Median nerve somatosensory evoked potential (SSEP) grading is useful for predicting functional outcome after traumatic and anoxic brain injury.Reference Guerit 1 - Reference Houlden, Taylor and Feinstein 3 Bilaterally preserved SSEPs obtained in the early stage after traumatic brain injury is associated with good outcome, whereas absent ones are associated with poor outcomes. With anoxic-ischemic insults, the absence of the N20 response with median nerve stimulation is almost always associated with an outcome no better than vegetative state, whereas N20 preservation is prognostically unreliable.Reference Wijdicks, Hijdra, Young, Bassetti and Weibe 4 , Reference Rossetti, Oddo, Logroscino and Kaplan 5
Giant SSEPs have been found mainly in progressive myoclonic epilepsyReference Shibasaki, Yamashita, Neshige, Tobimatsu and Fukui 6 and occasionally in nonconvulsive status epilepticus,Reference Schorl 7 in which there is increased cortical excitability. The finding of giant SSEPs in the context of acute encephalopathies is very unusual and has not been addressed for its pathogenesis and prognostic significance. In this article, we describe giant SSEPs that occurred in the context of frequent epileptiform discharges following acute brain insults. These cases illustrate the importance of interpreting the finding of giant SSEPs in association with etiology and pathophysiology.
Methods
SSEPs
SSEPs were obtained using a Cadwell Cascade (Cadwell Instruments, Kennewick, WA) at Sunnybrook Health Science Centre, Toronto, Ontario, for patient 1 and a VikingQuest (Viasys/Nicolet) at University Hospital, London, Ontario, for patient 2. For patient 1, the SSEP studies were obtained 1 day postoperatively (1 day after the onset of tonic-clonic seizures). The studies were repeated 5 days postoperatively. For patient 2, only one SSEP study was obtained (2 days after the onset of axial myoclonus). The median nerves were alternately stimulated by surface stimulators at each wrist with a stimulus intensity two times that necessary to evoke a visible thenar muscle twitch. The stimulus pulse duration was 0.1-0.2 milliseconds and the stimulus rate was 3.1-4.7 Hz depending on institution. SSEP recordings were obtained from surface electrodes (130 Meditrace Mini, Kendall) and from subdermal needle electrodes (Rochester Electromedical, Miami, FL) or electroencephalograph (EEG) disc electrodes (Natus Neurology Inc., Middleton, WI) on the scalp. Skin/electrode impedance was kept below 5000 ohms. The locations of the recording electrode pairs were (1) ipsilateral Erb’s point (EP)–frontal pole electrode (Fpz); (2) the C2 spinous process (Cv2)–Fpz; and (3) the scalp overlying the contralateral somatosensory cortex (C3’ or C4’)–Fpz. For patient 1, two additional channels were used that were not used in patient 2; contralateral somatosensory cortex (C3’ or C4’)–ipsilateral somatosensory cortex (C3’ or C4’) and contralateral somatosensory cortex (C3’ or C4’)–contralateral EP. For patient, 2 one additional channel was used that was not used in patient 1; contralateral EP–ipsilateral EP (EP1–EP2). The SSEP was the averaged result of at least 500 stimulus presentations. The amplifier gain was 20 µV/division and the recording bandpass was 3-2000 Hz. The sweep time was 75 milliseconds on the left and 80 milliseconds following right median nerve stimulation for patient 1 and 50 milliseconds bilaterally for patient 2. The display scale was adjusted for optimum presentation of the SSEP waveforms (up to 1.2 µV/division). Two SSEPs from each limb were superimposed for waveform reproducibility.
EEG
For patient 1, routine 20-channel EEGs were performed with sampling rates of 256 Hz/channel. A bipolar montage with standard disk electrodes was used, and impedances were kept under 5 KOhms. Recordings were 20-30 minutes in duration and were obtained on the same day as the SSEP studies for patient 1 (1 and 5 days postoperatively). Auditory and nail bed pressure stimuli were applied when patients were comatose.
For patient 2, continuous EEG was obtained 2 days after she became unresponsive (same day as SSEP for patient 2). A four-channel bipolar EEG monitor (Datex-Ohmeda S/5 M-EEG Module; model #898683-00 plugged into Datex-Ohmeda Critical care monitor, GE Healthcare, Helsinki, Finland), surface electrodes (3M Health Care2670-5), and a sub-hairline montage was used. The sampling rate was 100 Hz.
Results
Patient 1
Patient 1 was an 87-year-old male who underwent cardiac surgery at Sunnybrook Health Science Centre, Toronto, Ontario. His medical history included prior bypass surgery, critical aortic stenosis, angina, peripheral vascular disease, hyperlipidemia, hypertension, and atrial fibrillation. He underwent a repeat sternotomy, aortic valve replacement, and repair of the right coronary artery. The time on the bypass pump was 158 minutes and cross-clamp time was 108 minutes. He received 4 grams of tranexamic acid before surgery and a total of 14 grams during surgery as well as 2 grams of cefazolin before and 1 gram after surgery.
One day postoperatively, he developed recurrent generalized tonic-clonic seizures that lasted 5-60 seconds each. The patient was started on phenytoin and propofol was initiated given persistent convulsions. Over the next few days, propofol was successfully tapered with no recurrent seizures. Clonazepam was then introduced. Hypoxic injury was presumed.
SSEP studies
One day postoperatively, the peripheral nerve compound action potential recorded from EP–Fpz and the subcortically generated P/N13 waveform recorded from Cv2–Fpz were normal bilaterally. EP–P/N13 and P/N13 conduction times were normal bilaterally. Both recording montages (EP–frontal scalp [Fz], Cv2–Fz) showed unusual large amplitude late responses likely recorded from the Fz electrode. The cortically generated response recorded from somatosensory cortex (contralateral to side of stimulation)–Fpz had giant N20–P25 amplitude bilaterally; 34 µV on the right and 15 µV on the left side of the brain (upper limit of normal=5.5 µV]) (Figure 1).Reference Chiappa 8 Even N20 amplitude (measured from baseline to peak of N20) had giant amplitude bilaterally; 11 µV on the right and 6.3 µV on the left side of the brain (upper limit of normal=2.7 µV).Reference Chiappa 8 Five days postoperatively, the SSEP study was repeated and showed a large decrease of N20–P25 amplitude bilaterally compared with that obtained 1 day postoperatively; 4.8 µV on the right and 6 µV on the left side of the brain (compared with 34 µV and 15 µV, respectively) (Figure 2). N20 amplitude was also decreased compared with the previous study; 3.3 µV on the right and 4 µV on the left side of the brain (compared with 11 µV and 6.3 µV, respectively).
EEG studies
One day postoperatively, the EEG showed rare interictal spike and wave discharges generally less than one every 10 seconds and no more frequent than 1 Hz noted maximally around the right frontal/temporal regions (Figure 3A). There were no changes in the EEG after auditory or nail bed pressure stimulation. After propofol was tapered, the discharges became more frequent and spread bilaterally with the left arm jerking clinically, eventually becoming confluent (Figure 3B). The patient had a generalized seizure following this recording; propofol was immediately resumed. Five days postoperatively, generalized slowing was noted on the EEG with prominent theta (in the context of multiple metabolic abnormalities which gradually resolved) and beta activity (reflecting medication effects). There were no stimulus-induced rhythmic, periodic, or ictal discharges.
Imaging and outcome
The patient emerged from coma 5 days postoperatively. A magnetic resonance imaging of the brain conducted 8 days postoperatively demonstrated atrophy and chronic ischemic changes but no acute infarct. He had a lengthy intensive care unit stay that was complicated by atrial fibrillation, pneumonia, and renal failure, and eventually the patient required tracheostomy. However, he made a slow but continuous recovery over time. Upon discharge to rehabilitation, he was ambulating with assistance and his swallowing had improved dramatically. He provided signed informed consent to publish details of his case in accordance with our hospital research ethics board protocol.
Patient 2
Patient 2 was an 82-year-old female who underwent laparoscopic-assisted radical nephroureterectomy for an invasive transitional cell carcinoma at University Hospital, London, Ontario. Her postoperative course was uncomplicated; however, 3 days postoperatively she suffered pulseless electrical cardiac activity (PEA) and had hospital cardiopulmonary resuscitation measures for approximately 10 minutes before return of spontaneous circulation. She was transferred to the intensive care unit and was sedated, intubated, and required high vasopressor support. She subsequently developed significant axial myoclonus that day and was started on treatment of propofol and phenytoin and subsequently changed to valproic acid at the time when EEG and SSEP testing were performed.
SSEP study
The SSEP study (performed 2 days after PEA) showed a mildly delayed right EP response (probably related to a mild neuropathy from her chronic renal failure) but the morphology of the subcortically generated P/N13 waveform, the EP–P/N13 conduction time and the P/N13–N20 conduction time were normal bilaterally. As with patient 1, both recording montages (EP–Fz, Cv2–Fz) showed unusually large amplitude late responses. The cortically generated response recorded from somatosensory cortex (contralateral to side of stimulation)–Fpz had giant N20–P25 amplitude bilaterally; 29 µV on the right and 16.2 µV on the left side of the brain (Figure 4). The N20 amplitude (measured from baseline to peak of N20) was high amplitude: 3.8 µV on the right and 2.3 µV on the left side of the brain. The P/N13–N20 interpeak latency was normal bilaterally.
EEG study
The EEG was also performed 2 days after PEA and revealed generalized periodic epileptiform discharges against a suppressed background (Figure 5). There were no stimulus-induced rhythmic, periodic, or ictal discharges.
Imaging and outcome
One day after PEA cardiac arrest, computed tomography of the head showed generalized brain atrophy and chronic ischemic changes (in keeping with the patient’s age) but no acute intracranial abnormality. Two days after PEA, the neurological examination revealed absent corneal, gag, and pupillary reflexes. Given her clinical findings, the patient’s prognosis was quite grim so substitute decision-makers decided to remove life-sustaining therapies and family wishes were honoured.
Discussion
This is the first time that the relationship between giant SSEPs and EEG has been reported in patients whose coma was related to complications following surgery. The same phenomenon has been reported in a traumatic brain-injured patients.Reference Schorl 7 Giant median nerve SSEPs have been more commonly reported in patients with myoclonic epilepsy.Reference Takeuchi, Touge, Miki, Yamada, Deguchi and Nishioka 9 - Reference Shibasaki and Hallett 11
In addition to giant SSEP N20–P25 amplitudes, we found the amplitude of N20 itself (onset to negative peak) was also increased. N20 is thought to originate mainly from the thalamus with some contributions from thalamocortical radiation fibres and primary somatosensory cortex. We agree that cell bodies contribute to N20 amplitude because the temporary amplitude increase we observed in patient 1 can only occur by temporary facilitation or disinhibition of cell bodies (in thalamus and/or somatosensory cortex).
We propose the favourable outcome in patient 1 (despite initial seizures and transient giant SSEPs) was related to a reversible effect of drugs, notably tranexamic acid, used during cardiac surgery. We had previously noted this effect in two separate centers after the switch of aprotinin to high-dose tranexamic acid.Reference Murkin, Falter, Granton, Young, Burt and Chu 12 Both tranexamic acid and cefazolin, used in this patient, can block gamma-amino butyric acid type A receptors, causing a reversible reduction in inhibition in the brain.Reference Bechtel, Slaughter and Moore 13 - Reference Santos, Araujo, Fanchini, Pena, Delerue-Matos and Montenegro 15
The second patient suffered severe anoxic-ischemic cortical injury that proved fatal. The finding of giant SSEPs in this situation is unusual. We speculate that this resulted from an imbalance of neurotransmitter functions, with either decreased gamma-amino butyric acid or augmented excitatory transmitters, especially glutamic acid, or both, as has been described in neonatal hypoxic encephalopathy with seizures.Reference Takeuchi, Touge, Miki, Yamada, Deguchi and Nishioka 9 However, unlike patient 1, whose giant SSEPs were likely from a reversible drug effect, the neuronal damage was permanent and severe in patient 2.
Median nerve SSEP N20–P25 amplitude in the early stage after coma is used for prediction of clinical outcome,Reference Guerit 1 - Reference Houlden, Taylor and Feinstein 3 so it is important to consider the relationship between epileptiform discharges and SSEP amplitude when interpreting the SSEP. Unfortunately, this relationship was not elucidated in previous SSEP prognostic studies (seizure history was not considered in previous studies). It is possible that the prognostic power of the SSEP was weakened in previous studies because some patients with present but abnormal SSEPs may have had underlying facilitation or disinhibition that erroneously increased cortically generated SSEP amplitude. For example, in a previous study, we showed that 4 of 30 patients with bilaterally normal SSEPs on day 1 had bad outcomes.Reference Houlden, Taylor and Feinstein 3 Perhaps those four patients had concomitant epileptiform discharges and an erroneously increased SSEP N20–P25 amplitude on day 1. Given the findings in the current study, the prognostic information gained from intact SSEPs in comatose patients should be used with caution when epileptiform discharges are present.
In the current study, both patients 1 and 2 had giant SSEP N20–P25 amplitude and epileptiform discharges. The relationship between epileptiform discharges and SSEP amplitude was shown in patient 1 who had a marked reduction in epileptiform discharges between days 1 and 5 that was coincident with a decrease (normalization) in the cortically generated SSEP bilaterally. The patient emerged from coma in the same time frame.
Conclusion
Those who use cortically generated SSEP amplitude for predicting outcome in comatose patients should consider the presence of epileptiform discharges (detected by EEG) as a potential confounding factor.
Disclosures
DH, BM, LN, and BY do not have anything to disclose. MB is a researcher for and has received research support from the Canadian Stroke Network and ResMed Corporation and is a researcher for and received a grant from APF Innovation.