Combined Tilt- and Focal-Series Tomography for HAADF-STEM

Published online by Cambridge University Press:  28 April 2016

Tim Dahmen*
Affiliation:
German Research Center for Artificial Intelligence (DFKI) GmbH, 66123 Saarbrücken, Germany
Holger Kohr
Affiliation:
KTH Royal Institute of Technology, Lindstedtsvägen 25, Stockholm, SE 100 44, Sweden
Andrew R. Lupini
Affiliation:
Oak Ridge National Laboratory, Materials Science and Technology Division, Oak Ridge, TN 37831-6071
Jean-Pierre Baudoin
Affiliation:
Aix-Marseille Université, Faculté de Médecine, 27 Bd Jean Moulin, 13005 Marseille, France
Christian Kübel
Affiliation:
Karlsruhe Institute of Technology (KIT), Institute of Nanotechnology (INT) and Karlsruhe Nano Micro Facility (KNMF), 76344 Eggenstein-Leopoldshafen, Germany
Patrick Trampert
Affiliation:
German Research Center for Artificial Intelligence (DFKI) GmbH, 66123 Saarbrücken, Germany
Philipp Slusallek
Affiliation:
German Research Center for Artificial Intelligence (DFKI) GmbH, 66123 Saarbrücken, Germany
Niels de Jonge
Affiliation:
INM – Leibniz Institute for New Materials, 66123 Saarbrücken, Germany

Abstract

Type
Electron Microscopy
Copyright
Copyright © Microscopy Society of America 2016 

Introduction

The primary means to study cellular ultrastructure in three dimensions (3D) with nanoscale resolution is currently tilt-series transmission electron microscopy (TEM) [Reference Hoenger and McIntosh1Reference Kourkoutis2]. In this imaging modality, images are acquired at different projection angles realized by mechanically tilting the sample stage. However, high-quality TEM imaging strongly depends on the condition that the sample is thinner than a few hundreds of nanometers because undesired inelastic and multiple electron scattering can otherwise lead to blurring of the image [Reference Lucić3]. For biological materials, sectioning is one way to prepare sufficiently thin samples. It has the disadvantage of limiting the tomogram to a subsection of the sample. For thicker samples containing, for example, whole cells, both inelastic and multiple elastic scattering contribute substantially to the projection data. Scanning transmission electron microscope (STEM), where the sample is scanned by a focused electron beam, is capable of recording images with nanoscale resolution of thicker samples [Reference Kübel4Reference Aoyama6]. However, because of geometric blurring and scattering of the probe through the sample thickness, resolution decreases with increasing sample thickness. Especially for slab-like samples, the effective thickness of the specimen as seen by the electron beam is increasing for higher tilt angles (Figure 1a). For example, a specimen titled to 70° exhibits a three-fold larger effective thickness than at 0° tilt. Especially for samples thicker than a few hundred nanometers, this limits the achievable resolution. It thus would be advantageous if the maximum tilt could be limited to ~40°, where the thickness increase is only a factor of 1.3. However, tomographic reconstructions then suffer from severe missing wedge artifacts, visible as elongated objects (axial elongation).

Figure 1 (a) The thickness of the sample as seen by the electron beam increases with 1/cos(β) as the sample is tilted by an angle β. (b) Schematic overview of the CTFS recording scheme. The sample is tilted by relatively large angular increments, and for every tilt position a focal series is recorded.

Avoiding tilting altogether is also possible by recording a stack of images with different focal values. Making use of the limited depth of field (DOF) of aberration-corrected STEM, a focal series can be used to obtain 3D STEM information [Reference Van Benthem7Reference Van den Broek13]. Therein, axial information is retrieved from a stack of images with different focal values. The 3D data set can be deconvolved to improve the axial resolution and to remove out-of-focus information [Reference Ramachandra and de Jonge14]. However, the focal series data suffers from a vertically elongated point spread function, and the vertical resolution is further reduced by shadowing effects below strongly scattering objects [Reference Behan10Reference de Jonge11].

In this article, we present a combined tilt- and focal-series (CTFS) scheme as a new recording method for 3D STEM imaging [Reference Dahmen15Reference Hovden16]. In this scheme, the specimen is rotated in relatively large tilt increments over a limited tilt range, and for every tilt position a focal series is recorded (Figure 1b). The CTFS method is an alternative to tilt-series STEM tomography, but with improved resolution in the axial direction for a given tilt range.

Materials and Methods

Image acquisition

To record a CTFS, the objective aperture semi-angle α is chosen as high as possible in order to achieve the smallest possible DOF, typically up to about 49 mrad (≈2.8°). The available tilt range is then covered in tilt increments of at most 2α, that is, the tilt increment is set to ≤5.6°. The reason for this condition is that the Fourier transform of a focal series is nonzero in a set that has the shape of an inverted double cone of the opening angle 2α [Reference Dahmen17]. As long as this condition is fulfilled, the volume corresponding to the tilt step is covered in reciprocal space.

For every tilt angle, a focal series is recorded. The distance from one focal plane to the next is selected such that an object perfectly in focus in one focal plane appears blurred by a disc filter of not more than one pixel radius on the next focal plane. This condition ensures that the interpolation error in the backprojection remains small and holds if the distance between consecutive focal planes is ≤1/sin(α)·pixelsize. In the case of α = 49 mrad, this condition corresponds to a vertical distance of ≈20 pixels.

Data Processing – Alignment

As it is with TEM tilt series, CTFS data need to be aligned to compensate for sample movements during image acquisition. In-plane alignment is initially sufficient between images recorded at the same tilt angle, that is, within a focal stack. If necessary, an additional alignment step can be added to align the images within each focal stack. However, the sample shifts do not only consist of in-plane movements like translation and rotation, but also of an axial shift. The exact focal positions of the images are, thus, unknown.

Lateral alignment errors can be corrected by averaging the images in a focal series and aligning the resulting conventional tilt series using a standard method [Reference Kremer18]. In order to compensate for the axial shifts, particles are detected in the tilt series. The particles are tracked through the tilt series, that is, the same particle is identified in several frames of the series [Reference Sorzano19]. The 3D position of each particle is then computed using triangulation from its position in the different images. In the next step, it is determined in which image of the series the particle is best in focus. The image with the highest intensity at the particle’s position is defined as the best focus for this particle and has to be searched in the focal stack. By combining both relations, a given image in a focal series can be assigned a focal depth, and the accuracy of this estimate is improved by averaging over many values gained by repeating this procedure for many particles.

Data Processing – Tomographic Reconstruction

The reconstruction of a 3D volume from CTFS data is based on an iterative Kaczmarz-type method [Reference Natterer20]. It is related to the well-known simultaneous iterative reconstruction scheme (SIRT), but the parallel projection and backprojection models are replaced with weighted integrations of the specimen function over double cones [Reference Dahmen17]. Thus, the focusing property of a real convergent electron probe is captured in the aberration-corrected STEM. The forward projection can be computed using a cone-tracing implementation based on stochastic sampling and ray tracing, a technique originally used for the simulation of optical camera systems with limited DOF in movie productions. The backprojection can be implemented based on a convolution with a kernel, which corresponds to a lateral cut through the double cone, that is, a disc. The method is described in detail in [Reference Dahmen17]. A software implementation is available as a plug-in for the software package Ettention [Reference Dahmen21].

Results

The method was evaluated on a whole-mount macrophage cell with gold nanoparticles coated with native low-density lipoprotein (LDL) taken up into vesicles [Reference Dahmen15Reference Baudoin22] (Figure 2). The cell had a total thickness of about 1 µm. The sample was tilted from −40° to +40° at 5° increments, resulting in 17 focal stacks. Each stack consisted of 20 images recorded at focal values 60 nm apart. The entire dataset therefore consisted of 340 images. Each image was recorded at 160,000 magnification with a pixel size of 2.3 nm and 5122 pixel resolution. The dataset was aligned as described in the methods section and reconstructed to a tomogram of 5123 voxels in size (Figure 3). The voxel size of the tomogram was 2.3 nm in each dimension.

Figure 2 Two images of the CTFS dataset taken at the same tilt angle (-40°) but at different focal values. The difference between the focus planes was 780 nm, which corresponds to 13 images. The orange arrows indicate sections of the images that are in focus. Image adapted from [Reference Dahmen15].

Figure 3 Results of the tomographic reconstruction. (a) A 3D rendering (perspective projection). (b) Sliced view in xy. (c) Sliced view in xz and yz directions. The largest nanoparticle is marked with an orange arrow in all views. All images are coded with false color for better visibility. The slice in yz direction was rotated by 90°.

For thick specimens, such as the whole cell investigated in this study, axial elongation as a result of the limited tilt range is the main factor limiting the interpretation of tomograms. In order to quantify axial elongation, we measured the ratio of axial and lateral resolution. Both were determined using the full width at half maximum (FWHM) criterion on a small number of (n = 17) spherical particles. The CTFS tomogram had an axial elongation factor eyz = 2.2 ± 0.5. Measured using the same method, a conventional tilt series recorded at the same tilt range and reconstructed using the SART method has an axial elongation factor of eyz = 2.8 ± 0.5 [Reference Hoenger and McIntosh1], so the CTFS leads to a significant (p < 7.0 × 10−4) reduction of the axial elongation factor.

In order to evaluate the improved axial resolution, it is of interest which frequencies are covered by a focal series, providing a look at the imaging with limited DOF in Fourier space. For the case of a parallel projection with unlimited DOF, the Fourier slice theorem states that the Fourier transform of a single projection contains those frequencies that lie on a plane perpendicular to the projection direction in reciprocal space. Comparably, the Fourier transform of the STEM transform is non-zero in a set with the shape of a complementary double cone of semi-angle π/2-α (Figure 4a) [Reference Hovden16Reference Dahmen17]. Considering the focal stacks from two adjacent tilt directions, one can see that the double cones overlap in a complex shape (Figure 4b). This means that certain spatial frequencies are contained in more than one projection, which might be the factor that stabilizes the reconstruction process and, thus, leads to improved resolution in the axial direction.

Figure 4 Fourier transform of a focal series. (a) The frequencies covered by one focal stack correspond to the shape of a double-wedge of opening semi-angle α. (b) If the tilt increment is chosen as Δβ = 2α, neighboring wedges overlap in a non-trivial shape (orange). (c) When considering a cross section through the origin and perpendicular to the tilt axis, the wedges seem to seamlessly cover the entire frequency space. (d) A cross section shifted along the tilt axis reveals a complex-shaped region in frequency space that contains information from more than one tilt direction. (e) A cross section further along the tilt axis toward highest frequencies exposes that the region containing information from both tilt directions expands toward higher frequencies. Image adapted from [Reference Dahmen17].

The reduced axial elongation of the CTFS compared to a tilt series is basically a result of additional data acquired in the recording scheme. The impact of the CTFS can further be understood by considering the Fourier transform of a reconstructed tomogram. In the case of the tilt series (Figure 5a), sharp streaks are present corresponding to the tilt directions. The missing wedge effect is also clearly visible, and the sequential algebraic reconstruction technique (SART) algorithm used to compute the tomogram can reconstruct almost no information in the missing wedge region. In the case of the CTFS method (Figure 5b), the streaks corresponding to the tilt directions are less pronounced and spread over an angular region equal to the beam-opening angle, so additional information is present between the tilt directions. The missing wedge is still present in the CTFS recording, but in the central vertical region low spatial frequency signal components are now present (orange arrow).

Figure 5 Comparison of tilt-series and CTFS reconstructions using SART. (a) Spatial frequency spectrum (Fourier transform) of an xz slice of the tilt-series reconstruction. The orange arrow indicates a single tilt increment highlighted in the image. (b) Spatial frequency spectrum of an xz slice of the CTFS reconstruction. The orange arrow indicates information reconstructed in part of the missing wedge. The white lines mark the border of the missing wedge. Image adapted from [Reference Dahmen15].

Discussion

Alternative approaches exist to reduce axial elongation for situations with a very limited tilt range using prior knowledge reconstructions. Making the assumption that the gradient of the reconstruction volume is sparse, one can formulate reconstruction algorithms that generate only a finite number of different gray values, for example discrete algebraic reconstruction technique (DART) [Reference Batenburg and Sijbers23] or compressed sensing approaches [Reference Leary24]. These approaches are independent of the CTFS recording scheme. We expect that new algorithms using a sparse gradient assumption for the reconstruction of CTFS data will be developed in the future.

Conclusion

For the imaging of thick samples of 0.5 µm or more, the HAADF-STEM mode is a powerful alternative to conventional TEM. Because aberration-corrected STEM images have a very limited DOF, tilt-series STEM tomography can be supplemented by depth sectioning. This approach helps to gather 3D information, particularly in situations where the accessible tilt range is very limited, as it is typically the case for very thick specimens. The recording scheme is called CTFS and consists of a tilt series, where for each tilt direction an entire focal series is recorded. Reconstructions from CTFS data were reconstructed using an iterative reconstruction algorithm that considered the convergent nature of the electron beam in the forward model. The CTFS led to a significant reduction of the axial elongation, a significant reduction of the missing wedge effect, and a better representation of the 3D shape of objects in the sample.

Acknowledgements

This research was partially funded by the European Research Project NOTOX (FP7-267038) and the DFG grant IMCL (AOBJ: 600875). The authors thank Eduard Arzt for his support through INM and the DFKI GmbH for additional funding. The authors acknowledge the electron microscopy supported by the Materials Sciences and Engineering Division, Basic Energy Sciences, Office of Science, U.S. Department of Energy (ARL), using instrumentation at the Oak Ridge National Laboratory’s Center for Nanophase Materials Sciences (CNMS), which is sponsored by the Scientific User Facilities Division, Office of Basic Energy Sciences, U.S. Department of Energy, as well as the electron microscopy support provided by the Karlsruhe Nano Micro Facility (KNMF), a Helmholtz large-scale user facility operated at the Karlsruhe Institute of Technology (KIT).

References

[1]Hoenger, A and McIntosh, JR, Curr Opin Cell Biol 21 (2009) 8996.CrossRefGoogle Scholar
[2]Kourkoutis, LFet al., Ann Rev Mater Res 42 (2012) 3358.CrossRefGoogle Scholar
[3]Lucić, Vet al., Annu Rev Biochem 74 (2005) 833865.CrossRefGoogle Scholar
[4]Kübel, Cet al., Microsc Microanal 11 (2005) 378400.Google Scholar
[5]Hohmann-Marriott, MFet al., Nat Meth 6 (2009) 729731.CrossRefGoogle Scholar
[6]Aoyama, Ket al., Ultramicroscopy 109 (2008) 7080.CrossRefGoogle Scholar
[7]Van Benthem, Ket al., Appl Phys Lett 87 (2005) 13.CrossRefGoogle Scholar
[8]Frigo, SPet al., Appl Phys Lett 81 (2002) 2112.Google Scholar
[9]Borisevich, AYet al., Proc Natl Acad Sci USA 103 (2006) 30443048.Google Scholar
[10]Behan, Get al., Phil Trans A, Math Phys Eng Sci 367 (2009) 38253844.Google Scholar
[11]de Jonge, Net al., Microsc Microanal 16 (2010) 5463.Google Scholar
[12]Dukes, MJet al., J Struct Biol 174 (2011) 552562.Google Scholar
[13]Van den Broek, Wet al., Ultramicroscopy 141 (2014) 2231.CrossRefGoogle Scholar
[14]Ramachandra, R and de Jonge, N, Microsc Microanal 18 (2012) 218228.CrossRefGoogle Scholar
[15]Dahmen, Tet al., Microsc Microanal S3 20 (2014) 113.Google Scholar
[16]Hovden, Ret al., Ultramicroscopy 140 (2014) 2631.Google Scholar
[17]Dahmen, Tet al., Microsc Microanal 21 (2015) 725738.Google Scholar
[18]Kremer, JRet al., J Struct Biol 116 (1996) 71–776.Google Scholar
[19]Sorzano, COSet al., BMC Bioinformatics 10 (2009) 124135.CrossRefGoogle Scholar
[20]Natterer, F, The Mathematics of Computerized Tomography, Society for Industrial and Applied Mathematics, Philadelphia, 2001.Google Scholar
[21]Dahmen, Tet al., at Microscopy & Microanalysis 2015, Portland, OR.Google Scholar
[22]Baudoin, J-Pet al., PLoS One 8 (2013) e55022-1e55022-8.Google Scholar
[23]Batenburg, K and Sijbers, J, Image Processing 2007, ICIP 2007, 4(2) (2007) IV 133.Google Scholar
[24]Leary, Ret al., Ultramicroscopy 131 (2013) 7091.Google Scholar
Figure 0

Figure 1 (a) The thickness of the sample as seen by the electron beam increases with 1/cos(β) as the sample is tilted by an angle β. (b) Schematic overview of the CTFS recording scheme. The sample is tilted by relatively large angular increments, and for every tilt position a focal series is recorded.

Figure 1

Figure 2 Two images of the CTFS dataset taken at the same tilt angle (-40°) but at different focal values. The difference between the focus planes was 780 nm, which corresponds to 13 images. The orange arrows indicate sections of the images that are in focus. Image adapted from [15].

Figure 2

Figure 3 Results of the tomographic reconstruction. (a) A 3D rendering (perspective projection). (b) Sliced view in xy. (c) Sliced view in xz and yz directions. The largest nanoparticle is marked with an orange arrow in all views. All images are coded with false color for better visibility. The slice in yz direction was rotated by 90°.

Figure 3

Figure 4 Fourier transform of a focal series. (a) The frequencies covered by one focal stack correspond to the shape of a double-wedge of opening semi-angle α. (b) If the tilt increment is chosen as Δβ = 2α, neighboring wedges overlap in a non-trivial shape (orange). (c) When considering a cross section through the origin and perpendicular to the tilt axis, the wedges seem to seamlessly cover the entire frequency space. (d) A cross section shifted along the tilt axis reveals a complex-shaped region in frequency space that contains information from more than one tilt direction. (e) A cross section further along the tilt axis toward highest frequencies exposes that the region containing information from both tilt directions expands toward higher frequencies. Image adapted from [17].

Figure 4

Figure 5 Comparison of tilt-series and CTFS reconstructions using SART. (a) Spatial frequency spectrum (Fourier transform) of an xz slice of the tilt-series reconstruction. The orange arrow indicates a single tilt increment highlighted in the image. (b) Spatial frequency spectrum of an xz slice of the CTFS reconstruction. The orange arrow indicates information reconstructed in part of the missing wedge. The white lines mark the border of the missing wedge. Image adapted from [15].