Article contents
On the Active Volume of Cadmium Zinc Telluride Gamma-Ray Spectrometers
Published online by Cambridge University Press: 10 February 2011
Abstract
In this paper we develop quantitative models to predict the active volume of cadmium zinc telluride (CZT) detectors operated as gamma-ray pulse height spectrometers. Three cases are considered: a conventional planar detector, a unipolar device, and a detector in which electronic signal processing has been applied to correct for charge trapping effects. We find that existing detectors are very limited in their maximum attainable active volume, but unipolar devices with charge correction show promise for producing large active volume devices.
- Type
- Research Article
- Information
- Copyright
- Copyright © Materials Research Society 1998