Introduction
Reference BaderBader (1948, Reference Bader1950) has suggested that the liquid-water content of a snow sample can be determined in the field by adding a solution of known concentration and measuring the equilibrium temperature of the mixture. Reference ColbeckColbeck (1978) mentioned Bader’s method in his survey of various techniques of measuring the water content of snow and has given an estimate of its potential accuracy. This note describes an improved version of Bader’s method which has been used successfully in a wide variety of field conditions.
Theory
Suppose that a sample of mass b of wet snow contains mass a 0 of ice and mass x of water, then
A volume of a solution containing mass c0 of water and mass M of a solute at temperature t 1 is added to the sample in a Dewar of heat capacity H and initial temperature t 2. Phase equilibrium is established at a temperature t by melting a mass a m of ice. If a m < a 0, ❘t❘ is the freezing-point depression of the solution, which has been diluted by the addition of the mass am+x of water. The mass concentration A of a given solute, in grammes per 100 g of solution, which produces a given freezing-point depression can be found from standard chemical tables. Thus, if t is known, the mass of water c 1 in the solution at equilibrium can be calculated as
Conservation of mass of H2O gives
Conservation of energy gives
where L is the latent heat of fusion of ice (3.34 x 102 J g-1), δ the specific heat of the solution. and γ the specific heat of ice (2.09 J g-1 K-1). The volumetric water content of the snow is
where V is the volume of the sample and ρw the density of water. The relative error in the water content is approximately
where
If A ≪ 100,
and.
Reference BaderBader (1950) tested the freezing-point depression method in conditions such that:
-
(i) the solution and Dewar were initially at 0°C, i.e.
-
(ii) the solution was sufficiently dilute for a linear relation between t and A to be specified;
-
(iii) the depression of the freezing point was small enough for the heat content —Ht of the Dewar to be neglected.
However, in the author’s opinion these conditions are too restrictive for a general field method. Even though snow is available as a coolant, it is not always possible to keep the initial temperatures of Dewar and solution to 0°C in the field, especially if measurements are being made at short intervals. The freezing point depression has to be at least 2–3 K, so that it can be measured to a reasonable precision, and this will normally mean that the heat content of the Dewar cannot be neglected. The method may also be more accurate if a fairly concentrated solution is used.
Field tests
The method was tested in the field with an apparatus consisting of a wide-mouthed I 1 Dewar fitted with a polystyrene stopper 5 cm thick. Temperatures were measured using a “Digitron” platinum resistance probe and pre-prepared 100 ml samples of NaCl solution were carried to the field site in light plastic bottles. The snow samples were taken with a 500 cm3 sampling tube and weighed using a spring balance. The relative error in the weight of wet snow was ∣ Δb/b∣ ≈ 0.06, including instrumental error in the spring balance and random error in the volume of the snow sample. The relative error in the measured heat capacity of the Dewar and its stopper was ∣ΔH/H∣ ≈ 0.21. This error could certainly be reduced if the design of the lid were improved. The absolute error in the temperature measurements was Δt ═ Δt 1 ═ Δt 2 ═ ±0.1 K. This is probably the minimum value that can be achieved at the moment in the field with lightweight battery-operated equipment. Since the brine solution was made up into samples of known weight and concentration in the laboratory, the errors in c0 and H were very small; ∣Δc 0/c 0∣ ≈ ∣ΔH/H∣ ≈ 0.001.
Given these fixed errors, the most effective brine concentration could be estimated by calculating the relative error in the water content, ∣Δθ/θ∣, as a function of the mass of salt M added to the snow. Figure 1 shows this variation for t1 ═ t 2 ═ 0°C, b ═ 250 g, c 0 ═ 100 g, and H ═ 627 J K-1. The relative error decreases fairly rapidly for brine concentrations of up to about 5% by weight, but increasing the mass of salt beyond M ≈ 5 g is not very effective in reducing Δθ for water contents in the expected range of θ ≈ 5–10%.
It was found that using brine samples with M ═ 5 g, values of water content which were repeatable to within the estimated error Δθ ≈ ±0.02 could be obtained. This error is equivalent to about ±4% by weight and comparable to the 5% bands given by Reference WakahamaWakahama (1968) and Reference Wakahama, Wakahama, Nakamura and EndōWakahama and others (1968), who used a calorimeter to determine water content. Reference LeafLeaf (1966) reports an accuracy of Δθ ═ ±0.005 using a freezing calorimeter but as Reference ColbeckColbeck (1978) remarks this is unlikely to be a sensible estimate. In environments ranging from winter sub-arctic, with high winds and sub-zero temperatures, to summer high alpine, with intense solar radiation and warm air temperatures, the precision of the method could be maintained provided that
-
(i) the initial temperatures of the Dewar and brine were measured,
-
(ii) a relatively concentrated brine solution was used, and
-
(iii) the heat capacity of the Dewar and lid was included in the calculations.
The absolute accuracy of the method in the field could not be determined directly, since an alternative field apparatus for measuring water content was not available, but the measured values of 2–4% for the liquid water content of a ripe pack rising to 9–10% as the melt wave passed through agree with those reported by other authors. Measurements on frozen snow gave θ ═ 0.0±0.02.