No CrossRef data available.
Article contents
Magnetic Spin Valve Sensors with Different Geometry to Magnetic Bead Detection for Biosensor Application
Published online by Cambridge University Press: 26 February 2011
Abstract
We have demonstrated the use of highly sensitive spin valve sensors for the detection of micron magnetic beads. By using a ring type, cross type, and meander line type sensors, we were able to detect the presence of 2.8 μm size magnetic beads in real time by direct measurement of magnetic dipole fields from magnetic beads. The sensitivity of the ring, cross and meander line sensors were obtained about 50 μV/Oe, 7 μV/Oe, 30 μV/Oe and sensor output signals of 50 μV , 30 μV, 90 μV were obtained in an external applied field of 10 Oe and 1 mA sense current. Our results shows that ring, cross, and meander line shape spin valve sensors are very promising candidates for the detection of biomolecules with magnetic labels.
- Type
- Research Article
- Information
- Copyright
- Copyright © Materials Research Society 2008