Missing LINQ: extrusion of a new-generation implantable loop recorder in a child

Published online by Cambridge University Press:  20 June 2016

Ahmad S. Chaouki*
Affiliation:
Department of Pediatrics, Division of Cardiology, The Heart Institute, Cincinnati Children’s Hospital Medical Center, University of Cincinnati College of Medicine, Cincinnati, Ohio, United States of America
Richard J. Czosek
Affiliation:
Department of Pediatrics, Division of Cardiology, The Heart Institute, Cincinnati Children’s Hospital Medical Center, University of Cincinnati College of Medicine, Cincinnati, Ohio, United States of America
David S. Spar
Affiliation:
Department of Pediatrics, Division of Cardiology, The Heart Institute, Cincinnati Children’s Hospital Medical Center, University of Cincinnati College of Medicine, Cincinnati, Ohio, United States of America
*
Correspondence to: A. S. Chaouki, MD, PhD, Cincinnati Children’s Hospital Medical Center, 3333 Burnet Ave, MLC 2003, Cincinnati, OH 45229, United States of America. Tel: 513 636 2921; Fax: 513 636 6952; E-mail: a-chaouki@northwestern.edu

Abstract

Cardiac rhythm monitoring has been facilitated by the use of implantable loop recorders. New models of these devices are 87% smaller than before allowing for easier implantation and use in the paediatric population. Recommendations are for closure with adhesive. We report a device extrusion in a 6-year-old patient. Based on this, our practice has changed to include subcutaneous sutures this complication.

Type
Brief Reports
Copyright
© Cambridge University Press 2016 

Access options

Get access to the full version of this content by using one of the access options below. (Log in options will check for institutional or personal access. Content may require purchase if you do not have access.)

References

1. Bloemers, BL, Sreeram, N. Implantable loop recorders in pediatric practice. J Electrocardiol 2002; 35 (Suppl): 131135.Google Scholar
2. Krahn, AD, Klein, GJ, Yee, R, Norris, C. Maturation of the sensed electrogram amplitude over time in a new subcutaneous implantable loop recorder. Pacing Clin Electrophysiol 1997; 20: 16861690.Google Scholar
3. Hutchinson, LJ, Stuart, G, Walsh, MA. Implantation of the new Medtronic LINQ loop recorder in an infant with ventricular tachycardia. Cardiol Young 2015; 25: 12211223.CrossRefGoogle Scholar
4. Rossano, J, Bloemers, B, Sreeram, N, Balaji, S, Shah, MJ. Efficacy of implantable loop recorders in establishing symptom-rhythm correlation in young patients with syncope and palpitations. Pediatrics 2003; 112: e228e233.Google Scholar
5. Seidl, K. Diagnostic assessment of recurrent unexplained syncope with a new subcutaneously implantable loop recorder. Europace 2000; 2: 256262.Google Scholar
6. Bayliss, CE, Beanlands, DS, Baird, RJ. The pacemaker-twiddler’s syndrome: a new complication of implantable transvenous pacemakers. Can Med Assoc J 1968; 99: 371373.Google Scholar
7. Abrams, S, Peart, I. Twiddler’s syndrome in children: an unusual cause of pacemaker failure. Br Heart J 1995; 73: 190192.Google Scholar
8. Tomson, TT, Passman, R. The Reveal LINQ insertable cardiac monitor. Expert Rev Med Devices 2015; 12: 718.Google Scholar