No CrossRef data available.
Article contents
Efficacy of Gyrus diego® microdissector at varying oscillation speeds
Published online by Cambridge University Press: 13 March 2007
Abstract
To determine the efficacy of the Gyrus diego® microdissector at increasing oscillation speeds, using an in vitro tissue model.
It had not previously been established whether microdissectors were more efficient at higher or lower speeds.
We investigated the effect of varying microdissector oscillation speeds on the weight of material aspirated in a given time. A 4 mm straight blade was used with constant suction strength. Jelly and liver were used to simulate polyps and muscle plus connective tissue, respectively. Water was used as a control. Repeat readings were taken at speeds of 1000, 2000, 3000, 4000 and 5000 rpm. Data were analysed using linear bivariate regression.
The results showed significant linear trends in the cases of liver and jelly, with faster cutter speeds being associated with higher aspiration rates.
Conclusion. These results suggest that microdissector efficacy increases with speed, up to 5000 rpm.
- Type
- Main Article
- Information
- Copyright
- Copyright © JLO (1984) Limited 2007