Published online by Cambridge University Press: 06 March 2019
The use of polarized x rays as the excitation source for x-ray fluorescence (XRF) measurements has been shown to significantly improve signal-to-background ratios. However, previous studies on polarized x rays applied to XRF techniques have concentrated on low-energy fluoresced x rays (<30 keV). In many cases strong matrix effects exist or the analyte is encased by a material that strongly attenuates low-energy x rays. These situations may preclude accurate assays based on L x-ray detection, and techniques based on the detection of higher energy K x rays may be more suitable because of the increased penetrability of higher energy x rays. The measurements and calculations reported in this work were made to assess the improvement in signal-to-background ratios and the increase in accuracy and detection sensitivity achievable by using polarized x rays as the excitation source for fluoresced x rays having energies between 25 and 110 keV.