Hostname: page-component-78c5997874-m6dg7 Total loading time: 0 Render date: 2024-11-14T05:12:55.822Z Has data issue: false hasContentIssue false

Plasma Deposition, Plasma Coating, and Ion Implantation to Improve Metallic Implants and Prostheses

Published online by Cambridge University Press:  17 June 2015

Get access

Extract

In spite of the success of surgical implants such as artificial hip joints, the materials used to make them are not always quite up to the job. Even stainless steel and titanium alloys can break under the enormous stress on load-bearing joints and corrode in the salty environment of the body. Deposits of inorganic salts can scratch weight-bearing surfaces, making artificial joints stiff and awkward. As a result, the lifetime of an implant is, at most, 10–15 years.

Metallurgists and engineers often treat the surfaces of metal parts to improve their properties. The use of advanced surface-treatment techniques such as glow-discharge ion implantation, plasma deposition, and plasma coating can significantly improve the strength, hardness, and corrosion resistance of metal implants. At the same time, these methods should also improve the biocompati-bility of the implanted devices.

Cobalt-based alloys are widely used for joint replacements. However, other compounds, such as titanium alloys, have excellent potential biocompatibility and interesting but imperfect mechanical properties. Stainless steel, namely 316L, has a good price-to-mechanical-properties ratio, but has the lowest corrosion resistance of the most commonly used metallic biomaterials.

Type
Aspects of Reconstructive Biomaterials
Copyright
Copyright © Materials Research Society 2000

Access options

Get access to the full version of this content by using one of the access options below. (Log in options will check for institutional or personal access. Content may require purchase if you do not have access.)

References

1. Mosser, A., Speisser, C., and Muster, D., Eur. Mater. Res. Soc. Monographs 3 (1992) p. 143.Google Scholar
2. Muster, D., Humbert, P., and Mosser, A., Bio-materials 11 (1990) p. 57.Google Scholar
3. Buchanan, R.A., Rigney, E.D., and Williams, J.R., J. Biomed. Mater. Res. 21 (1987) p. 355.Google Scholar
4. Higham, P.A., in Biomedical Materials, edited by Williams, J.P., Nichols, M.F., and Zingg, W. (Mater. Res. Soc. Symp. Proc. 55, Pittsburgh, 1986) p. 253.Google Scholar
5. Johansson, C.B., Lausmaa, J., Röstlund, T., and Thomson, P., J. Mater. Sci.: Mater. Med. 4 (1993) p. 132.Google Scholar
6. Shrivastava, S., Jain, A., Sethuramiah, A., Vankar, V.D., and Chorra, K.L., Nucl. Instrum. Methods Phys. Res., Sect. B 21 (1987) p. 591.Google Scholar
7. Dearnaley, G., Goode, P.D., and Minister, F.J., Vacuum 36 (1986) p. 807.Google Scholar
8. Billard, A., Steinmetz, J., and Frantz, C., Mater. Sci. Eng., A 140 (1991) p. 802.CrossRefGoogle Scholar
9. Saker, A., Leroy, Ch., Michel, H., and Frantz, C., Mater. Sci. Eng., A 140 (1991) p. 702.Google Scholar
10. Guemmaz, M., Mosser, A., Raiser, D., Grob, J.J., Cornet, A., and Paletto, S., in Beam-Solid Interactions for Materials Synthesis and Characterization, edited by Jacobson, D.C., Luzzi, D.E., Heinz, T.F., and Iwaki, M. (Mater. Res. Soc. Symp. Proc. 354 Pittsburgh, 1995) p. 243.Google Scholar
11. Rie, K.T., Stucky, T., Silva, R.A., Leitao, E., Bordji, K., Jouzeau, J.Y., and Mainard, D., Surf. Coat. Technol. 74–75 (1995) p. 973.Google Scholar
12. Hirvonen, J.K., J. Vac. Sci. Technol. 15 (1978) p. 1662.CrossRefGoogle Scholar
13. Buchanan, R.A., Rigney, E.D., and Williams, J.M., J. Biomed. Mater. Res. 21 (1987) p. 367.Google Scholar
14. Pivin, J.C., Pons, F., and Takadoum, J., J. Mater. Sci. 22 (1987) p. 1087.Google Scholar
15. Bordji, K., Jouzeau, J.Y., Mainard, D., Payan, E., Netter, P., Rie, K.T., Stucky, T., and Hage-Ali, M., Biomaterials 17 (1996) p. 929.Google Scholar
16. Bordji, K., Jouzeau, J.Y., Mainard, D., Payan, E., Delagoutte, J.P., and Netter, P., Biomaterials 17 (5) (1996) p. 491.Google Scholar
17. Sioshansi, P., Nucl. Instrum. Methods Phys. Res., Sect. B 19/20 (1987) p. 204.Google Scholar
18. Fayeulle, S. and Treheux, D., Nucl. lnstrum. Methods Phys. Res., Sect. B 19/20 (1987) p. 216.Google Scholar
19. Jianqiang, Z., Xiaozphong, Z., Zintang, G., and Hendge, L., in Biomedical Materials, edited by Williams, J.P., Nichols, M.F., and Zingg, W. (Mater. Res. Soc. Symp. Proc. 55, Pittsburgh, 1986) p. 229.Google Scholar
20. Martinella, R. and Giovanardi, S., Mater. Sci. Eng. 69 (1985) p. 247.Google Scholar
21. Martinella, R. and Giovanardi, S., Soc. Franç. Tribologie éd. (1985) p. 1.Google Scholar