Introduction
Delineating the proximal urethra for radiotherapy planning is commonly done for the treatment of prostate cancer. With the growing use of high-dose per fraction treatment paradigms such as stereotactic body radiation therapy and the dose escalation of magnetic resonance imaging (MRI)-detected prostatic lesions, this practice may become increasingly critical. Higher urethral dose with such treatment approaches is associated with increased toxicity including urinary frequency or dysuria. Reference Hsu, Hunt and Straube1–Reference Budaus, Bolla and Bossi3 Interest is growing in the use of urethra-sparing radiation therapy for prostate cancer; however, results from these approaches have been conflicting. Reference Zilli, Franzese and Bottero4–Reference Vainshtein, Abu-Isa and Olson6 In part this may be due to challenges in accurately contouring the proximal urethra.
Identifying the proximal urethra can be challenging on computerised tomography (CT) scan alone, and the geometric centre of the prostate has been used as a surrogate. Reference Bucci, Spadinger and Hilts7,Reference Waterman and Dicker8 However, the geometric centre has been shown to vary significantly from the Foley catheter-defined urethra. Reference Lee, D’Souza and Yamal9 Accurately contouring the proximal urethra often relies on Foley catheter insertion, retrograde urethrogram or using the geometric centre of the prostate. Drawbacks to these approaches include patient discomfort and possible distortion of normal anatomy from catheter placement. Reference Lee, D’Souza and Yamal9,Reference Dekura, Nishioka and Hashimoto10 We hypothesised that we would be able to visualise and contour the proximal urethra by using a MRI scan to delineate urine flow during voiding. We believed that during the act of voiding urine would fill the potential space in the proximal urethra and that this flowing urine would be observed during an MRI scan. By visualising the urine flow and proximal urethra non-invasively, we expected that this non-invasive approach would be preferable to invasive methods due to greater patient satisfaction and reduced risk of infection and trauma compared to current methods for visualising the prostatic urethra. Our aim was to determine if it was feasible to visualise the proximal urethra using a non-invasive novel technique.
We prospectively trialed a novel non-invasive technique to allow visualisation of the proximal urethra during simulation scanning using a rapid sequence MRI protocol to visualise the urinary flow of patients voiding during the MRI scan.
Materials and Methods
After institutional review board approval and informed consent, we enrolled patients with prostate cancer undergoing CT simulation scans and 0·35T MRI scans on the same day. We utilised our Viewray MRIdian Linear Accelerator, which consists of a 6 MV flattening-filter-free linear accelerator integrated between two 0·35 T split superconducting magnets, for the MRI-simulation scans. Seven patients were enrolled on study and asked to urinate into a condom catheter in the supine treatment position. All patients were instructed to have a comfortably full bladder for simulation scan. Patients verbally indicated when voiding commenced and a 45 × 45 × 24 cm field of view, with an in-plane resolution of 1·5 mm × 1·5 mm and slice thickness of 3 mm, 17-second true fast imaging with steady-state free precession (TRUFI) MRI scan was obtained. TRUFI is a balanced steady-state free precession sequence, yielding a T2/T1-weighted contrast. Reference Green, Rankine and Cai11,Reference Bieri and Scheffler12 Patients additionally underwent a CT simulation scan with or without a Foley catheter in place, per current institutional standard. The MRI and CT images were fused together, based on anatomic registration with a focus on the prostate. The proximal urethra was defined between bladder neck and the base of the pubic symphysis and contoured separately on the voiding MRI scan and CT simulation scan. The proximal urethra-segmented volume between MRI and CT contours was compared and the Dice coefficient was calculated as a measure of spatial agreement.
Results
We prospectively enrolled seven patients with prostate cancer. Of these, four were able to void in the supine position during the MRI scan. For these four patients, direct visualisation of urinary flow through the proximal urethra was readily achieved. A representative scan showing the clearly defined proximal urethra during voiding is shown in Figure 1.
Of the four patients who were able to void in the supine position, one underwent CT simulation with Foley catheter and four underwent CT simulation without catheter placement. For the four patients without Foley catheter, the proximal urethra on CT was contoured as the geometric centre of the prostate, as previously discussed. Reference Bucci, Spadinger and Hilts7,Reference Waterman and Dicker8 The average volume of the proximal urethra contoured on voiding MRI was significantly greater than the average proximal urethra volume contoured on CT (4·07 versus 1·60 cc, respectively; p = 0·02). The Dice coefficient averaged 0·28 (range 0–0·616). Representative images of patients with and without a Foley catheter, as compared with voiding MRI scan, are shown in Figures 1c and 1d. The voiding urethra tended to be non-central, larger and more curvilinear, as compared to the CT-defined urethra.
Discussion
In this small, proof-of-concept prospective clinical trial, the volume of the proximal urethra differed significantly when contoured on a voiding MRI compared to the volume determined on conventional CT simulation. This is the first study comparing the three-dimensional position of the voiding urethra to that of a CT scan-defined urethra.
The Foley catheter-defined urethra was defined as ‘ground truth’ in the Groupe Européen de Curiethérapie/European Society for Radiotherapy and Oncology recommendations. Reference Kovacs, Potter and Loch13 However, there is a growing body of evidence that this may not accurately encompass the urethra in external beam treatments where a Foley catheter is not used for each treatment. Dekura et al. found that the urethral position was significantly different when using a soft guidewire as compared to when a Foley catheter was in place. Reference Dekura, Nishioka and Hashimoto10 Additionally, Litzenberg et al. found that removing a Foley catheter could cause a significant change in the rotation of the prostate. Reference Litzenberg, Muenz and Archer14 In our study, we found minimal overlap between the voiding MRI urethra and the CT-defined urethra. The location of the MRI-voiding proximal urethra is likely more anatomically correct and suggested a more curved shape compared to the shape determined with a semi-rigid catheter in place (Figure 2).
Foremost among the limitations of this study is the small number of patients and the lack of multiple participants undergoing CT simulation with Foley catheter in place. This prevents us from drawing firm conclusions; rather, our observation of changes in the urethra positing between CT scan with Foley catheter and voiding MRI scan is hypothesis generating. Four patients underwent CT simulation where the urethra was defined as the geometric centre, which is a common practice for defining the proximal urethra. Reference Bucci, Spadinger and Hilts7,Reference Waterman and Dicker8 In these patients, we noted similar patterns of increased size, increased curvature and non-central location of the proximal urethra defined on MRI voiding scan, as compared to the geometric centre urethra using CT imaging. Another limitation of this study is the fact that three of the seven patients enrolled were unable to void under trial conditions, this may be due to patient discomfort of urinating in a public space or while supine.
Urination allows for clear visualisation of the proximal urethra by MRI; however, the act of voiding may change the bladder, prostate and pelvic floor muscle positions, making fusion with a full-bladder CT scan more difficult. Reference Dekura, Nishioka and Hashimoto10,Reference Rai, Sidhom, Lim, Ohanessian and Liney15 To overcome this, we initially complete a full-bladder MRI scan and we use the short voiding sequence MRI for defining the prostatic urethra position. In our analysis of the pre-void and voiding MRIs, we did not observe large changes in the prostate or pelvic floor musculature positioning. However, clinically we have observed such shifts in prostate and pelvic floor musculature with a Foley catheter simulation as compared to daily imaging during treatment with no catheter. This may result from pelvic floor muscle contraction secondary to discomfort from the Foley catheter during simulation and delineation of urethra by voiding MRI may overcome this challenge.
This trial supports further investigation into delineating the prostatic urethra for radiotherapy treatment planning using non-invasive MRI imaging. Future studies may include comparison of urethral contours for a series of patients undergoing treatment simulation using a voiding MRI scan and then a Foley catheter. This may allow for more accurate image segmentation and contouring of the prostatic urethra as well as evaluation of whether and how a Foley catheter may deform this structure. Further, this could enable additional studies to evaluate approaches aimed at reducing genitourinary toxicities in these patients by more effectively constraining the dose of radiation delivered to the proximal urethra.
Conclusion
This study provides a novel non-invasive technique to allow for visualisation of the proximal urethra during simulation scanning using a rapid sequence MRI protocol to visualise the urinary flow of patients voiding during the MRI scan.
Acknowledgements
Thank you to our medical physics and radiation therapy team for making this possible.
Conflicts of Interest
The authors declare none.
Financial Support
There was no financial support for the data presented in this manuscript.
Ethical Standards
The authors assert that all procedures contributing to this work comply with the ethical standards of the relevant national guidelines on human experimentation, the Belmont Report and with the Helsinki Declaration of 1975, as revised in 2008, and have been approved by the University of Wisconsin Institutional Review Board.