Article contents
New CdTe Quantum Sensor, With Improved Spatial Resolution, for Fast X-Ray and γ-Ray Imaging
Published online by Cambridge University Press: 10 February 2011
Abstract
We have designed and tested a new compact, very sensitive (>1 μRad), X-ray quantum sensor for high energy (0.5 -10 MeV) flash imaging. Monte Carlo simulations have been computed for several interesting detector materials (CdZnTe, HgI2-a, BGO, CsI, LSO...). From these calculations, we obtain general design rules and curves which allow one to optimize the pixel size of a sensor to the dose levels. An experimental CdZnTe detector array and its performance is presented.
- Type
- Research Article
- Information
- Copyright
- Copyright © Materials Research Society 1998
References
1/
Derenzo, S.E., Moses, W.W., Proceedings of the « cristal 2000 » international workshop, Sept. 22–26 (1992) p. 125–135, Ed. FrontieresGoogle Scholar
2/
Briesmeister, J.F., Memorandum MCNP4, Newsletter, 4 Apr. 1991, in: MCNP a general Monte Carlo Code, Los Alamos, New MexicoGoogle Scholar
4/
Gigot, J-F., Gerstenmayer, J-L., Dain, L. Le and Mathy, F., 9th Int. Workshop on Room Temperature Semiconductor X- and γ-Rays Detectors, Associated Electronics and Applications Grenoble, France, 1995, Nucl. Instr. and Meth. A
380 (1996) 479–482
Google Scholar
5/
Tubiana, M., Dutreix, J., Dutreix, A., Jockey, P., Bases physiques de la radiothérapie et de la radiobiologie, Masson et Cie Ed. 1963, page 62
Google Scholar
6/
Gerstenmayer, J-L., Gigot, J-F., Izycki, M., Hennion, C., Dain, L. Le, Proceedings, SPIE-The International Society for Optical Engineering, SPIE Vol.2551 p. 234–238
Google Scholar
7/
Cuzin, M., De la tomographie à la la voludensitométrie: les capteurs pour le contrôle industriel et l'instrumentation médicale, Actes des journées d'électronique 93, LETI, Saclay 1993, France, February 1993
Google Scholar
8/
Gerstenmayer, J-L., Gigot, J-F., Vibert, P., Cuzin, M., Peyret, O., Dain, L. Le, Mathy, F. and Verger, L., 9th Int. Workshop on Room Temperature Semiconductor X- and γ-Rays Detectors, Associated Electronics and Applications Grenoble, France, 1995, Nucl. Instr. and Meth. A
380 (1996) 462–466.Google Scholar
- 2
- Cited by