-component-cd9895bd7-7cvxr Total loading time: 0 Render date: 2024-12-27T07:36:23.968Z Has data issue: false hasContentIssue false

Increased User Safety from Glass-Free Liquid Nitrogen Dewars

Published online by Cambridge University Press:  21 January 2014

Steven Barlow*
Affiliation:
Electron Microscope Facility, San Diego State University, 5500 Campanile Drive, San Diego CA92182-4614

Abstract

Type
Microscopy 101
Copyright
Copyright © Microscopy Society of America 2014 

In a multi-user facility where students cycle through every few months as part of short-term projects and internships, maintaining safe operation of the equipment as well as ensuring the safety of users is of paramount concern. Thanks to films such as The Terminator, most users are aware of the freezing potential of liquid nitrogen (LN). However, users may not appreciate the explosive power associated with the rapid conversion of liquid nitrogen to gaseous nitrogen. When liquid nitrogen comes into contact with a surface warmer than its boiling point (−196 °C), a 700-fold expansion of liquid to gas occurs. This conversion can easily turn a broken glass vacuum dewar insert into lacerating projectile bits.

In response to two explosive incidents involving liquid nitrogen dewars in our biology department, I looked for an alternative that might minimize future dangerous events. We were fortunate that in both previous cases, which occurred in two unrelated laboratories, no one was injured. In the first, a break in the wall of a glass-lined dewar created a directed blast into a nearby paper towel dispenser, shattering the plastic cover. In the second case, an icicle of water, created by long-term use of the anti-contaminator on our transmission electron microscope (TEM), formed between the glass-lined dewar and the arm of the anti-contaminator assembly (Figure 1a). When a user gave the dewar a slight twist to break the icicle, the twist instead broke the LN-filled glass liner. When the LN hit the metal walls of the dewar, it vaporized. The rapid expansion of LN within the narrow bore of the dewar, coupled with the fragmentation of the glass into minute pieces, led to an upward explosion of glass fragments that the user described as a “stream of confetti shooting to the ceiling.” For weeks afterward, we found glass “confetti” in and around crevices in the TEM room. Discussions with our service engineers revealed that similar breaks in glass-lined dewars have occurred in labs around the world.

Figure 1a Standard 1-liter glass-lined dewar on a TEM.

As we looked for a replacement of our 1-liter glass-lined dewar (Figure 1a), we were also interested in finding a larger dewar that would allow for automated image capture over longer time periods than the 3–4 hours provided by our 1-liter dewar. Another local laboratory tried a 2-liter stainless steel wide-mouthed coffee dewar obtained from a local big-box store. The wide mouth was useful for insertion of the copper braid into the LN (Figure 1b). Since then we have used a similar dewar with great success. Because the commercial coffee dewar was not as tall as the 1-liter scientific dewar normally used, we initially used a surplus Styrofoam™ shipping container for gallon bottles of solvents (Figure 2a). The inner diameter of this container measured almost exactly the outer diameter of the stainless steel dewar and raised the dewar up the inch or so necessary for proper location on the TEM manufacturer's support stand. With usage over time, however, the Styrofoam container shed little Styrofoam particles around the room. We subsequently replaced the Styrofoam container with a surplus plastic pipette washing cylinder cut down in height (Figure 2b) and filled it with a thin foam spacer to hold the dewar in the cylinder.

Figure 1b Copper braid that transmits cooling from the dewar to the specimen chamber inside the microscope column.

Figure 2a Coffee carafe in a surplus Styrafoam™ container.

Figure 2b Coffee carafe in a surplus plastic pipette washing cylinder.

Although the major benefit to using the stainless steel coffee carafe dewar is the reduced chance of user injury, there are financial benefits as well. The 2-liter coffee carafe dewar costs approximately $50, compared to $160 for a 1-liter glass-lined dewar or $450 for a 1-liter scientific stainless steel dewar. Using the larger 2-liter dewar, our normal TEM operating vacuum improved with LN in the anti-contaminator dewar, and the time between dewar refills was significantly increased.

The interface between the coffee carafe and the Styrofoam lid on the instrument does frost up more than the standard 1-liter dewar; some care must be taken to catch the frost into the plastic or Styrofoam enclosure rather than have it fall onto the equipment during dewar refill or removal. As a result of catching this frost, water from the melting frost slowly accumulates in the space between the outer container wall and the outer wall of the carafe. This water should be poured off prior to refilling the dewar with liquid nitrogen.

In summary, a simple stainless steel coffee carafe can substitute for a scientific-grade glass-lined liquid nitrogen dewar. This low-cost coffee carafe provides increased lab safety and longer times between refillings.

Figure 0

Figure 1a Standard 1-liter glass-lined dewar on a TEM.

Figure 1

Figure 1b Copper braid that transmits cooling from the dewar to the specimen chamber inside the microscope column.

Figure 2

Figure 2a Coffee carafe in a surplus Styrafoam™ container.

Figure 3

Figure 2b Coffee carafe in a surplus plastic pipette washing cylinder.