Single crystal high purity CVD diamonds have been metallized and calibrated as photodiodes at the National Synchrotron Light Source (NSLS). Current mode responsivity measurements have been made over a wide range (0.2-28 keV) of photon energies across several beamlines. Linear response has been achieved over ten orders of magnitude of incident flux, along with uniform spatial response. A simple model of responsivity has been used to describe the results, yielding a value of 13.3±0.5 eV for the mean pair creation energy. The responsivity vs. photon energy data show a dip for photon energies near the carbon edge (284 eV), indicating incomplete charge collection for carriers created less than one micron from the metallized layer.