INTRODUCTION
Low dose-rate brachytherapy (LDR) is a popular radiation therapy modality used in the treatment of prostate adenocarcinoma. For low-grade disease, LDR offers several advantages compared to external beam radiotherapy including shorter course of treatment and lower normal tissue toxicity. Presently, treatment planning goals for LDR of the prostate are based on dose-volume histograms (DVH), isodose distributions and dose objectives to the target volume and organs at risk (OAR), as recommended in the American Association of Physicists in Medicine Task Group 137 report (TG-137).Reference Nath, Bice and Butler1
These quantifiers presume that the only quantity of importance is the delivered dose. However, the amount of biological damage attained with a certain dose is dependent on many factors, such as the dose rate, photon energy, tissue sensitivity, tissue seriality and sublethal damage repair during irradiation. These additional parameters vary greatly among the different radioisotopes used in prostate seed implants. Under some circumstances, plans with similar dose distributions have been shown to have different estimated radiobiological outcomes.Reference Mavroidis, Lind and Brahme2
We, hypothesise that, with the addition of radiobiological indices to the TG-137 criteria, a more clinically relevant, treatment planning analysis may be obtained. In this work, a radiobiologically-based treatment planning tool was produced to improve plan comparison through the addition of a small number of evaluation criteria. The aim of this work was to show that the recommended radiobiological indices are useful to differentiate patient treatment plans and that the calculated response estimates are realistic.
MATERIALS AND METHODS
This study is a retrospective analysis of ten patients receiving prostate seed implants as monotherapy. All the patients were treated with 125I seeds (BARD, BrachySource model, Covington, GA, USA) and prescribed 145 Gy for their clinical implant. For each patient, five additional treatment plans were created with different dose distributions. One of these additional plans used 103Pd seeds (Theragenics, model 200, Buford, GA, USA) and a prescription of 125 Gy. Each plan was created with a different goal in mind, given in Table 1, and therefore each had a unique dose distribution. However, while the isodose distributions for each plan were different, the plans had similar dosimetric characteristics. The goal of using these additional treatment plans was to simulate a realistic clinical situation where one must choose the best plan for a patient. Observations about calculated response accuracy were based on the follow-up data for the clinically used plan and responses reported in literature at similar dose levels.
Dose quantifiers and DVH distribution for these treatment plans are given in Table 3 and Figure 2. Pre-implant prostate volume studies were performed using transrectal ultrasound (TRUS). These ultrasound images were used for treatment planning on the Prowess Panther 3D Brachy Pro (Prowess, Concord, California, USA) system. Volumes for the prostate, urethra, rectum and bladder were drawn by a physician.
The physical dose was calculated with the in-house software using the TG-43 formalism, which used the seed strengths and coordinates from Prowess.Reference Nath, Anderson and Luxton3 The dose calculation was validated by comparing point doses and isodose distributions for a single source and a simple distribution of five seeds. All point doses agreed with Prowess within one percent. Then, the organ contours for each ultrasound image slice was exported from Prowess. Based on the physical dose and type of tissue present in each voxel, the tumour control probability (TCP) was calculated.Reference Mavroidis, Lind and Brahme2
The value of TCP is based on a calculation of biologically effective dose (BED), which is calculated for the tumour and normal tissues using equations 1a and 1b, respectively.Reference Yu, Anderson and Li19–Reference Kim and Tomé21
where,
In equations 1a and 1b, R 0 is the initial dose rate and λ is the decay constant (for 125I λ = 0.01166 day−1, for 103Pd λ = 0.04079 day−1). The sublethal damage repair constant (μ) was calculated using equation 2.
This factor accounts for the decrease in cell kill as the cell repairs damage. Here, a general repair half-life of 15 minutes was assumed for both tumour and normal tissues, making μ = 2.8 hour−1.Reference Wang, Guerrero and Li22,Reference Brenner and Hall23 The tumour repopulation factor (K) accounts for the growth of new tumour cells during treatment and is calculated from equation 3. A potential doubling time (Tpot) of 42 days was used in this analysis resulting in a repopulation factor of 0.11 Gy*day−1.Reference Wang, Guerrero and Li22,Reference Armpilia, Dale, Coles, Jones and Antipas24
The effective dose (Deff) was calculated using equation 4. The effective treatment time (Teff) was determined from equation 5. The endpoint for brachytherapy has been defined as the point where the rate of cell kill is equal to the tumour repopulation factor.Reference Antipas, Dale and Coles25 For normal tissues it is assumed that Teff = ∞, hence the effective dose is taken to be equal to the total physical dose accumulated over the lifetime of the seeds.
The relative biological effectiveness (RBE) for 125I and 103Pd has been reported to be between 1.4 – 1.5 and 1.6–1.9.Reference Antipas, Dale and Coles25–Reference Ling, Li and Anderson27 For this study values of RBE= 1.45 and RBE= 1.75 were used for 125I and 103Pd, respectively. The specific radiobiological parameters α/β, D 50 and γ used for each tissue are given in Table 2. D 50 is the dose which gives a 50% response and γ is the maximum normalised dose-response gradient. The D 50 and γ parameters are derived from clinical materials and describe the shape of the dose-response curve.Reference Källman, Agren and Brahme28 These parameters and the radiobiological model used are similar to those recommended in TG-137.Reference Nath, Bice and Butler29 Voxel response probability (P) was then determined using equation 6, where BED becomes BEDTum or BEDNT depending on whether the given voxel belongs to the tumour or an OAR.Reference Strigari, Orlandini and Andriani30–Reference Haworth, Ebert, Waterhouse, Joseph and Duchesne32 The overall response probability for the tumour and normal tissues is calculated using equations 7a and 7b, respectively.
where N is the total number of voxels in the organ, s is the tissue-specific relative seriality parameter and Δvi is the fractional subvolume of the organ irradiated. The overall probability of tumour control (PB), the overall probability of injury to the involved normal tissues (PI) and the complication-free tumour control probability (P+) for the treatment were calculated using equations 8, 9 and 10, respectively. The biologically effective uniform dose () which is the uniform dose that causes the same tumour control as the actual dose distribution for a given treatment, was calculated from equation 11.
The biologically effective uniform dose () calculates the uniform dose that would provide the same clinical outcome as the inhomogeneous dose distribution. It is a function of physical dose and tissue specific radiobiological parameters. The general expression of is derived numerically from the first part of the following equation, where for a tissue of uniform radiosensitivity, is given from the analytical formula of the second part of equation 12.
where denotes the 3-dimensional dose distribution delivered to the tissue and P() is the response probability of the tissue. The second part of the equation has been derived using the Poisson model.
RESULTS
The results from the dosimetric analysis for common dosimetric qualifiers are presented in Table 3.Reference Stock and Stone4–Reference Wallner, Merrick and True8 The results for the radiobiological analysis are presented for each organ of interest in Table 4. Comparing Tables 3 and 4, the PTV D90 were well above the prescription level, indicating adequate dose to the tumour. Although the PTV V150 was considerably different, the radiobiological response for all plans was very high. Considering the rectum, the average dose for all the plans 1–5 was similar, while the maximum dose varied considerably among plans. Plan 1 had a substantially larger rectal Dmax than any of the other plans and also yielded the highest complication probability. Plans 2–4 have similar dosimetry and complication ratios. Plan 6, which used 103Pd seeds, had rectal doses lower than the other plans and also a complication response that was much lower than plans 2–4. The urethra V100 was similar for each plan. The urethra V150 varied significantly, perhaps explaining the differences in urethral response among plans.
Table 4 shows the complication-free tumour control probability (P+) for each treatment. The number of seeds used for each plan, given in Table 3, is consistent with goals of each plan. Plan 4 uses fewer seeds due to the increased seed strength and plan 5 uses more seeds due to the decreased seed strength. Plan 5 has the highest P+ value, and in a clinical situation would be the preferred treatment plan. However, since PB is very high and PI is significant, it appears that simple refinement of the seed strengths may improve the treatment plans. Figure 1 shows the response probabilities as a function of initial seed strength for each treatment plan, as an average of all patients.
Figure 1 shows that for each plan the relationship between the response curves is different. The P+ curve indicates that for each plan the clinical initial seed strength is above optimal. This is seen to increase normal tissue complication probabilities (NTCP) while providing no additional tumour control. Using the initial seed strength that maximises P+, a simplistic plan optimisation was performed and compared to the original treatment plans. It is important to note that this simple optimisation is meant only to produce better plans for comparison in this work. A true treatment plan optimisation would include the spatial position of the seeds in addition to their strengths and is beyond the scope of this report. Figure 2 shows the DVHs of each plan, as an average of all patients. Tables 5 and 6 give dosimetric information and response probabilities for the optimised dose prescription.
Figure 2 and Table 5 show that all of the dose quantifiers are significantly reduced from the original treatment plans. All but plan 4 meet the criteria D90 ≥ 100% for a successful plan. Table 6 shows that TCP is virtually unchanged while the NTCP is dramatically reduced. The optimised plans have P+ nearly 40 % greater than the original plans.
DISCUSSION
Presently, the appropriateness of a treatment plan is determined by evaluating DVH distributions and dose quantifiers for the tumour and OAR. This can prove challenging when different indices give contradictory results. The aim of this project was to create a criterion whereby a diverse group of plans maybe compared in a simple and intuitive manner. Using a radiobiological model, the suitableness of a plan may be determined by evaluating a few response probabilities for the tumour and OAR. These probabilities may be further combined to yield a single criterion, P+. A limited patient cohort was evaluated in this study. However, these patients show the utility of the recommended evaluation tool. Evaluation of additional patients would likely further reinforce this result.
In Figure 1 the response probabilities for each plan was evaluated for a range of initial seed strengths. Using P+ to determine the ideal balance of TCP and NTCP, it was clear that each plan used seeds strengths that were higher than ideal; resulting in increased NTCP with no added TCP. Using the optimal seed strengths for each plan, new treatment plans were created and evaluated for each patient. The DVH distribution of the optimised plans is shown in Figure 2, with the dosimetric and radiobiological response data given in Tables 5 and 6, respectively. Figure 2 shows that for each plan the doses to the tumour and OAR are reduced, with plan 3 showing the most dramatic reduction in OAR doses, particularly the urethra.
Table 5 indicates that plan 4 has dose levels slightly below what is commonly considered acceptable. A possible explanation for the lower dose yet high TCP may be related to the smaller number of seeds that were used. The DVH drops more rapidly in the mid-dose region, lowering D90, while the high-dose tail extends further. In short, plan 4 is a more inhomogeneous dose distribution than the other plans. Table 6 shows that while the tumour dose was decreased, the TCP was nearly unchanged and NTCP was greatly reduced. By using the optimal initial seed strength for each plan P+ is improved by about 40%.
Prostate seed implants are unique among radiotherapy modalities in that the delivered dose distribution is largely dependent on the skill of the implanting physician and subsequent prostate edema. The predicted TCP for each plan is very high, 99.8%. Other investigators have reported lower relapse-free survival of 89–94% among similar patient populations.Reference Potters, Cao and Calugaru9,Reference Blasko, Grimm and Sylvester10 This discrepancy is likely due to the slight differences between the planned and the delivered dose distribution from edema and inexact seed placement. The effects of prostate edema would likely be small for iodine-125 implants since the isotope half-life is much longer than the edema half-life.Reference Waterman, Yue, Corn and Dicker11,Reference Crook, McLean, Yeung, Williams and Lockwood12 The effect of edema on palladium-103 implants, however, may be more significant.
Concerning the calculated rectal response, the optimised plans are significantly lower than the original plans. Authors have reported rectal complication rates of 8–18% for plans with similar dose distribution as the original plans.Reference Merrick, Butler and Dorsey13,Reference Gelblum and Potters14 Our results are consistent with these response probabilities. The optimised plans have reduced rectal doses, as is seen in Figure 2, and the response probability is correspondingly less.
For both the original and optimised plans the bladder response was negligible. This is consistent with our clinical experience that the bladder is located relatively far from the prostate and receives limited dose. Concerning the urethra, reported complication rates vary widely. This is likely due to the variability of urethral doses that may occur. Thomas et al. has reported a urethra grade 2 complication rate of 36% for plans with dosimetry similar to those used here in the original plans, which is consistent with our results.Reference Thomas, Keyes, Liu and Moravan15 Others have reported higher complication rates, 85–92%, for plans with dosimetry much higher the original plans presented here.Reference Merrick, Butler and Wallner16,Reference Takeda, Jingu and Koto17 However, these rates are consistent with those determined for plans with high initial seed strengths, seen in Figure 1. The urethral response for the optimised plans is significantly lower than those of the original plans. This is due to a similarly significant drop in urethral doses in the optimised plans.
The response probabilities determined for the original plans are consistent with patient follow-up data. None of the patients required additional treatment for their prostate cancer. Some patients reported minor rectal complications, such as temporary diarrhea or bloody stool. The most common complaints were of urinary symptoms such as dysuria and increased nocturia. Patients that had the most severe symptoms also had the highest calculated response probabilities.
The main advantage of this treatment planning tool is that it uses the response calculated for each voxel to predict the clinical outcome. This allows for simpler plan comparison based on a few criteria, such as response probabilities and P+.Reference Mavroidis, Katsilieri and Kelfala18 It also allows for optimisation of the seed strengths in order to better balance increased TCP with increased NTCP.
CONCLUSION
The goal of this work was to apply and produce a radiobiologically-based treatment planning evaluation tool to improve and simplify plan comparison through the use of a small number of evaluation criteria. By analysing several treatment plans for each patient we have shown that a radiobiologically-based treatment plan evaluation is more intuitive than dose quantifiers and offers opportunities to optimise treatment plans.