Hostname: page-component-78c5997874-j824f Total loading time: 0 Render date: 2024-11-14T23:02:08.875Z Has data issue: false hasContentIssue false

10Be Analyses with a Compact AMS Facility—Are Bef2 Samples the Solution?

Published online by Cambridge University Press:  18 July 2016

L Wacker*
Affiliation:
Institute for Particle Physics, ETH Hönggerberg, CH-8093 Zürich, Switzerland.
M Grajcar
Affiliation:
Institute for Particle Physics, ETH Hönggerberg, CH-8093 Zürich, Switzerland.
S Ivy-Ochs
Affiliation:
Institute for Particle Physics, ETH Hönggerberg, CH-8093 Zürich, Switzerland.
PW Kubik
Affiliation:
Paul Scherrer Institut (PSI), c/o Institute for Particle Physics, ETH Hönggerberg, CH-8093 Zürich, Switzerland.
M Suter
Affiliation:
Institute for Particle Physics, ETH Hönggerberg, CH-8093 Zürich, Switzerland.
*
Corresponding author. Email: wacker@phys.ethz.ch.
Rights & Permissions [Opens in a new window]

Abstract

Core share and HTML view are not available for this content. However, as you have access to this content, a full PDF is available via the ‘Save PDF’ action button.

The injection of 10BeF- instead of 10BeO- into a compact accelerator mass spectrometry system with a terminal voltage of 0.58 MV was investigated, because BF- molecules are unstable and isobaric interference of 10B with 10Be can thus be significantly reduced. We describe the method we developed to prepare BeF2 samples. 10Be was measured in a segmented gas ionization detector. Separation of 10Be from 10B could be achieved both for ions in the 1+ charge state with an energy of 0.8 MeV and in the 2+ charge state with an energy of 1.4 MeV. The 2+ ions are better separated, whereas the 1+ charge state has a higher transmission. 10Be/9Be ratios (~10-12) in a suite of rock samples were successfully determined for exposure dating in either charge state and compared with measurements made on the 6MV tandem.

Type
Articles
Copyright
Copyright © 2004 by the Arizona Board of Regents on behalf of the University of Arizona 

References

Döbeli, M, Kottler, C, Stocker, M, Weinmann, S, Synal, HA, Grajcar, M, Suter, M. Forthcoming. Gas ionization chambers with silicon nitride windows for the detection and identification of low energy ions. Nuclear Instruments and Methods in Physics Research B. Google Scholar
Dreuw, A, Sommerfeld, T, Cederbaum, LS. 2002. Short-and long-lived electronic states of BF- . Journal of Chemical Physics 116(14):6039–44.Google Scholar
Grajcar, M, Döbeli, M, Kubik, PW, Maden, C, Suter, M, Synal, H-A. Forthcoming. 10Be measurements with terminal voltages below 1 MV. Nuclear Instruments and Methods in Physics Research B. Google Scholar
Ochs, M, Ivy-Ochs, S. 1997. The chemical behavior of Be, Al, Fe, Ca and Mg during AMS target preparation from terrestrial silicates modeled with chemical speciation calculations. Nuclear Instruments and Methods in Physics Research B 123:235–40.Google Scholar
Stocker, M, Bertschinger, R, Döbeli, M, Grajcar, M, Jakob, S, Scheer, J, Suter, M, Synal, H-A. Forthcoming. Status of the PSI/ETH Compact AMS facility. Nuclear Instruments and Methods in Physics Research B. Google Scholar
Synal, H-A, Bonani, G, Döbeli, M, Ender, RM, Gartenmann, P, Kubik, PW, Schnabel, Ch, Suter, M. 1997. Status report of the PSI/ETH AMS Facility. Nuclear Instruments and Methods in Physics Research B 123:62–8.Google Scholar
Synal, H-A, Jacob, S, Suter, M. 2000. The PSI/ETH small radiocarbon dating system. Nuclear Instruments and Methods in Physics Research B 172:17.Google Scholar
Zhao, X-L, Litherland, AE, Doup, JP, Kieser, WE. 2002. AMS measurements of 10Be using BeF- . Proceedings of the 9th International Accelerator Mass Spectrometry Conference, 9–13 September 2002, Nagoya, Japan. Poster presentation PTPd-5.Google Scholar