Published online by Cambridge University Press: 13 October 2021
This study investigated the audiometric and sound localisation results in patients with conductive hearing loss after bilateral Bonebridge implantation.
Eight patients with congenital microtia and atresia supplied with bilateral Bonebridge devices were enrolled in this study. Hearing tests and sound localisation were tested under unaided, unilateral and bilateral aided conditions.
Mean functional gain was higher with a bilateral fitting than with a unilateral fitting, especially at 1.0–4.0 kHz (p < 0.05, both). The improvement in speech reception threshold in noise with a bilateral fitting was a 2.3 dB higher signal-to-noise ratio compared with unilateral fitting (p < 0.05). Bilateral fitting had better sound localisation than unilateral fitting (p <0.001). Four participants who attended follow up showed improved sound localisation ability after one year.
Patients demonstrated better hearing threshold, speech reception thresholds in noise and directional hearing with bilateral Bonebridge devices than with a unilateral Bonebridge device. Sound localisation ability with bilateral Bonebridge devices can be improved through long-term training.
Prof S Zhao takes responsibility for the integrity of the content of the paper