In the present paper, we estimated backscattering factors for the TLD and active personaldosimeters using 662 keV gamma-ray photons on a PMMA surface, which are, for the TLD,1.12; Dosicard, 1.116; DRD, 0.99; and Saphydose, 1.073, and on a water slab phantomsurface; for the TLD, 1.115; Dosicard, 1.112; DRD, 1; and Saphydose, 1.059. Thebackscattering factor contribution for the Saphydose is nearly half that of the TLD andDosicard, whereas it is zero for the DRD. The DRD underestimates Hp(10)by 10.3–11.6%, whereas for the Saphydose it is 3.4–5% compared with the TLD or Dosicard.The response of the DRD to Hp(10) is nearly independent of the phantom,whereas the TLD, Dosicard and Saphydose require phantoms for calibration for personalmonitoring. It is observed that the PMMA phantom contributes slightly higherbackscattering compared with the water slab phantom. The backscattering factor of theselected dosimeters is comparable with ICRU tissue already reported.