A dual excitation microfluorimeter is described for measuring
rapidly changing, intracellular calcium signals. A spinning
sector wheel is used in conjunction with a beam masking device
to provide rapid, efficient switching between the 2 excitation
wavelengths. Exposure intervals as short as 120 μs can be
achieved, yielding ratio samples at a rate of 6 kHz. Emission
photons are collected using a photomultiplier tube operating
in counting mode. When tested using FURA-2 as the calcium reporting
dye, throughput noise in the system is demonstrated to be due
to the statistical fluctuation inherent in photon counting.
An example of the operation of the system, using a guinea pig
cardiac myocyte, demonstrates that sufficient ratio data may
be acquires to fully characterize the fastest components of
the intracellular calcium signal.