The esophageal detector device (EDD) recently has been found to assess endotracheal (ET) tube placement accurately. This study describes the reliability of the EDD in determining the position of the ET tube in clinical airway situations that are difficult.
This was a prospective, randomized, single-blinded, controlled laboratory investigation. Two airway managers (an emergency-medicine attending physician and a resident) determined ET-tube placement using the EDD in five swine in respiratory arrest. The ET tube was placed in the following clinical airway situations: 1) esophagus; 2) esophagus with 1 liter of air instilled; 3) trachea; 4) trachea with 5 ml/kg water instilled; and 5) right mainstem bronchus. Anatomic location of the tube was verified by thoracotomy of the left side of the chest.
There was 100% correlation between the resident and attending physician's use of the EDD. The EDD was 100% accurate in determining tube placement in the esophagus, in the esophagus with 1 liter of air instilled, in the trachea, and in the right mainstem bronchus. The airway managers were only 80% accurate in detecting tracheal intubations when fluid was present.