We use cookies to distinguish you from other users and to provide you with a better experience on our websites. Close this message to accept cookies or find out how to manage your cookie settings.
To save content items to your account,
please confirm that you agree to abide by our usage policies.
If this is the first time you use this feature, you will be asked to authorise Cambridge Core to connect with your account.
Find out more about saving content to .
To save content items to your Kindle, first ensure no-reply@cambridge.org
is added to your Approved Personal Document E-mail List under your Personal Document Settings
on the Manage Your Content and Devices page of your Amazon account. Then enter the ‘name’ part
of your Kindle email address below.
Find out more about saving to your Kindle.
Note you can select to save to either the @free.kindle.com or @kindle.com variations.
‘@free.kindle.com’ emails are free but can only be saved to your device when it is connected to wi-fi.
‘@kindle.com’ emails can be delivered even when you are not connected to wi-fi, but note that service fees apply.
This exploration is intended to measure tissue maximum ratios (TMRs) in smaller fields through CC01 detector and to compare CC01 measured TMRs with Pinnacle treatment planning software (TPS) calculated TMRs.
Materials and methods
CC01 compact chamber detector was used to measure TMR in water phantom for 6 and 18 MV beam delivered from Varian linear accelerator. Pinnacle TPS was employed in this study to calculate TMR from the measured percentage depth doses data. CC01 measured TMR data was compared with the calculated TMR data at depths from 5 to 20 cm for field sizes varying from 1 to 10 cm2.
Results
For the smallest given field size of 1 cm2, CCO1 measured 13·95% higher TMR value for 18 MV beam than that for 6 MV beam. At 20 cm depth for 1 cm2 field size, TMR due to 18 MV beam was 52·4% higher than the TMR due to 6 MV beam. For 6 MV beam, the maximum difference appeared between the measured TMR and pinnacle calculated TMR was 2·8% and for 18 MV beam, the maximum difference was 4%.
Conclusion
For both 6 and 18 MV beam, there was good agreement between CC01 measured and Pinnacle calculated TMRs for the field sizes ranging from 1 to 10 cm2. This exploration can be extended to the determination of other dosimetric parameters like TARs, TPRs in small fields.
Recommend this
Email your librarian or administrator to recommend adding this to your organisation's collection.