Electroconvulsive therapy (ECT) is the most effective treatment for major depressive disorder. 1,Reference Ebmeier, Donaghey and Steele2 However, its use is restricted, mostly owing to concerns about cognitive side-effects. Reference Eranti and McLoughlin3 Modifications to the treatment procedure that could reduce the cognitive side-effect burden are therefore of great clinical interest. Since the introduction of repetitive transcranial magnetic stimulation, it was noticed that a rare side-effect of this treatment was the induction of seizures, when it was used at high frequencies, high intensities, or with long train durations. The possibility and safety of deliberately inducing seizures with repetitive magnetic fields was first tested in non-human primates. Reference Lisanby, Luber, Finck, Schroeder and Sackeim4–Reference Lisanby, Sackeim, Dwork, Underwood, Wang, Kassir, Luber and Arango8 In 2000, the first human patient received a course of four magnetic seizure therapy sessions using 40 Hz at 100% intensity of the repetitive transcranial magnetic stimulation equipment available at that time. Reference Lisanby, Schlaepfer, Fisch and Sackeim9 A second patient was successfully treated with a full course of magnetic seizure therapy with similar parameters. Reference Kosel, Frick, Lisanby, Fisch and Schlaepfer10 Both patients tolerated the treatments well and responded to the treatment well. Following this, ten patients with depression received two magnetic seizure therapy sessions each during a course of ECT, in a masked, randomised within-participant cross-over trial. This time a custom-modified device capable of producing 50 Hz stimulation at a peak magnetic field of 1.2 T was used. Acute side-effects of magnetic seizure therapy were milder compared with ECT. Reference Lisanby11,Reference Lisanby, Luber, Schlaepfer and Sackeim12 Although seizures were induced in all patients, it was observed that in three patients the seizure threshold was at the maximum output of the device, highlighting the need for technological improvements in the equipment. Twenty patients with depression were treated with a full course of magnetic seizure therapy using the same 50 Hz magnetic seizure therapy device in 2003. Reference Lisanby, Husain, Morales, Thornton, White, Payne, Rush and Sackheim13 Magnetic seizure therapy improved depression scores and patients demonstrated remarkably rapid reorientation with few side-effects, although improvements were smaller than those seen in the simultaneously treated ECT group. Reference White, Amos, Zhang, Stool, Husain, Thornton, Downing, McClintock and Lisanby14 However, the maximum stimulation of 400 pulses per session was estimated to be on average only 1.3 times the magnetic seizure threshold, perhaps contributing to the suboptimal antidepressant efficacy, similar to some modes of ECT which are highly sensitive to dosage relative to seizure threshold. Reference Sackeim, Prudic, Devanand, Nobler, Lisanby, Peyser, Fitzsimons, Moody and Clark15
These early trials were a proof of the principle that magnetic seizure therapy could induce therapeutic seizures in a clinical setting, but they also indicated the need for improved magnetic seizure therapy devices (as discussed by Lisanby & Peterchev). Reference Lisanby, Peterchev, Marcolin and Padberg16 A new prototype device (Magstim Company Ltd, Whitland, Carmarthenshire, Wales) capable of stimulating continuously at 100 Hz at maximum stimulator output (1.2 T at the coil surface) for up to 10 s became available for animal use in 2004. This device was used for the first time in 2004 in a study of rhesus monkeys. Reference Lisanby, Peterchev, Marcolin and Padberg16 Since then, 275 magnetic seizure therapy sessions have been successfully performed in 11 rhesus monkeys. Seizures have been induced in all sessions and 100 Hz magnetic seizure therapy still demonstrates fewer cognitive side-effects in the monkey model than conventional ECT. Reference Peterchev, Kirov, Ebmeier, Scott, Husain and Lisanby17 A version of the 100 Hz magnetic seizure therapy device designed specifically for human use (the Magstim Theta; Magstim Company Ltd, Whitland, Carmarthenshire, Wales) became available in the middle of 2006. The current study was designed as a pilot to examine the feasibility of magnetic seizure therapy at 100 Hz in patients, its safety and side-effects, and recovery times compared with ECT.
Method
Patients
Eleven patients diagnosed with treatment-resistant major depressive episodes in the context of either recurrent major depression or schizoaffective disorders according to DSM–IV 18 criteria, who had been referred for ECT, were enrolled in this pilot study. Demographic details are presented in Table 1. Patients were treated in Whitchurch Hospital, Cardiff, Wales, and the Royal Edinburgh Hospital, Scotland. Eight of the patients were already receiving ECT, and one of their regular (twice weekly sessions) was substituted with magnetic seizure therapy. The remaining three patients received magnetic seizure therapy before ECT: two of these continued with ECT, the third decided against it. Local research ethics committee approval was obtained at both centres and patients gave written informed consent following the approved protocols. In accordance with usual clinical practice of ECT delivery in the UK, antidepressant medication was not stopped during the treatment. Every patient received at least one antidepressant, six were also taking one or more antipsychotics, and two patients (patients 5 and 10 in the table) were also taking sodium valproate.
Electroconvulsive therapy | Magnetic seizure therapya | |||||||
---|---|---|---|---|---|---|---|---|
Patient | Gender, age (years), setting | Seizure threshold, mC | Seizure duration, s Motor/EEG | Orientation time, min:s | Pulses, n | Position of coil | Seizure duration, s Motor/EEG | Orientation time, min:s |
1 | Female, 35, C | 126 BL | 36/60 | 26:49 | 250 | Vertex | 0/0 | - |
500 | Vertex | 25/21 | 4:36 | |||||
2 | Male, 21, C | 80 BL | 26/32 | 22:45 | 500 | Frontal | 86/83 | 5:23 |
3 | Female, 41, C | 841 BL | 42/49 | 25:01 | 250 | Vertex | 9.5/NA | - |
500 | Vertex | 21/NA | 4:20 | |||||
500b | Frontal | 0/- | - | |||||
1000b | Frontal | 70/NA | 4:40 | |||||
4 | Female, 47, E | 80 BL | 30/42 | 29:20 | 1000 | Frontal | 0/0 | - |
1000 | Vertex | 21/21 | 8:41 | |||||
5 | Female, 56, E | 170 BL | 30/31 | 22:30 | 1000 | Vertex | 28/29 | 5:36 |
6 | Female, 28, E | 80 BL | 22/38 | 21:10 | 1000 | Frontal | 0/NA | - |
1000 | Vertex | 18/NA | 9:00 | |||||
7 | Female, 48, E | 204 BL | 33/39 | 28:00 | 1000 | Frontal | 22/NA | 8:43 |
8 | Male, 39, C | No ECT | - | - | 600 | Frontal | 0/0 | 9:00 |
no seizure | ||||||||
600b | Vertex | 15/NA | 9:25 | |||||
9 | Male, 28, C | 80 UL | 28/28 | 16:00 | 600 | Vertex | 49/85 | 9:41 |
10 | Female, 70, C | 80 UL | 54/86 | 13:10 | 600 | Vertex | 0/0 | 7:48 |
no seizure | ||||||||
11 | Female, 56, E | 46 UL | 19/NA | 11:13 | 1000 | Vertex | 11/NA | 6:35 |
Magnetic seizure therapy
We used two custom-built Magstim Theta devices (Magstim Ltd, Whitland, Carmarthenshire, Wales). This stimulator is capable of producing 100 Hz magnetic stimuli at 1.2 T (at the centre of the coil) with a biphasic waveform with a pulse width of 340–400 μs for up to 10 s duration (i.e. a maximum of 1000 pulses). We used a round coil with an 80 mm average diameter (47 mm inside diameter, 115 mm outside diameter). For positioning of the coil we used standard 10–20 electroencephalogram (EEG) positions. The middle of the coil was applied firmly to the head of the patients, and positioned over Cz for vertical and Fz for frontal stimulation for up to 10 s. The direction of current induced in the brain was counter-clockwise. The inside of the coils heats from 20°C to 130°C after 1000 pulses at 100% output stimulation; therefore, coils were cooled down to 5–10°C in a refrigerator prior to stimulation and were changed if a patient required restimulation. All treatments were given at 100 Hz frequency and at maximum stimulator output. When a patient was restimulated, we allowed at least 20 s between stimulations. Staff and patients wore ear protectors during magnetic seizure therapy.
Anaesthesia
For anaesthesia we used intravenous etomidate (0.15–0.3 mg/kg) as it does not cause an increase in the seizure threshold and might even reduce it. Reference Khalid, Atkins and Kirov19 Muscle relaxation was achieved with intravenous succinylcholine; since patients recover more quickly from magnetic seizure therapy the dose was generally lower than that routinely used in ECT (0.5–1.0 mg/kg). Reference White, Amos, Zhang, Stool, Husain, Thornton, Downing, McClintock and Lisanby14
Seizure monitoring
Seizure duration during magnetic seizure therapy was measured from the start of stimulation to the termination of the observed seizure. Reference Lisanby, Luber, Schlaepfer and Sackeim12 Electroencephalogram seizure expression was monitored via bilateral fronto-mastoid EEG using magnetic resonance image-compatible plastic electrodes to prevent electrode heating during therapy.
Orientation assessment
Recovery of orientation after magnetic seizure therapy/ECT was assessed by asking the patient for their name, date of birth, age, place and day of the week. The point of orientation recovery was defined as the time when a patient was able to recall four of these five items.
Results
The first treatment session with the new device took place in Cardiff in June 2006. The patient was a 35-year-old woman, who had already received five bi-temporal ECT treatments, administered at 195.8 mC. For magnetic seizure therapy, the coil was positioned over the vertex (Cz). Stimulation with 250 pulses produced no seizure. She was restimulated 52 s later with 500 pulses and had a visible motor seizure of 25 s; EEG duration was about 21 s, but the end-point was difficult to estimate as there was no post-ictal suppression (EEG trace available from the authors on request).
Orientation was recovered after 4 min 36 s. Immediately upon awakening, the patient achieved a Mini-Mental State Examination Reference Folstein, Folstein and McHugh20 score of 27/30 points. On the next day, the patient's score was at the pre-ECT level of 30 points. A battery of further cognitive tests that included tests for verbal and visual memory, verbal fluency and executive speed was also administered and no relevant changes in performance from baseline were found (results not presented).
We have since treated 10 further patients. In order to explore optimal parameters of stimulation for this new procedure, we applied different numbers of pulses and changed the positioning of the coil between Cz and Fz. The results for each patient and the corresponding settings for their ECTs are presented in Table 1.
Seizures were elicited in 10 of the 11 patients. The one who did not fit was stimulated with only 600 pulses. Vertex stimulation appeared to be more effective in inducing seizure activity (Table 1; see patients 3, 4, 6 and 8). The mean duration of successful seizures was 31.3 s, range 9.5–86 s.
Orientation was recovered much faster after magnetic seizure therapy than after ECT. The mean time to recovery after successful seizures was 7 min 12 s (s.d.=2 min 7 s, range 4 min 20 s to 9 min 41 s). We compared these results with the recovery times of the same patients during their nearest ECT session(s) taking care that the order of ECT and magnetic seizure therapy sessions used for the calculation was approximately balanced. The mean recovery time after ECT was 26 min 35 s. When the recovery times of the nine patients who had both ECT and magnetic seizure therapy were compared in a paired-samples t-test, magnetic seizure therapy was shown to result in 15 min 35 s quicker recovery, and despite the small numbers, this result was highly significant at P<0.0001.
Patients uniformly commented that they felt less confused after magnetic seizure therapy. Side-effects of 100 Hz magnetic seizure therapy were restricted to the usual myotonic movements observed after etomidate anaesthesia. No serious immediate adverse events resulted from the use of magnetic seizure therapy.
Discussion
We report the first use of a new magnetic seizure therapy device capable of sustaining maximum stimulator output for 10 s at 100 Hz (1000 pulses). We treated 11 patients with a total of 18 stimulations. To explore the range of seizure thresholds, we used a different number of pulses and two positions of the coil: over vertex (Cz) or pre-frontally (Fz). Seizures were elicited in 10 of the 11 patients. The one who did not fit received only 600 pulses over the vertex. This was a 70-year-old woman who was on valproate; both her age and anticonvulsant medication could account for the difficulty to elicit a seizure. Eight more patients received stimulation over the vertex of between 250 and 1000 pulses. One of these patients (our first patient) did not have a seizure when we used 250 pulses, but she fitted when restimulated at 500 pulses. These findings correspond to the previous observations that the mean seizure threshold with 50 Hz magnetic seizure therapy was at 268 or 320 pulses. Reference Lisanby, Luber, Schlaepfer and Sackeim12,Reference White, Amos, Zhang, Stool, Husain, Thornton, Downing, McClintock and Lisanby14
Position of stimulation coil
We also tested whether stimulation at 100 Hz was capable of inducing seizures over the prefrontal cortex, which had been difficult to achieve at lower frequencies. Reference Lisanby, Schlaepfer, Fisch and Sackeim9,Reference Lisanby, Luber, Schlaepfer and Sackeim12,Reference Lisanby, Husain, Morales, Thornton, White, Payne, Rush and Sackheim13 We attempted seven prefrontal cortex stimulations in six patients. Of those, three were successful (one at 500 and two at 1000 pulses) and four were not successful (one at 500, one at 600, and two at 1000 pulses). Patients who did not fit with prefrontal stimulation fitted when stimulated over the vertex (Table 1). We conclude that even at the maximum setting of the machine, some patients will only fit if the coil is positioned over the vertex (i.e. closest to the motor cortex, which has a lower seizure threshold than the prefrontal or precentral cortices).
Seizure duration and recovery
We measured seizure duration during magnetic seizure therapy starting from the onset of stimulation. This is because we observed that the seizures in magnetic seizure therapy start during the stimulation train. In contrast, in ECT the convulsion typically does not start during electrical stimulation and a latent phase is usually seen immediately after stimulation. Reference Scott21
The mean duration of successful magnetic seizure therapy seizures was 31.3 s, range 10–86 s. Four patients had short seizures of 10, 18, 15 and 11 s (Table 1), which would not be considered therapeutic if evoked by ECT. Two of these patients were stimulated with only 600 pulses, raising the possibility that they may have had adequate seizures if stimulated at the maximum duration output (10 s) of the device.
In line with previous results, Reference Lisanby, Luber, Schlaepfer and Sackeim12 the recovery of orientation after magnetic seizure therapy was much faster than after ECT. Despite the small sample size, this difference was highly statistically significant and, more importantly, clinically meaningful. The ability to combine antidepressant efficacy with low neurocognitive adverse effects would be invaluable for patients who require neurostimulation therapies. Reference Sobin, Sackeim, Prudic, Devanand, Moody and McElhiney22 All patients felt less confused after magnetic seizure therapy. Many patients felt as if they had received no treatment and remembered details of what had happened immediately prior to the therapy. For instance, patients were able to continue conversations after recovery that had begun just prior to therapy.
EEG changes during seizures
It has been noted that the EEG after magnetic seizure therapy differs markedly from that after ECT, with a lower amplitude and relative absence of post-ictal suppression. Reference Lisanby, Luber, Schlaepfer and Sackeim12,Reference White, Amos, Zhang, Stool, Husain, Thornton, Downing, McClintock and Lisanby14 We confirmed these differences after stimulation at 100 Hz. Electroencephalogram traces during ECT showed high amplitude, synchronised EEG activity and clear post-ictal suppression which were markedly different from the EEG recorded after magnetic seizure therapy (traces available on request from the authors). The observed differences between ECT and magnetic seizure therapy ictal expression on EEG could be due to the more focal stimulation achieved with magnetic seizure therapy, which spares deeper brain regions such as the hippocampus that may be implicated in the cognitive side-effects of ECT. Reference Lisanby, Moscrip, Morales, Luber, Schroeder and Sackeim6 Differences in patterns of seizure expression might also explain the much faster recovery after magnetic seizure therapy. Another explanation for differences in ictal EEG expression between magnetic seizure therapy and ECT may stem from the fact that we were not recording EEG from directly under the magnetic coil, where the induced currents and seizure expression should be at its strongest. Specifically, our scalp EEG recordings were collected from bilateral prefrontal cortex, whereas the most effective coil placement was over the vertex. We have since observed that placing the electrodes over the motor cortex during magnetic seizure therapy produces clearer seizure activity, confirming our impression that these seizures are more localised (S. H. Lisanby, personal communication, 2008).
Outlook
Limitations of this work include the small sample size, open design and non-randomised nature. Nevertheless, this initial pilot study found that magnetic seizure therapy delivered with the new Magstim Theta device was well-tolerated and reliably produced seizures in the majority of patients, while resulting in much less post-ictal confusion. These encouraging initial results beg the question of the efficacy of this new investigational intervention for severe major depression. Previous open studies using 40 Hz and 50 Hz magnetic seizure therapy Reference Lisanby, Luber, Finck, Schroeder and Sackeim4,Reference White, Amos, Zhang, Stool, Husain, Thornton, Downing, McClintock and Lisanby14 showed promising results, although magnetic seizure therapy did not reach the effect size of optimal ECT. The ability to provide higher-dosage seizures relative to seizure threshold may narrow the gap in efficacy. This will be tested in the context of new trials now underway using the 100 Hz device to assess the effectiveness and safety of high-dose magnetic seizure therapy relative to ECT.
Acknowledgements
We thank the team from the Magstim Company (John H. Starzewski, Andrew Thomas, Anthony Thomas and Reza Jalinous) for constructing the new device and for always responding to our continuous requests for further refinements to the equipment. We thank the anaesthetists Mousa Saber Ali, John McClure, Charles Morton, John Wilson and Kate Harvey, and the nurses Maureen Giles, Karen Champney-Smith, John Tredget, Morag Gardner and Tracy Fraser who were involved in treating the first patients. These results were presented in part at the 2007 Annual Meeting of the American College of Neuropsychopharmacology, the 2007 Annual Meeting of the Royal College of Psychiatrists and the Medical Research Council Neurosciences Showcase Meeting in 2006.
Support for this work came from the Cardiff and Vale NHS Trust (to G.K., N.K. and M.A.) for the purchase of the Magstim Theta at Cardiff; and from a Trial Platform Grant of the UK Medical Research Council (G0401083) and the Gordon Small Charitable Trust (to K.P.E., R.E.OC and A.S.) for the magnetic seizure therapy trial in Edinburgh. The Magstim Company supported travel for S.H.L. and M.M.H. to attend the first magnetic seizure therapy treatments at Cardiff and Edinburgh. The development and preclinical testing of the prototype 100 Hz magnetic seizure therapy device were supported by a US National Institute of Health Grant (NIH R01 MH60884 to S.H.L.). S.H.L. and M.M.H. received a grant from the Stanley Medical Research Foundation for a randomised controlled trial of magnetic seizure therapy v. ECT. S.H.L. has also received grants to support magnetic seizure therapy development from National Alliance for Research on Schizophrenia and Depression, the American Federation for Aging Research, and New York State Foundation for Science, Technology and Innovation. For other work not the focus of this report, S.H.L. and M.M.H. have received funding from Neuronetics Inc. and Cyberonics Inc. Columbia University has submitted a patent on a novel transcranial magnetic stimulation technology developed in the laboratory of S.H.L. (not the topic of this report). None of the authors hold patents, office, or stock in magnetic seizure therapy or magnetic seizure therapy-related companies.
eLetters