Hostname: page-component-78c5997874-lj6df Total loading time: 0 Render date: 2024-11-10T11:57:00.984Z Has data issue: false hasContentIssue false

Colour Doppler imaging of the interspinous and epidural space

Published online by Cambridge University Press:  16 August 2006

T. Grau
Affiliation:
University of Heidelberg, Department of Anaesthesiology, Im Neuenheimer Feld 110, D-69120 Heidelberg, Germany
R. W. Leipold
Affiliation:
University of Heidelberg, Department of Anaesthesiology, Im Neuenheimer Feld 110, D-69120 Heidelberg, Germany
J. Horter
Affiliation:
University of Heidelberg, Department of Anaesthesiology, Im Neuenheimer Feld 110, D-69120 Heidelberg, Germany
E. Martin
Affiliation:
University of Heidelberg, Department of Anaesthesiology, Im Neuenheimer Feld 110, D-69120 Heidelberg, Germany
J. Motsch
Affiliation:
University of Heidelberg, Department of Anaesthesiology, Im Neuenheimer Feld 110, D-69120 Heidelberg, Germany
Get access

Abstract

Background and objective In recent studies, ultrasonic diagnostic imaging has proved useful in the screening of the trajectory of the epidural needle. With regard to possible side-effects of spinal and epidural anaesthesia caused by vessel injury, we aimed to evaluate the usability of Colour Doppler imaging for the depiction of interspinous vessels in prepuncture examination.

Methods Ultrasonic examination of the L¾ interspace area was performed in 20 volunteers. Using a 4-MHz and a 7-MHz probe with B-mode and Colour Doppler imaging, respectively, we compared four settings for the quality of vessel depiction in the puncture area. Overall resolution was evaluated according to the distinction of landmarks. Vascular structures were identified by pulsation (B-mode) or blood flow (Doppler).

Results Colour Doppler imaging of the L3/4 interspace was unachievable using the 7-MHz transducer. Vessel detection was possible in 50% of the B-mode images and in all of the 4-MHz Doppler images. Vessels were perceptible from a diameter of 0.5 mm. Veins were the predominantly visible structures. Overall vessel visibility was best using 4-MHz Colour Doppler.

Conclusions Prepuncture Doppler imaging can provide the epiduralist with information regarding the position of vessels in the needle trajectory. This might help to reduce complications in regional anaesthesia.

Type
Original Article
Copyright
2001 European Society of Anaesthesiology

Access options

Get access to the full version of this content by using one of the access options below. (Log in options will check for institutional or personal access. Content may require purchase if you do not have access.)