Hostname: page-component-cd9895bd7-gvvz8 Total loading time: 0 Render date: 2024-12-28T02:33:14.623Z Has data issue: false hasContentIssue false

A rare complication after radiofrequency catheter ablation in an adolescent case: skin burn

Published online by Cambridge University Press:  31 May 2023

Emine Gulsah Torun*
Affiliation:
Department of Pediatric Cardiology, Ankara City Hospital, Ankara, Turkey
Nevin Özdemiroğlu
Affiliation:
Department of Pediatric Cardiology, Ankara City Hospital, Ankara, Turkey
Serhat Koca
Affiliation:
Department of Pediatric Cardiology, Ankara City Hospital, Ankara, Turkey
*
Corresponding author: E. G. Torun; Email: drgtorun@gmail.com
Rights & Permissions [Opens in a new window]

Abstract

Radiofrequency catheter ablation is a preferred treatment method for cardiac arrhythmias in children due to its high success rate and low complication risk. We present an adolescent patient who underwent radiofrequency catheter ablation for Wolff–Parkinson–White syndrome and developed a skin burn at the site of the electrode patch. Skin burns can catastrophic consequences, especially in patients with life-threatening arrhythmias; therefore, clinicians should be aware of this complication.

Type
Brief Report
Copyright
© The Author(s), 2023. Published by Cambridge University Press

Radiofrequency catheter ablation is a preferred treatment method for cardiac arrhythmias in children due to its high success rate and low complication risk. Reference Hindricks1,Reference Rostock, Risius, Ventura, Klemm and Weiss2 Cardiac tamponade, pericarditis, pneumothorax, phrenic nerve palsy, thromboembolism, and atrioventricular block are among the major complications reported in patients undergoing radiofrequency catheter ablation. Reference Kugler, Danford, Houston and Felix3,Reference Ibrahim, Finta and Jubran4 Although there are studies on skin burns developing after radiofrequency ablation, which is used in the treatment of hepatic tumours, there are not sufficient data in the literature on skin burns that occur during cardiac arrhythmia treatment. Reference Mulier, Mulier, Ni, Miao and Dupas5

In this paper, we present an adolescent patient who underwent radiofrequency catheter ablation for Wolff–Parkinson–White syndrome and developed a skin burn at the site of the electrode patch.

Case report

A 14-year-old girl with Wolff–Parkinson–White syndrome was referred to our clinic for radiofrequency catheter ablation. The patient’s body mass index was 25.8. In the procedure performed under general anaesthesia, a high-risk left posterolateral manifest accessory pathway was detected in the patient. An electrode patch (9 x 10 mm) was placed on the anterior aspect of the trunk of the patient, and radiofrequency catheter ablation was successfully performed using a sheath placed on the right femoral artery, with a retrograde aortic approach, a 7-F multi-curve radiofrequency ablation catheter (MarinR, Medtronic, Minneapolis, USA), and a radiofrequency generator (Amper, St Jude Medical, St Paul, Minnesota, USA) (Fig. 1). Radiofrequency energy was applied three times for one minute each, at a maximum power of 50 W, and a temperature of 60°C. Impedance was kept constant during the application of energy (150–155 Ω). When the electrode patch was removed at the end of the procedure, A third-degree 5x6 cm skin burn was observed on the anterior surface of the patient’s body where the electrode patch had been placed (Fig. 2). The burn area healed with topical treatment.

Figure 1. a . Radiofrequency catheter ablation with the retrograde aortic approach. b . Accessory pathway ablation.

Figure 2. Third-degree skin burn at the electrode patch site.

Discussion

We presented a rare skin burn in an adolescent patient with high-risk Wolff–Parkinson–White syndrome after a successful radiofrequency catheter ablation procedure. This complication, which we encountered for the first time in the electrophysiology laboratory of our centre, has been rarely reported in paediatric patients in the literature and is important. Reference Ertuğrul, Karagöz and Aykan6 Although patients with life-threatening arrhythmias are successfully treated with radiofrequency catheter ablation, complications such as skin burns can result in catastrophic consequences.

Electrode patches provide an equal amount of current and heat in the electrodes and patches during radiofrequency ablation therapy, providing the intensification of the current patches passing through the patient to complete the electromagnetic circuit, acting as grounding. A large surface area and good skin contact are required to dissipate high energy charges. Reference Goldberg, Solbiati, Halpern and Gazelle7

In the study of Ibrahim et al, Reference Ibrahim, Finta and Jubran4 the incidence of skin burns after the radiofrequency catheter ablation procedure was found to be 0.28% over the age of 18 years. Ibrahim et al Reference Ibrahim, Finta and Jubran4 reported that the median body mass index of the patients with skin burns was also higher than the control group (36.6 versus 30.6, p = 0.044), their total procedure time was longer than the control group, and none of these patients had electrode patch malattachment. In the same study, the difference between the maximum current watt used and the maximum temperature reached did not statistically significantly different between the patients with skin burns and controls. Reference Ibrahim, Finta and Jubran4 In another paediatric case, the lack of contact between the pad and skin and the long duration of the procedure were considered as factors that could cause skin burns. Reference Ertuğrul, Karagöz and Aykan6 In our case, the patient’s body mass index was normal, and the total procedure time and the Watt and temperature used were not above the risk limits specified in the literature. Reference Ibrahim, Finta and Jubran4,Reference Goldberg, Solbiati, Halpern and Gazelle7 We did not observe any malattachments to lack of contact after the procedure, but the skin burn being mainly in the upper right corner of the patch electrode and presence of no other risk factors during the procedure suggest that the partial loosening of the skin-to-skin contact may have caused the skin burn.

Strategies such as the use of more and/or wider patches in radiofrequency ablation therapy, good contact between the electrode patches and skin, short-term use of low-energy currents, and monitoring of the temperature in the electrode patch area have been suggested to prevent skin burns. Reference Mulier, Mulier, Ni, Miao and Dupas5 Reference Steinke, Gananadha, King, Zhao and Morris8 In the study of Ibrahim et al, Reference Ibrahim, Finta and Jubran4 current Watt and maximum temperature used were not found to be risk factors for skin burn development, while body mass index and procedure time were found to be risk factors.

The incidence of skin burns at the electrode patch site after radiofrequency catheter ablation is low. Reference Ibrahim, Finta and Jubran4 Since the use of high-energy applications and multiple ablation procedures in the treatment of cardiac arrhythmia are expected to increase in future, it can be predicted that the incidence of such complications will also increase. Although there are not methods to prevent the development of skin burns associated with the cardiac radiofrequency procedures, this complication can be prevented by establishing good skin contact with the electrode patches and avoiding high-energy currents. Skin burns can be prevented using a new patch in each ablation procedure, checking the patch frequently during the procedure, and ensuring that the patch and the skin are dry. Further research or case reports can provide incidence and a better understanding of the causes of such complications and develop various methods to prevent them.

Acknowledgements

None.

Financial supports

This research received no specific grant from any funding agency in commercial or not-for-profit sectors.

Competing interests

None.

References

Hindricks, G. The multicentre european radiofrequency survey (MERFS): complications of radiofrequency catheter ablation of arrhythmias. the multicentre european radiofrequency survey (MERFS) investigators of the working group on arrhythmias of the European society of cardiology. Eur Heart J 1993; 14: 16441653.10.1093/eurheartj/14.12.1644CrossRefGoogle Scholar
Rostock, T, Risius, T, Ventura, R, Klemm, HU, Weiss, C, et al. Efficacy and safety of radiofrequency catheter ablation of atrioventricular nodal reentrant tachycardia in the elderly. J Cardiovasc Electrophysiol 2005; 16: 608610.10.1111/j.1540-8167.2005.40717.xCrossRefGoogle ScholarPubMed
Kugler, JD, Danford, DA, Houston, KA, Felix, G. Pediatric radiofrequency ablation registry of the pediatric radiofrequency ablation registry of the pediatric electrophysiology society pediatric radiofrequency catheter ablation registry success, fluoroscopy time, and complication rate for supraventricular tachycardia: comparison of early and recent eras. J Cardiovasc Electrophysiol 2002; 13: 336341.10.1046/j.1540-8167.2002.00336.xCrossRefGoogle ScholarPubMed
Ibrahim, H, Finta, B, Jubran, Rind J. Incidence and factors predicting skin burns at the site of the indifferent electrode during radiofrequency catheter ablation of cardiac arrhythmias. Cardiol Res Pract 2016; 2016: 5265682.10.1155/2016/5265682CrossRefGoogle Scholar
Mulier, S, Mulier, P, Ni, Y, Miao, Y, Dupas, B, et al. Complications of radiofrequency coagulation of liver tumors. Br J Surg 2002; 89: 12061222.10.1046/j.1365-2168.2002.02168.xCrossRefGoogle Scholar
Ertuğrul, İ., Karagöz, T, Aykan, H. A rare complication of radiofrequency ablation: skin burn. Cardiol Young 2015; 25: 13851386.10.1017/S1047951114002741CrossRefGoogle ScholarPubMed
Goldberg, SN, Solbiati, L, Halpern, EF, Gazelle, GS. Variables affecting proper system grounding for radiofrequency ablation in an animal model. J Vasc Interv Radiol 2000; 11: 10691075.10.1016/S1051-0443(07)61341-4CrossRefGoogle ScholarPubMed
Steinke, K, Gananadha, S, King, J, Zhao, J, Morris, DL. Dispersive Pad Site Burns With Modern Radiofrequency Ablation Equipment. Surg Laparosc Endosc Percutan Techn 2003; 13(6): 366371.10.1097/00129689-200312000-00003CrossRefGoogle Scholar
Figure 0

Figure 1. a. Radiofrequency catheter ablation with the retrograde aortic approach. b. Accessory pathway ablation.

Figure 1

Figure 2. Third-degree skin burn at the electrode patch site.