Material methods:VMAT plans of 35 patients, planned on the Eclipse Treatment Planning System (Aria 15), were included in the study. Plan acceptability was checked using statistical analysis, which includes homogeneity index, radical and median homogeneity index, coverage and uniformity index. Dose–volume histograms (DVH) of the plans were also studied to check prescribed dose (PD), Dmax, Dmin, D5 and D95. Portal dosimetry was also done by gamma analysis using 3%/3 mm criterion. SD and mean SD error were also calculated and analysed.
Results:Statistical analysis showed a mean HI of 1·054, coverage 0·959, UI 1·055, mDHI 0·962 and rDHI 0·866. SD of HI, coverage, UI, mDHI and rDHI was 0·019, 0·019, 0·014, 0·013 and 0·030, respectively. From the DVHs, mean of D5, D95, Dmin and Dmax was calculated at 6,252·9, 5,757·4, 6,413·3 and 5,657·7 cGy, respectively, with a prescribed dose of 6,000 cGy. According to gamma analysis, area gamma < 1 was 99·12% with a tolerance limit of 95%, maximum gamma was 1·466 with a tolerance limit of 3·5, average gamma was 0·388 with a tolerance limit of 0·5, area gamma > 1·2 was 0·242% with a tolerance limit of 0·5%, maximum dose difference was 0·6 with a tolerance limit of 1·0 and average dose difference was 0·029 with a tolerance limit of 0·2.