Hostname: page-component-cd9895bd7-jn8rn Total loading time: 0 Render date: 2024-12-27T09:17:09.757Z Has data issue: false hasContentIssue false

Quantification of physical activities by means of ambulatory accelerometry: A validation study

Published online by Cambridge University Press:  04 January 2001

JOHANNES B.J. BUSSMANN
Affiliation:
Department of Rehabilitation, Erasmus University Rotterdam, The Netherlands
JOKE H.M. TULEN
Affiliation:
Department of Psychiatry, Erasmus University Rotterdam, The Netherlands
EVA C.G. VAN HEREL
Affiliation:
Department of Rehabilitation, Erasmus University Rotterdam, The Netherlands
HENK J. STAM
Affiliation:
Department of Rehabilitation, Erasmus University Rotterdam, The Netherlands
Get access

Abstract

The objective of the study was to assess the validity of an activity monitor (AM) within a psychophysiological study. The AM was based on four body-fixed accelerometers and discriminated postures, transitions, and dynamic activities. Three subjects participated in each of two 4-hr sessions. During each session, consisting of two protocols, ambulatory accelerometer and heart rate measurements were made. The output of the AM was compared with simultaneously recorded video tapes. An overall agreement between AM and video of 88 and 96% was found. The number of transitions and dynamic periods, and the duration of activities were well determined. Posture-related heart rate changes were demonstrated. A three-sensor configuration hardly influenced the validity scores. The AM appeared to be a valid instrument to quantify aspects of physical activity, and offers new possibilities for ambulatory psychophysiological research.

Type
Research Article
Copyright
© 1998 Society for Psychophysiological Research