Hostname: page-component-78c5997874-v9fdk Total loading time: 0 Render date: 2024-11-14T05:02:22.253Z Has data issue: false hasContentIssue false

A Hunsaker Mon-Jet tube trapped in the larynx

Published online by Cambridge University Press:  07 June 2011

I Khan*
Affiliation:
Department of Otolaryngology-Head and Neck Surgery, Aberdeen Royal Infirmary, Scotland, UK
M Shakeel
Affiliation:
Department of Otolaryngology-Head and Neck Surgery, Aberdeen Royal Infirmary, Scotland, UK
R Nagaraja
Affiliation:
Department of Anaesthesia, Aberdeen Royal Infirmary, Scotland, UK
B Ram
Affiliation:
Department of Otolaryngology-Head and Neck Surgery, Aberdeen Royal Infirmary, Scotland, UK
A D Thomas
Affiliation:
Department of Anaesthesia, Aberdeen Royal Infirmary, Scotland, UK
*
Address for correspondence: Mr Imran Khan, ENT Specialist Registrar, Ward 45, Aberdeen Royal Infirmary, Aberdeen AB25 2ZN, Scotland, UK Fax: +44 (0)1224 554569 E-mail: drkhanimran@gmail.com

Abstract

Objective:

We report a unique complication arising from the use of a Hunsaker Mon-Jet ventilation tube during microlaryngeal surgery, and we briefly review the literature on the use of this tube for jet ventilation.

Method:

A case report on the safe and successful management of a potential airway compromise, and a brief literature review on using the Hunsaker Mon-Jet ventilation tube during microlaryngeal surgery.

Case report:

A 46-year-old woman was scheduled to undergo removal of a vocal cord polyp under general anaesthesia. However, the polyp became stuck in the basket of the Hunsaker tube during intubation. The polyp and the Hunsaker tube were removed safely after a microlaryngeal tube was passed beyond the vocal cords. To our knowledge, this complication has not previously been reported.

Conclusion:

This case highlights the potential risk of laryngeal growths, especially vocal cord polyps, becoming trapped in the Hunsaker tube during intubation. Both the anaesthetist and the operating surgeon should be aware of this possible complication and, more importantly, how to deal with such a problem.

Type
Clinical Records
Copyright
Copyright © JLO (1984) Limited 2011

Access options

Get access to the full version of this content by using one of the access options below. (Log in options will check for institutional or personal access. Content may require purchase if you do not have access.)

References

1Barakate, M, Maver, E, Wotherspoon, G, Havas, T. Anaesthesia for microlaryngeal and laser laryngeal surgery: impact of subglottic jet ventilation. J Laryngol Otol 2010;124:641–5Google Scholar
2Davies, MJ, Hillel, DA, Maronian, NC, Posner, KL. The Hunsaker Mon-Jet tube with jet ventilation is effective for microlaryngeal surgery. Can J Anaesth 2009;56:284–90Google Scholar
3Orloff, LA, Parhizkar, N, Ortiz, E. The Hunsaker Mon-Jet ventilation tube for microlaryngeal surgery: optimal laryngeal exposure. Ear Nose Throat J 2002;81:390–4CrossRefGoogle ScholarPubMed
4Hunsaker, DH. Anesthesia for microlaryngeal surgery: the case for subglottic jet ventilation. Laryngoscope 1994;104:130Google Scholar
5Cook, TM, Alexander, R. Major complications during anaesthesia for elective laryngeal surgery in the UK: a national survey of the use of high-pressure source ventilation. Br J Anaesth 2008;101:266–72Google Scholar