Thermo-luminescent dosimeter evaluation of extra-target dose in intensity modulated sequential tomotherapy for pancreatic cancer
Published online by Cambridge University Press: 27 September 2006
Abstract
The proximity of abdominal tumors to non-target organs has motivated exploration of intensity modulated radiation therapy (IMRT), with its demonstrated capability to create conformal dose gradients, in abdominal tumors. However, substantial extra-target dose may be observed to radiosensitive organs remote from the treatment field.
We present dosimetric data from two patients receiving abdominal IMRT using a sequential tomotherapy technique, demonstrating substantial dose to extra-target organs.
Keywords
- Type
- Case Study
- Information
- Copyright
- 2006 Cambridge University Press