Hostname: page-component-78c5997874-ndw9j Total loading time: 0 Render date: 2024-11-10T12:53:34.750Z Has data issue: false hasContentIssue false

A balloon-borne instrument to measure total water content in low-level clouds

Published online by Cambridge University Press:  04 April 2001

Jeremy D Price
Affiliation:
Meteorological Office Research Unit, Cardington, Shortstown, Bedfordshire, MK42 0TH, UK
Tony A Jones
Affiliation:
Meteorological Office Research Unit, Cardington, Shortstown, Bedfordshire, MK42 0TH, UK
John Shields
Affiliation:
Meteorological Office Research Unit, Cardington, Shortstown, Bedfordshire, MK42 0TH, UK
Get access

Abstract

This paper describes a new instrument designed to measure cloud or fog total water content mounted on a tethered balloon facility. The instrument is an evaporator type, which is inexpensive and relatively easy to construct. The device measures total water content with an uncertainty of 0.3 g kg–1, and a resolution of 0.03 g kg−1, at a frequency of 1 Hz. This typically gives a spatial resolution of 5–20 m and allows turbulent statistics to be calculated. Tethered balloons provide an ideal platform for the instrument to study low cloud, allowing very detailed measurement of cloud structure to be made. Some typical data are presented as an illustration.

Type
Research Article
Copyright
© 1998 Meteorological Society

Access options

Get access to the full version of this content by using one of the access options below. (Log in options will check for institutional or personal access. Content may require purchase if you do not have access.)