We use cookies to distinguish you from other users and to provide you with a better experience on our websites. Close this message to accept cookies or find out how to manage your cookie settings.
To save content items to your account,
please confirm that you agree to abide by our usage policies.
If this is the first time you use this feature, you will be asked to authorise Cambridge Core to connect with your account.
Find out more about saving content to .
To save content items to your Kindle, first ensure no-reply@cambridge.org
is added to your Approved Personal Document E-mail List under your Personal Document Settings
on the Manage Your Content and Devices page of your Amazon account. Then enter the ‘name’ part
of your Kindle email address below.
Find out more about saving to your Kindle.
Note you can select to save to either the @free.kindle.com or @kindle.com variations.
‘@free.kindle.com’ emails are free but can only be saved to your device when it is connected to wi-fi.
‘@kindle.com’ emails can be delivered even when you are not connected to wi-fi, but note that service fees apply.
Non-invasive peripheral pulse oximeters are routinely used to measure oxyhaemoglobin saturation (SpO2) in cyanotic congenital heart disease. These probes are calibrated in healthy adult volunteers between arterial saturations of ~75 and 100%, using the gold standard of co-oximetry on arterial blood samples. There are little data to attest their accuracy in cyanotic congenital heart disease.
Aims
We aimed to assess the accuracy of a commonly used probe in children with cyanotic congenital heart disease.
Methods
Children with cyanotic congenital heart disease admitted to the Paediatric Intensive Care Unit with an arterial line in situ were included to our study. Prospective simultaneous recordings of SpO2, measured by the Masimo SET® LNCS Neo peripheral probe, and co-oximeter saturations (SaO2) measured by arterial blood gas analysis were recorded.
Results
A total of 527 paired measurements of SpO2 and SaO2 (using an ABL800 FLEX analyser) in 25 children were obtained. The mean bias of the pulse oximeter for all SaO2 readings was +4.7±13.8%. The wide standard deviation indicates poor precision. This mean bias increased to +7.0±13.7% at SaO2 recordings <75%. The accuracy root mean square of the recordings was 3.30% across all saturation levels, and this increased to 4.98% at SaO2 <75%.
Conclusions
The performance of the Masimo SET® LNCS Neo pulse oximeter is poor when arterial oxyhaemoglobin saturations are below 75%. It tends to overestimate saturations in children with cyanotic congenital heart disease. This may have serious implications for clinical decisions.