We use cookies to distinguish you from other users and to provide you with a better experience on our websites. Close this message to accept cookies or find out how to manage your cookie settings.
To save content items to your account,
please confirm that you agree to abide by our usage policies.
If this is the first time you use this feature, you will be asked to authorise Cambridge Core to connect with your account.
Find out more about saving content to .
To save content items to your Kindle, first ensure no-reply@cambridge.org
is added to your Approved Personal Document E-mail List under your Personal Document Settings
on the Manage Your Content and Devices page of your Amazon account. Then enter the ‘name’ part
of your Kindle email address below.
Find out more about saving to your Kindle.
Note you can select to save to either the @free.kindle.com or @kindle.com variations.
‘@free.kindle.com’ emails are free but can only be saved to your device when it is connected to wi-fi.
‘@kindle.com’ emails can be delivered even when you are not connected to wi-fi, but note that service fees apply.
Accuracy and reproducibility of the patient’s position is crucial for successful delivery of radiotherapy (RT). Data on palliative patients’ set-up uncertainties are sparse. The aim of this study was to calculate set-up errors observed for palliative patients positioned using one skin mark (Group 1) versus three skin marks (Group 2) and to assess the accuracy of both approaches.
Methods:
Displacements in the left–right (L–R) and superior–inferior (S–I) directions were retrospectively analysed for 175 sites treated with a course of fractionated palliative RT. Population mean, systematic and random errors were calculated in both directions for patients positioned with one and three skin marks. Frequency of deviations was also examined for both groups.
Results:
The population mean, systematic and random errors for Group 1 and 2 for the L–R direction were 0·0, 4·4, 4·8 and 0·4, 3·1 and 3·3 mm, respectively, and in the S–I direction: 0·1, 3·4, 4·2 and 1·2, 2·7 and 3·3 mm, respectively. Frequency of images within the clinical tolerance of 5 mm was 47·1% for Group 1 and 65·9% for Group 2.
Conclusion:
Three skin marks are recommended for patients receiving a fractionated course of palliative RT, as it reduces set-up error, reduces the number of gross displacements (>10 mm) and increases the number of displacements within the clinically acceptable tolerance of 5 mm.
Imaging protocols are implemented to identify and minimise set-up errors. A crucial component to the success of these protocols is staff compliance.
Materials and methods
This is case report describing a retrospective review of radiation therapists’ compliance to a palliative imaging protocol in a single large institution in one calendar year.
Results
The review showed a non-compliance to protocol for 8% of treatments. The most frequent protocol deviation was a failure to calculate the mean set-up displacement after 2/3 days of consecutive imaging.
Conclusion
Despite the presence of institutional evidence-based palliative imaging protocol unwanted deviations in practice can occur.