Safe and Cost-Effective Cleaning of Pressure-Monitoring Transducers
Published online by Cambridge University Press: 21 June 2016
Abstract
The CDC category I recommendation for high-level disinfection of pressure-monitoring transducers between uses imposes substantial costs. This practice has not been shown to be useful when disposable transducer domes are used. A prospective, randomized, double-blind study assessed the consequences of wiping transducers with alcohol between uses rather than sterilizing with ethylene oxide. We studied 5,197 transducer courses with disposable domes involving 2,202 patients in intensive care units during 15 months. There was no significant difference between the two treatment regimens in the risk of positive cultures of fluids from monitored lines, of cultures of cannula tips or of a positive blood culture. We conclude that under normal conditions, ie, in the absence of a cluster of transducer-related bacteremias, wiping transducers with alcohol between uses is sufficient when disposable domes are used.
- Type
- Original Articles
- Information
- Copyright
- Copyright © The Society for Healthcare Epidemiology of America 1988