Background:Device embolisation remains a major complication in transcatheter closure of atrial septal defect and patent ductus arteriosus. Although percutaneous retrieval is feasible in the majority of cases, surgical retrieval may be required in complicated circumstances. However, the methods of transcatheter device retrieval have not been completely established.
Methods:Bench tests of device retrieval were performed to verify the appropriate retrieval method according to device type/size. The devices used for testing were Amplatzer Septal Occluder (Abbott, Chicago, IL, United States of America), Figulla Flex II (Occlutech GmbH, Jena, Germany), Amplatzer Duct Occluder-I (Abbott), Amplatzer Duct Occluder-II (Abbott), and Amplatzer Vascular Plug-II (Abbott). The retrieval equipment constituted diagnostic catheters (multipurpose catheter and right Judkins catheter, 4-Fr or 5-Fr, Gadelius Medical, Tokyo, Japan), delivery sheath and cables for each device, Amplatz goose neck snares (Medtronic, Minneapolis, MN, United States of America), OSYPKA CATCHER (Osypka ag, Rheinfelden-Herten, Germany), and OSYPKA LASSOS (Osypka). We investigated the retrieval equipment and sheath sizes required for a successful retrieval procedure for variously sized devices.
Results:For patent ductus arteriosus devices, the type of snare and the snaring position are considered important. For atrial septal defect devices, simple snare capture or a double-snare technique with a sufficiently large sheath is effective. Special care should be taken when using the OSYPKA CATCHER for device retrieval.