Hostname: page-component-cd9895bd7-dk4vv Total loading time: 0 Render date: 2024-12-28T04:07:16.583Z Has data issue: false hasContentIssue false

Low-Level Gas Proportional Counting in An Underground Laboratory

Published online by Cambridge University Press:  18 July 2016

H H Loosli
Affiliation:
Physics Institute, University of Bern, Sidlerstrasse 5, CH-3012 Bern, Switzerland
Martin Heimann
Affiliation:
Physics Institute, University of Bern, Sidlerstrasse 5, CH-3012 Bern, Switzerland
Hans Oeschger
Affiliation:
Physics Institute, University of Bern, Sidlerstrasse 5, CH-3012 Bern, Switzerland
Rights & Permissions [Opens in a new window]

Abstract

Core share and HTML view are not available for this content. However, as you have access to this content, a full PDF is available via the ‘Save PDF’ action button.

Results of measurements in an underground laboratory, 70m of water equivalent below surface, are given. Background values of proportional gas counters with volumes between 16cc and 1.5L are lower by a factor of 2—4.5 compared to the values in the previously used laboratory (7m w.e.). High counting gas pressures, up to 5 at CH4 and up to 36 at P-10, enable the use of relatively small counters with correspondingly small background contributions from the walls. A separation of the residual background into different components is attempted, distinguishing a pressure dependent volume effect and surface correlated contributions. It can be shown that the selection of radioactively pure counter construction material is very important for a good low level counting system.

Type
Techniques
Copyright
Copyright © The American Journal of Science 

References

Nydal, Reidar, Gulliksen, Steinar, and Lövseth, Knut, 1975, Proportional counters and shielding for low level gas counting, in Internatl conf on low radioactivity measurements and applications, High Tatras: p 7784.Google Scholar
Oeschger, Hans, 1963, Low-level counting methods, in Radiocarbon dating: IAEA, Vienna, p 1334.Google Scholar
Oeschger, Hans and Loosli, H H, 1975, New developments in sampling and low level counting of natural radioactivity, in Internatl conf on low radioactivity measurements and applications, High Tatras: p 1322.Google Scholar
Oeschger, Hans and Wahlen, Martin, 1975, Low level counting techniques: Ann Rev Nuclear Science, v 25, p 423463.Google Scholar
Pulfer, Peter, 1974, Grundlagen zum Bau eines Tieflabors zur Messung schwach radioaktiver Proben: Lizentiatsarbeit Univ Bern, unpub.Google Scholar
Schoch, Hilla, Bruns, Michael, Münnich, K O, and Münnich, Marianne, 1980, A multicounter system for high precision 14C measurements, in Stuiver, Minze and Kra, Renee, eds, Internatl radiocarbon conf, 10th, Proc: Radiocarbon, v 22, no. 2, p 442447.Google Scholar
Stuiver, Minze, Robinson, S W, and Yang, I C, 1979, 14C dating up to 60,000 years bp with high efficiency proportional counters, in Berger, Rainer and Suess, H E, eds, Radiocarbon dating, Internatl radiocarbon conf, 9th, Proc: Berkeley/Los Angeles, Univ California Press, p 202215.Google Scholar
Tans, P P and Mook, W G, 1979, Design, construction and calibration of a high accuracy carbon-14 counting set up: Radiocarbon, v 21, p 2240.Google Scholar