Hostname: page-component-78c5997874-fbnjt Total loading time: 0 Render date: 2024-11-10T09:06:56.617Z Has data issue: false hasContentIssue false

An Updated Gas Source Focused Ion Beam Instrument for TEM Specimen Preparation

Published online by Cambridge University Press:  21 February 2011

Reza Alani
Affiliation:
Gatan Inc., 6678 Owens Drive, Pleasanton, CA 94588
Joseph S. Jones
Affiliation:
Gatan Inc., 6678 Owens Drive, Pleasanton, CA 94588
Peter R. Swann
Affiliation:
Gatan Inc., 6678 Owens Drive, Pleasanton, CA 94588
Get access

Abstract

The construction and performance of an updated gas source precision ion milling system are described. The system is based on an existing focused ion beam machine which is able to image and mill selected areas of specimens that are too thick for TEM studies. The specimen image is formed using either secondary electrons or secondary ions, captured by a dual detector. The work chamber consists of three major components: the ion gun, the ion column and the specimen chamber. The ion gun is an electron impact ionization type with an optimized source size and allows the use of variety of gases. The updated system employs an objective lens with shorter focal length to enhance the resolution. The specimen chamber with an improved specimen eucentric stage, accepts side entry TEM specimen holders. This enables the specimen to move between the TEM and the instrument for further precision thinning as required without removal of the specimen from the holder and consequent risk of damage. The upgraded system resolves features <1μm in thickness. Its point milling rate for Ni is 1.4μm/min. The ability of the instrument for imaging and localized milling is demonstrated by a number of TEM images of semiconductors, metals, ceramics and composites.

Type
Research Article
Copyright
Copyright © Materials Research Society 1992

Access options

Get access to the full version of this content by using one of the access options below. (Log in options will check for institutional or personal access. Content may require purchase if you do not have access.)

References

1. Shaver, D.C. and Ward, B.W., Solid State Technology, Dec., 73.(1985)Google Scholar
2. Fisher, A.T. and Bentley, J., Proceedings of the 44th Annual Meeting of the EMSA, 830 (1986)Google Scholar
3. Walmsley, J.C. and Chern, D., EUREM 88, Inst. Phys. Conf. Ser. #, 93: 2, 439 (1988)Google Scholar
4. Vecchio, K.S. and Williams, D.B., Philosophical Magazine B,.57, # 4, 535 (1988)Google Scholar
5. Omlor, R.E., Robertson, E.. and Lloyd, P.F., Proceedings of the 46th Annual Meeting of the EMSA, 984 (1988)CrossRefGoogle Scholar
6. Lundquist, T.R., Alani, R. and Swann, P.R., Ultramicroscopy, 24, 27.(1988)Google Scholar
7. Lin, T., Alani, R. and Lambeth, D.N., Journal of Magnetism and Magnetic Materials,.78, 213 (1989)Google Scholar
8. Alani, R., Jones, J.S. and Swann, P.R., Hitachi Scientific Instrument News (in Japanese), 34, # 4, 3424 (1991).Google Scholar