Hostname: page-component-78c5997874-t5tsf Total loading time: 0 Render date: 2024-11-10T08:50:58.525Z Has data issue: false hasContentIssue false

Smart sensor tights: Movement tracking of the lower limbs in football

Published online by Cambridge University Press:  29 November 2021

Annemarijn Steijlen*
Affiliation:
Faculty of Electrical Engineering, Mathematics and Computer Science, Delft University of Technology, Delft, The Netherlands
Bastiaan Burgers
Affiliation:
Faculty of Electrical Engineering, Mathematics and Computer Science, Delft University of Technology, Delft, The Netherlands
Erik Wilmes
Affiliation:
Faculty of Behavioral and Movement Sciences, Vrije Universiteit Amsterdam, Amsterdam, The Netherlands
Jeroen Bastemeijer
Affiliation:
Faculty of Electrical Engineering, Mathematics and Computer Science, Delft University of Technology, Delft, The Netherlands
Bram Bastiaansen
Affiliation:
Centre for Human Movement Sciences, University Medical Centre Groningen & University of Groningen, Groningen, The Netherlands
Patrick French
Affiliation:
Faculty of Electrical Engineering, Mathematics and Computer Science, Delft University of Technology, Delft, The Netherlands
Andre Bossche
Affiliation:
Faculty of Electrical Engineering, Mathematics and Computer Science, Delft University of Technology, Delft, The Netherlands
Kaspar Jansen
Affiliation:
Faculty of Industrial Design Engineering, Delft University of Technology, Delft, The Netherlands
*
*Corresponding author. Email: a.s.m.steijlen@tudelft.nl

Abstract

This article presents a novel smart sensor garment with integrated miniaturized inertial measurements units (IMUs) that can be used to monitor lower body kinematics during daily training activities, without the need of extensive technical assistance throughout the measurements. The smart sensor tights enclose five ultra-light sensor modules that measure linear accelerations, angular velocities, and the earth magnetic field in three directions. The modules are located at the pelvis, thighs, and shanks. The garment enables continuous measurement in the field at high sample rates (250 Hz) and the sensors have a large measurement range (32 g, 4,000°/s). They are read out by a central processing unit through an SPI bus, and connected to a centralized battery in the waistband. A fully functioning prototype was built to perform validation studies in a lab setting and in a field setting. In the lab validation study, the IMU data (converted to limb orientation data) were compared with the kinematic data of an optoelectronic measurement system and good validity (CMCs >0.8) was shown. In the field tests, participants experienced the tights as comfortable to wear and they did not feel restricted in their movements. These results show the potential of using the smart sensor tights on a regular base to derive lower limb kinematics in the field.

Type
Research Article
Creative Commons
Creative Common License - CCCreative Common License - BY
This is an Open Access article, distributed under the terms of the Creative Commons Attribution licence (http://creativecommons.org/licenses/by/4.0), which permits unrestricted re-use, distribution and reproduction, provided the original article is properly cited.
Copyright
© The Author(s), 2021. Published by Cambridge University Press

Introduction

Over the years, the physical demands of football have increased. In the English Premier league for example, high intensity running distance and sprinting distance increased by more than 30% between 2006 and 2012 (Barnes et al., Reference Barnes, Archer, Hogg, Bush and Bradley2014). Likewise, in World Cup Soccer finals between 1966 and 2010, ball speed and passing rate increased by 15 and 35%, respectively (Wallace and Norton, Reference Wallace and Norton2014). The high physical demands in football lead to a substantial injury risk. In professional football 8.1 injuries per 1,000 hr of play were reported based on epidemiological data of 44 studies (López-Valenciano et al., Reference López-Valenciano, Ruiz-Pérez, Garcia-Gómez, Vera-Garcia, De Ste Croix, Myer and Ayala2020). Most injuries occur during competitive matches (Owoeye et al., Reference Owoeye, VanderWey and Pike2020), and almost one-third of all time-loss injuries are muscle related. More than 90% of all muscle injuries are lower limb injuries, of which 37% concern the hamstrings (Ekstrand et al., Reference Ekstrand, Hägglund and Waldén2011). Several studies conclude that increased age and previous injury significantly increase the risk of hamstring injury (Arnason et al., Reference Arnason, Sigurdsson, Gudmundsson, Holme, Engebretsen and Bahr2004; Freckleton and Pizzari, Reference Freckleton and Pizzari2013). Moreover, Ekstrand et al. (Reference Ekstrand, Hägglund and Waldén2011) reported more cases of injury at the end of each half of a football game, which suggests that fatigue is also an important risk factor. Therefore, investigating physical player load and its relation to hamstring injury prevalence can expand the knowledge on the etiology of hamstring injuries, and help develop new methods to prevent these injuries in the future.

The use of wearable electronic measurement equipment was approved by the FIFA in 2015 (Dunn et al., Reference Dunn, Hart and James2018). Several commercial systems that can track individual athletes by GPS (Global Positioning System, e.g., Zephyr Performance Systems, US, Catapult Sports Ltd., Australia and JOHAN Sports, The Netherlands) or RFID technology (Radio Frequency Identification, e.g., Inmotio Object Tracking, The Netherlands) have been developed. In addition, the systems contain a 3-axis accelerometer. Based on the data from these systems, the total distance covered can be calculated, whole-body acceleration and deceleration data can be derived (Barrett et al., Reference Barrett, Midgley and Lovell2014) and activities can be classified (Datson et al., Reference Datson, Drust, Weston, Jarman, Lisboa and Gregson2017). However, these data do not provide insight in the kinematics of the lower limbs. For detailed kinematic analysis of the lower limbs, the current practice is to use an optoelectronic measurement system (Cuesta-Vargas et al., Reference Cuesta-Vargas, Galán-Mercant and Williams2010; Schache et al., Reference Schache, Dorn, Blanch, Brown and Pandy2012; Malfait et al., Reference Malfait, Dingenen, Smeets, Staes, Pataky, Robinson, Vanrenterghem and Verschueren2016), which is restricted to a lab environment and does not allow for on-field measurements during training or competition. Kinematic measurements outside the lab are commonly performed using inertial measurement units (IMUs), which are attached to body segments. IMUs measure linear accelerations (accelerometer), angular velocities (gyroscope) and the earth magnetic field (magnetometer) in three different axes making the nine degrees of freedom (Ahmad et al., Reference Ahmad, Ghazilla, Khairi and Kasi2013). The orientation of an IMU can be derived by combining these measurements using a sensor fusion algorithm. After a calibration of the sensors to each limb segment, the lower limb kinematics can be obtained (Luinge et al., Reference Luinge, Veltink and Baten2007; Roetenberg et al., Reference Roetenberg, Slycke and Veltink2007).

Recently, several applications of IMU systems have been presented. These IMUs consist of a box, typically with a size of 30 mm × 40 mm × 10 mm, containing all electronics including the battery. In most cases, the modules are tied around the limbs, chest, or hands (Stiefmeier et al., Reference Stiefmeier, Roggen, Ogris, Lukowicz and Tröster2008; Rawashdeh et al., Reference Rawashdeh, Rafeldt and Uhl2016; Anwary et al., Reference Anwary, Yu and Vassallo2018; Chen et al., Reference Chen, Lin, Tsai, Chuang and Lee2020; Teague et al., Reference Teague, Heller, Nevius, Carek, Mabrouk, Garcia-Vicente, Inan and Etemadi2020). Other researchers attach them to the skin with an adhesive at predetermined locations (Gaidhani et al., Reference Gaidhani, Moon, Ozturk, Lee and Youm2017; Hu et al., Reference Hu, Ding, Peng, Liu and Wen2020). These sensor units are less suitable for long-term monitoring studies and little attention is paid to the integration of the IMUs in clothing in these papers. The company Xsens developed a motion tracking suit including software to derive body kinematics (MVN Link, Xsens, The Netherlands). Although, they found an elegant way to place the hardware in pockets in the garment, the electronics are relatively large, and it takes time and additional help to position all the electronics in the right locations when wearing the suit. The integration of electronics in textile is part of another research field that is more focused on materials science (Varga, Reference Varga, Schneegass and Amft2017; de Mulatier et al., Reference de Mulatier, Nasreldin, Delattre, Ramuz and Djenizian2018; Komolafe et al., Reference Komolafe, Torah, Wei, Nunes-Matos, Li, Hardy, Dias, Tudor and Beeby2019). In this field, some researchers have integrated IMUs in textiles. For example, Wicaksono et al. created a suit with an IMU on a flexible printed circuit board (PCB) right below the sternum for measuring heart, and breathing activity (Wicaksono et al., Reference Wicaksono, Tucker, Sun, Guerrero, Liu, Woo, Pence and Dagdeviren2020), and Wang et al. (Reference Wang, Chen, Timmermans, Karachristos, Martens and Markopoulos2015) created a garment with integrated IMUs for posture monitoring. However, in all of the above mentioned monitoring systems, either the sensor modules are relatively large and not integrated in the garment, or the IMU sampling rate and detection range are limited (100 Hz or below, and 16 g or lower, respectively). Therefore, as far as we are aware of, a garment for unobtrusive monitoring of lower limb kinematics in everyday training situations which is accurate enough and can be used without the need of technical assistance does not yet exist.

To enable reliable long-term monitoring at the football field, we are developing an easy-to-use wearable monitoring system with integrated IMUs, which will be referred to as smart sensor tights. This is an embedded system with IMUs integrated in textile on each segment of the lower limbs. The sensors are connected by flexible wiring to a central processing unit at the waist band that contains a single microprocessor and a power source. The advantages of this specific architecture are that the sensor modules are small, unobtrusive, and of low weight. Moreover, no additional synchronization of the sensors is needed, since they are all read out by one microprocessor. To the best of our knowledge, this is the first IMU system that has the sensors integrated in shorts or tights, which allows for easy on-field measurements of fast movements of the lower limbs. By making use of a real-time operating system, high sample rates can be reached to track fast football-specific movements accurately. The measurement range of the IMUs is larger (±4,000°/s for the gyroscopes and ±30 g for the accelerometers) than in the systems presented above. The tights can be easily put on by the players themselves and can be worn in regular matches and trainings. The integration of the sensors in a garment facilitates long-term monitoring of players and larger scale studies outside a lab environment, which are required to find injury risk factors for injury prevention.

This article presents the design and development of the smart sensor tights. First, the smart sensor tights system design is presented, followed by the detailed design of hardware and software. Secondly, a controlled test with football-specific movements was executed in the lab to concurrently validate the novel sensor tights with an optoelectronic measurement system. Third, a validation study with the complete system was executed. The sensor tights were worn during football-specific exercises to test the performance in the field. Lastly, user experience is assessed.

Materials and Methods

To improve the technical reliability and user acceptance of the new system, an iterative design approach was followed. Multiple prototypes were made to enable simultaneous improvement of the hardware and software, as well as improvement of usability and comfort. Prototype one included widely used IMUs (MPU-9250, InvenSense, San Jose, CA) and an easy-to-use microcontroller development board (Arduino Due). With this prototype, a proof of concept was created and based on field tests, major improvements were identified. These improvements included increasing the sampling rates, increasing the measurement range, and improving the robustness of the electronics. This formed the starting point for the design of the second prototype that is presented here.

Design

The design of the sensor tights had to meet several requirements. To investigate the kinematics of the legs, five IMUs are integrated in this design. IMUs are placed on the thighs, shanks, and pelvis. The exact locations are explained below. Since the product will be used by researchers in biomechanics to derive lower limb kinematics, it is important that the IMUs have a sample rate of 250 Hz in each direction and that raw data can be obtained via an SD card. At a later stage, a wireless connection will be integrated. Furthermore, the risk that the electronics harm the wearer or fellow players during a match should be negligible. The garment should not restrict the players movement and it should be comfortable to wear. Lastly, the IMUs need to be tight to the skin and of low mass, to prevent shaking due to inertia of the sensors.

Hardware design

For the hardware, it was chosen to centralize the data acquisition electronics and the power supply. This resulted in tiny and low weight sensor modules that reduced the measurement artifacts due to inertia of the sensors. Figure 1 shows an overview of the system. The two subsystems are referred to as the IMU node and the central unit. A new PCB was designed for each subsystem. Linear accelerations and angular velocities are measured by the ICM-20649 (InvenSense, San Jose, CA). This IMU was selected based on the wide measurement range of ±4,000°/s and ± 30 g, which enables more representative analysis of kinematics of highly dynamic movements. Preliminary experiments showed that with sensors with a smaller measurement range (16 g), clipping occurred at moments of impact with the ground. Via an I2C bus, a magnetometer (AK8963) with a range of ±4,900 μT (Asahi Kasei Microdevices, Japan) is connected to the ICM-20649. The central unit is responsible for reading out the sensor, power handling and data storage at an SD card. The microcontroller that was chosen for the central PCB is the Arm Cortex -M4 with FPU processor (Arm Ltd., UK) which runs at 100 MHz maximum. The STM32F411 Nucleo-64 development board, was used for software development (STMicroelectronics, Switzerland). Lastly, a CE marked power bank (1,350 mAh, Xqisit, Germany) was used as a power supply. Data acquisition from the sensor nodes occurs via an SPI connection. The SPI protocol was chosen, because the ICM-20649 can then directly be read out through SPI and no extra electronics are needed. Furthermore, the SPI protocol is less prone to errors over long distances at high data transfer rates than the I2C protocol.

Figure 1. System overview and location of the electronics.

Sensor locations

Two requirements needed to be considered to identify the best locations for the sensors. First, soft tissue artifacts (STAs) need to be minimized (Barré et al., Reference Barré, Jolles, Theumann and Aminian2015). STAs are defined as measurement errors that arise from the relative motion between bone and sensor, for example due to muscle bulging during a contraction. Second, we needed to choose a location that is safe and comfortable for use in a football context. To investigate the best sensor locations based on comfort and safety, football training sessions were observed from the Dutch national football team under 21 and a survey was held among the team (Steijlen et al., Reference Steijlen, Bastemeijer, Plaude, French, Bossche and Jansen2020). In consultation with human movement scientists and physiotherapists, locations that were expected to introduce the least STAs were identified. Based on the outcomes, the upper leg sensors are placed halfway the hip and knee joints on the lateral side of the thighs. STAs were expected to be relatively low because of the stiff underlying connective tissue. The lower leg sensors are placed at the inner sides of the shanks, close to the knees. The trunk sensor is placed in the waistband and the central unit, including the battery, is placed in a pocket located above the sacrum (Figure 1).

Software design

To make sure that each sensor node can be read out with a sample rate of 250 Hz for the gyroscopes and accelerometers and 100 Hz for the magnetometers, a real-time operating system (FreeRTOS) was used for this design. An RTOS was used to implement a pre-emptive priority-based scheduling scheme (Guan et al., Reference Guan, Peng, Perneel and Timmerman2016). Pre-emptive scheduling means that a task with lower priority can be interrupted by a task with higher priority. Once the task with a higher priority is finished, it will continue with the lower priority task. Table 1 shows the different tasks priority level. With this software, the required sample rates can easily be reached. When the microcontroller is running at 100 MHz, it turns out that the microcontroller only needs 40% of available time to execute all tasks. The power consumption is around 110 mAh, which means that with the current battery the system can run for 12 hr.

Table 1. Scheduling scheme for microprocessor

Lossless data compression

Future prototypes must be able to send data wirelessly to the football coach or medical staff. Given that the total amount of bits per sample is 16, the data rate of one prototype is 144 kbit/s. This implies that, when 22 players at the field are wearing the system, a data rate of 3.168 Mbit/s would be required. Based on the overview of different wireless protocols and their maximum bit rates and ranges by Kos et al. (Reference Kos, Milutinović and Umek2019), it can be concluded that only Wi-Fi protocols would be suitable. On top of that, experimental findings from literature show that the presence of a human body between the transmitter and receiver can lead to significant disturbances in the radiation pattern (Kurusingal et al., Reference Kurusingal, Dhamdhere and Sivaraman2010; Sivaraman et al., Reference Sivaraman, Grover, Kurusingal, Dhamdhere and Burdett2010) To improve the reliability of wireless data transmission for future versions, the amount of data retrieved from the sensors can be compressed. Therefore, the FELACS data compression algorithm (Kolo et al., Reference Kolo, Shanmugam, Lim and Ang2015) was implemented in the software. The performance of the algorithm was thoroughly tested in a football-specific setting.

Prototype

Figure 2 shows images of the tights and important details. As can be seen in Figure 2g an insulated stranded silver-plated copper conductor with a conductor area of 0.03 mm2 was chosen and laced in a serpentine pattern to allow for stretchability on top of a base garment (stretchable running tights, Under Armour, Dallas, TX). PTFE was chosen as insulation material, because of its high chemical inertness, hydrophobicity and mechanical strength, which will be of great value when the tights will absorb sweat and when they are washed. Miniature connectors for use in wearables are not yet commercialized. As an alternative, a regular miniature wire-to-board connector (Pico-clasp, Molex, Lisle, IL) was chosen to connect to the central unit PCB and IMU node PCBs. To test the impact of washing with soap and water on the reliability of the connections, a washing test (40°C, 1,200 RPM, including other sports clothes) was performed with these connectors (n = 12) and a 4-point resistance measurement was performed after each washing cycle. The resistance slightly increased after seven washing cycles. However, it stayed below 1 Ω. The PCBS were placed in 3D-printed casings made from a photopolymer (Connex 3, Objet 350, Stratasys Ltd., Israel). The sensor nodes weigh 3.6 g. The sensor nodes and central unit were placed in sleeves, which have a waterproof lining, to protect the electronics against sweat. For washing the tights, the sensors and central unit, can be disconnected. The wiring and connectors remain embedded in the garment. The central unit has an interface with three colored LEDs and two buttons, to start and stop recording and to indicate a special event. Figure 2d shows the interface.

Figure 2. (a) Side view sensor tights, (b) Back view sensor tights, (c) Pocket with central unit PCB and battery, (d) Battery and central unit with user interface, (e) Pocket of a sensor node, (f) A sensor node, (g) Interlaced wiring that allows for 100% stretching, and (h) Central unit during assembling.

Validation Study

The validation of the design was split in two parts. The first part was a validation test in the lab. The aim of this test was to assess the concurrent validity of the prototype with the golden standard optoelectronic motion analysis system. The second part was a validation test of the system on the football field. In this experiment, the functioning of the device was tested, and user experience tests were performed. It was verified if the datasets are complete, and the performance of the data compression algorithm for different football-specific movements was measured. The study was approved by the Human Research Ethics Committee of Delft University of Technology. All participants gave informed consent.

Lab validation

As explained in the introduction, detailed lower limb kinematics can be derived by applying sensor fusion algorithms to the IMU data and performing sensor-to-segment calibration (Roetenberg et al., Reference Roetenberg, Slycke and Veltink2007). Recently, a paper has been published which explains the model that we use to derive lower limb kinematics with the sensor tights (Wilmes et al., Reference Wilmes, de Ruiter, Bastiaansen, Zon, Vegter, Brink, Goedhart, KAPM and GJP2020). In that study, individual sensor modules that are taped to the skin were used to validate the models and good concurrent validity with an optoelectronic system was shown. To validate the novel sensor tights, a similar protocol was performed. Since the previous study was already performed with 11 participants, it was chosen to compare the previous results and the results with the new garment of 1 participant to validate the new product. In the current experiment, one male participant (age: 23 years, height: 1.89 m) was wearing the garment with integrated IMUs, and the measurements of the garment were compared with the optoelectronic system measurements. The sensor-to-segment calibration consists of two steps. First, the participant is asked to stand still for 5 s in a neutral upright pose to identify the longitudinal axis of each segment by using the direction of gravity. Thereafter, the participant performs three movements in the sagittal plane to determine the frontal axis of each segment; a rise of the right upper leg, a rise of the left upper leg, and a bow forward of the trunk. A gradient descent Madgwick algorithm is used to estimate the sensor orientation during the measurements (Madgwick et al., Reference Madgwick, Harrison and Vaidyanathan2011).

The optoelectronic system used eight cameras (Vicon V5 cameras, Vicon Motion Systems Ltd., UK), and 20 reflective markers were placed at the lower body. More information about marker placement can be found in Appendix A. The motion capture area was 25 m2. Five types of football-specific movement were executed: an acceleration run, a run with a cutting movement, a run with a 180° turn, kicking a ball (preceded by a few steps) and a jump. Each movement was performed at three different intensities (50, 80, and 100% of maximum effort respectively). Each trial was repeated three times. A more detailed description of the protocol can be found in Wilmes et al. (Reference Wilmes, de Ruiter, Bastiaansen, Zon, Vegter, Brink, Goedhart, KAPM and GJP2020). Data were processed in MATLAB (The MathWorks, Santa Clara, CA). Hip and knee joint angles and angular velocities were calculated with the data from the optoelectronic system, and with the data from the sensor tights. Thereafter, the root mean square differences (RMSDs) and coefficients of multiple correlation (CMCs) were calculated between the sensor tights and the optoelectronic system for all types of movements separately.

Field tests

Five male participants (recreational football players, age: 21.8 ± 1.3 years) were asked to wear the garment and to perform a high intensity football-specific training drill (Kelly et al., Reference Kelly, Gregson, Reilly and Drust2013) This training drill is aimed to replicate the physical movements and technical actions during match-play. The drill was performed at least two times for each participant. Next to the training drill, two participants performed an extra series of isolated football-specific movements at 50, 80, and 100% of maximum effort. These movements included an instep soccer kick, running, jumping vertically, running with the ball, and running sideways. After the field experiment, all participants were asked to fill in a short questionnaire about their experience with wearing the tights. Data were processed in MATLAB (The MathWorks, Santa Clara, CA). Differentiated signals and probability distributions of the decompressed accelerometer, gyroscope, and magnetometer data were calculated to check the completeness of the dataset, and functioning of the sensors. Furthermore, compression ratios for each type of movement were calculated, to test the performance of the data compression algorithm. The compression ratio (CR) is defined as:

(1) $$ \mathrm{CR}=100\times \left(1-\frac{\mathrm{compressed}\ \mathrm{filesize}}{\mathrm{original}\ \mathrm{filesize}}\;\right)\%.\hskip0.84em $$

Results and Discussion

This section describes the results and discussion of the validation study. First, a comparison of football-specific movements measured by the optoelectronic system and by the prototype is made. Secondly, the results of the recorded training sessions and user tests are discussed.

Lab Validation

One participant performed six types of football-specific movements, and these were tracked with the IMU system and the optoelectronic measurement system at the same time. To compare the IMU data with the optoelectronic measurement results, the raw IMU data were converted to joint angles and joint angular velocities. An example of the comparison of the joint angles and angular velocities of the right leg during a kick and a jump measured with the prototype and the optoelectronic system, is shown in Figures 3 and 4, respectively.

Figure 3. Joint angles and angular velocities of the right leg during a kick. The vertical gray line indicates the moment of ball contact.

Figure 4. Joint angles and angular velocities of the right leg during a jump. The first vertical gray line indicates the time of push-off and the second indicates the landing.

To evaluate the comparison, the RMSDs and coefficients of multiple correlation (CMCs) were calculated for each joint respectively. An overview of all RMSDs and CMCs at the different intensities and types of movement, is given in Appendix B. As shown in Table B1, RMSDs of the knee joint angles for all types of actions (each movement at a certain intensity) ranged between 14.11° and 2.98°, and CMCs between 0.911 and 0.997. For hip joint angles for all types of movement, RMSDs ranged between 18.46° and 7.07° and CMCs between 0.830 and 0.983. Errors in hip joint angles are likely to be larger than errors in knee joint angles, because the hip has a larger range of motion. RMSDs of joint angular velocities for all separated actions ranged between 168.2°/s and 59.1°/s for the hips and between 141.2°/s and 37.7°/s for the knees, respectively. CMCs ranged between 0.809 and 0.968 for the hip and between 0.979 and 0.9938 for the knee joint angular velocities.

In general, it was found that the RMSDs and CMCs for all types of movement are within the same range as in the previous measurements of 11 players with the taped IMUs (Wilmes et al., Reference Wilmes, de Ruiter, Bastiaansen, Zon, Vegter, Brink, Goedhart, KAPM and GJP2020). In some cases (e.g., Figure 4, upper right) it seems that the differences are related to stronger filtering (peak suppression) by the Vicon system, rather than measurement accuracy. Furthermore, the RMSDs and CMCs of joint angles found in the previous study by Wilmes et al. and in this study were comparable to the results from other studies. For example, Nüesch et al. (Reference Nüesch, Roos, Pagenstert and Mündermann2017) measured RMSDs of 27.6° for the hip and 17.9° for the knee during jogging (~2.9 m/s), before they performed an offset correction. Additionally, Tadano et al. (Reference Tadano, Takeda and Miyagawa2013) measured RMSDs of 9.0° for the hip and 7.1° for the knee during walking. Errors are expected to be higher for high intensity movements due to STAs and errors originating from the orientation filter. During this study, the participants performed higher intensity movements than reported in the other studies (e.g., acceleration runs with a mean running speed up to ~6.6 m/s), and the new IMU system still provides valid joint angle measurements (CMCs > 0.8) during these higher intensity movements. The advantage of the garment compared to the taped IMUs is that it is easier to use, comfortable to wear, and allows for longer-term monitoring studies during trainings and matches in the field.

Field tests

This section describes the results and discussion of the field tests. First, the technical validation results are discussed. Second, user experience insights are shown.

Technical validation

Datasets of the training sessions of the five participants were checked for completeness by plotting the probability distributions of the differentiated signals and visually inspecting the sensor data. It was found that for one participant, malfunctioning of a connector resulted in data loss of one of the lower leg sensors. For the other measurements, the differentiated signals of the sensors showed a Gaussian distribution. Figure 5a shows typical accelerometer data of the left lower leg in a single direction for the accelerometer. They were recorded during execution of a series of isolated movements. The figure shows results at different running intensities.

Figure 5. (a) Results from the left lower leg accelerometer (x-axis) of participant 1, during walking (1), jogging (2), running (3), sprinting (4), and running out (5). The spacings between each movement type indicate that the player is standing still or turning. (b) A recording of a shot. Results from the right lower leg accelerometer and gyroscope in a single direction. (c) Power spectrum analysis of accelerometer recordings (left lower leg, x-axis) at different running intensities (walking, jogging, running, and sprinting). For each running intensity, a recording of 1,000 samples (4 s) is used. A zoomed-in view of the range between 80 and 125 Hz is shown at the top right. The training session measurements showed that the accelerometers and gyroscopes worked properly and data transfer using the SPI protocol through the laced wiring appeared to be no problem. However, the magnetometers occasionally showed double readings. Most likely, this is caused by a not-updated register of the sensor itself. In future versions, the magnetometer can be replaced with a newer version, which may improve reliability of the readings.

Detailed analysis showed that the compression ratios varied with movement intensity as expected. Lower compression ratios were reached at higher movement intensities, whereas the lowest compression ratios were observed while sprinting (~24%). During walking the compression ratios were around 48%. As a comparison, Chiasson et al. (Reference Chiasson, Xu and Shull2020) were able to achieve compression ratios of ~16% of IMU data during walking in a recent study. It should be noted that their data were converted using the CR definition presented in the methods section, because they used another CR definition. The sample frequency of their dataset was 60 Hz, with a lower measurement range, which probably explains the large difference with our data. On top of that, the encoding part of the FELACS algorithm improves the representation of the quotient, most likely outperforming normal Golomb-rice encoding used by Chiasson et al. In this study, the average compression ratios of all sensors during the training session were between 43–45% for the different participants. In other words, the files of the training session data were 43–45% smaller than the original data set, which is beneficial for wireless data transfer in future prototypes. The algorithm for data compression consumes only 1.6% of the total power consumption extra power. These results are promising for wireless data transfer because the algorithm would markedly reduce the amount of data to be transferred at only a very small energetic cost.

At the start of this research, it was assumed that high sample rates (of 250 Hz) are required to prevent aliasing, and to ensure that the resolution is high enough to track all football-specific movements accurately. First, when inspecting the raw data of the recordings of the isolated movements, sudden large changes were observed during high intensity movements such as sprinting and kicking. Figure 5b shows the data of the gyroscope and accelerometer of the right lower leg during a shot. The recording comprises 25 samples. Due to the limited number of samples, no transient behavior was recorded during impact with the ball. The accelerometer measured a difference of almost 15 g within 12 ms. This indicates that maybe even higher sampling rates than 250 Hz could be beneficial for high intensity actions like shooting. In lower intensity movements walking, transient information and even noise were measured during a stride, so the sample rate is high enough for this type of movement.

Secondly, to analyse if the high sampling rate of 250 Hz is sufficient, a fast Fourier transform (FFT) was used to represent the data at different running intensities in the frequency domain. A total of 1,000 samples of each running intensity were used for a separate power spectrum analysis. The results are plotted in Figure 5c. The first peak in the power spectrum corresponds to the dominant movement frequency of the different activities. As can be seen from Figure 5c, for walking, jogging, running, and sprinting, the dominant frequencies increase from 2.0, 2.5, 2.75 to 3.75 Hz respectively. When zooming in at the last part of the graph, it can be seen that between 80 and 125 Hz the amplitude is higher for sprinting than for walking. A shock usually consists of a high magnitude main peak, associated with lower magnitude components at the higher frequency part of the spectrum. The high frequency components observed in Figure 5c are therefore attributed to these associated lower magnitude components. From the power spectrum analysis (Figure 5c) in combination with the abrupt changes in accelerometer and gyroscope data during shooting (Figure 5b) and sprinting, it can be concluded that this high sample rate adds information in case of high intensity movements. Further research is needed to find out if this extra information adds accuracy in deriving kinematics and estimating player load. Furthermore, a balance between high sample rates and practical implementation needs to be found. High sampling rates will increase the load on a wireless link in the future and it will increase energy use. Based on these considerations, a sampling frequency of 250 Hz is a justified choice. Future work may also include the use of adaptive sampling. For sprinting and shooting, the sample rates maybe even higher than 250 Hz, while the sample rates could be reduced for walking without losing accuracy.

User experience

To investigate user experience, each participant filled in a short questionnaire. Table 2 summarizes the results. Testing the prototype in the field provided insight not only in user experiences, but also in reliability of the prototype and efficiency and effectiveness of the movement monitoring system. User experience scores on comfort and freedom of movement were relatively high. For both attributes a rating of 4 was given on average (Likert 5-point scale). A tight fit of the tights is required for minimizing motion artifacts of the sensors. Four subjects experienced the tightness of the garment as a positive aspect. One subject, with the largest clothing size, experienced the garment as too tight. In the future, multiple sizes and shorts need to be developed to enhance comfort and freedom of movement for every player. In general, the subjects did not feel restricted in their movements by the electronics in the tights. However, one subject noted that the garment would seem too fragile for sliding’s. In future versions, connections will be made more robust to enable the players to make sliding’s. Furthermore, the thickness of the sensor modules will be reduced, and the use of a cushioning layer can be explored.

Table 2. Results of the user experience tests. Positive feedback is indicated with a (+) and negative feedback with a (−)

When a novel prototype is developed and a wireless module is added to the tights, it would be beneficial when the athletes can activate the sensor tights themselves. Therefore, future work will include optimization of the user interface to start and stop measurements for use by athletes. Future work should also include better integration and miniaturization of the sensor nodes and central unit to enhance safe use and comfort. Furthermore, the connections with the sensor nodes need to be improved. During the tests, it was found that even after very short instants of disconnection, for example during dressing, the data collection is interrupted for this sensor. Since the connectors chosen were not designed for use in clothing, alternative solutions need to be explored. A very robust, yet not very sustainable solution, is to solder the wires to the PCB and encapsulate sensor nodes with an acrylate or silicone material. An alternative solution would be to develop a new type of connection, for example based on inductive coupling. In the current version of the garment, a waterproof textile layer was placed in the pockets between the electronics and the skin to protect against sweat. When the sensor nodes are encapsulated, the central unit casing is sealed, and an extra waterproof layer will be placed between the environment and the body, the system will also be resistant to rain. Lastly, although power consumption did not cause feasibility issues, it can be optimized to minimize the size of the battery. A smaller battery can enhance comfort and reduce safety issues. Power consumption can be reduced by optimizing the software and reducing power dissipation through the wiring.

Conclusions

In this study, a fully functional prototype of a garment was developed for monitoring lower limb kinematics during every day, on-the-field training situations. Attention was paid to the aspects of unobtrusiveness, easiness of use, as well as the accuracy and reliability of the recorded signals. In a human performance lab, hip and knee joint angles and angular velocities recorded with the tights were compared with the same measurements obtained with an optoelectronic measurement system and good validity was shown. This demonstrates that the sensor tights can be used to accurately monitor lower limb kinematics. During the field tests, the tights scored high on ease-of-use, comfort, and freedom of movement, which allows the sensor garment to be used in field training sessions and matches, and this shows that the tights can be used for long term monitoring studies. In follow-up studies, it will be researched how the lower limb kinematics, measured by the garment, can be used to estimate individual player load on the lower limbs. Future work includes making a wireless connection to a dashboard for the coach and medical staff. This dashboard should show physical player load parameters. The ultimate goal is to relate these player load parameters to the prevalence of hamstring injuries, so that the tights can be used to prevent hamstring injuries in the future.

Acknowledgments

The authors would like to thank Linda Plaude for textile design and prototyping support. Furthermore, the authors would like to thank all participants in this study for their time and effort.

Funding Statement

This work is part of the research program Citius, Altius, and Sanius with project number P16-28, which is financed by the Dutch Research Council (NWO).

Competing Interests

The authors declare no competing interests exist.

Authorship Contributions

Conceptualization: A.S., E.W., B. Ba., K.J. Design and prototyping: A.S., B.Bu., J.B. Field testing: A.S., B.Bu. Lab validation: E.W., A.S., B.Bu. Writing first draft: A.S., K.J. Reviewing/writing, supervision, and funding acquisition: P.F., A.B., K.J.

Data Availability Statement

The datasets used in this study are available from the corresponding author upon request.

Ethical Standards

The validation measurements were approved by the Human Ethics Research Committee of Delft University of Technology (ID1237). The authors assert that all procedures contributing to this work comply with the ethical standards of the relevant national and institutional committees on human experimentation and with the Helsinki Declaration of 1975, as revised in 2008.

A. Appendix A

Figure A1. The setup for the validation experiment from posterior (left) and anterior (right) view: the IMUs are highlighted in blue. The Vicon markers (white dots) are placed at the following anatomical locations: At the posterior and anterior superior iliac spines, Halfway the length of hip and knee on the lateral and posterior sides of the thighs, at the medial and lateral femoral epicondyles, halfway the length of knee and ankle on the lateral and anterior sides of the shanks, and at the medial and lateral malleoli.

B. Appendix B

B.1. Results of the Lab Validation Study

Table B1. Root mean square differences between the optoelectronic measurement system and the prototype for the different movement types at different intensity levels

Table B2. Coefficients of multiple correlation between the optoelectronic measurement system and the prototype for the different movement types at different intensity levels

References

Ahmad, N, Ghazilla, RAR, Khairi, NM and Kasi, V (2013) Reviews on various inertial measurement unit (IMU) sensor applications. International Journal of Signal Processing Systems 1(2), 256262.CrossRefGoogle Scholar
Anwary, AR, Yu, H and Vassallo, M (2018) Optimal foot location for placing wearable IMU sensors and automatic feature extraction for gait analysis. IEEE Sensors Journal 18(6), 25552567. https://doi.org/10.1109/JSEN.2017.2786587CrossRefGoogle Scholar
Arnason, A, Sigurdsson, SB, Gudmundsson, A, Holme, I, Engebretsen, L and Bahr, R (2004) Risk factors for injuries in football. The American Journal of Sports Medicine 32(1_suppl), 516. https://doi.org/10.1177/0363546503258912CrossRefGoogle ScholarPubMed
Barnes, C, Archer, DT, Hogg, B, Bush, M and Bradley, PS (2014) The evolution of physical and technical performance parameters in the English premier league. International Journal of Sports Medicine 35(13), 10951100. https://doi.org/10.1055/s-0034-1375695Google ScholarPubMed
Barré, A, Jolles, BM, Theumann, N and Aminian, K (2015) Soft tissue artifact distribution on lower limbs during treadmill gait: Influence of skin markers’ location on cluster design. Journal of Biomechanics 48(10), 19651971. https://doi.org/10.1016/j.jbiomech.2015.04.007CrossRefGoogle ScholarPubMed
Barrett, S, Midgley, A and Lovell, R (2014) PlayerLoad™: Reliability, convergent validity, and influence of unit position during treadmill running. International Journal of Sports Physiology and Performance 9(6), 945. https://doi.org/10.1123/ijspp.2013-0418CrossRefGoogle ScholarPubMed
Chen, Y-P, Lin, C-Y, Tsai, M Jr, Chuang, T-Y and Lee, OK-S (2020) Wearable motion sensor device to facilitate rehabilitation in patients with shoulder adhesive capsulitis: Pilot study to assess feasibility. Journal of Medical Internet Research 22(7), e17032. https://doi.org/10.2196/17032CrossRefGoogle ScholarPubMed
Chiasson, D, Xu, J and Shull, P (2020) Lossless compression of human movement IMU signals. Sensors 20(20), 5926.CrossRefGoogle ScholarPubMed
Cuesta-Vargas, AI, Galán-Mercant, A and Williams, JM (2010) The use of inertial sensors system for human motion analysis. Physical Therapy Reviews 15(6), 462473. https://doi.org/10.1179/1743288X11Y.0000000006CrossRefGoogle ScholarPubMed
Datson, N, Drust, B, Weston, M, Jarman, IH, Lisboa, PJ and Gregson, W (2017) Match physical performance of elite female soccer players during international competition. The Journal of Strength & Conditioning Research 31(9), 23792387. https://doi.org/10.1519/jsc.0000000000001575CrossRefGoogle ScholarPubMed
de Mulatier, S, Nasreldin, M, Delattre, R, Ramuz, M and Djenizian, T (2018) Electronic circuits integration in textiles for data processing in wearable technologies. Advanced Materials Technologies 3(10), 1700320. https://doi.org/10.1002/admt.201700320CrossRefGoogle Scholar
Dunn, M, Hart, J and James, D (2018) Wearing electronic performance and tracking system devices in association football: Potential injury scenarios and associated impact energies. Multidisciplinary Digital Publishing Institute Proceedings 2(6), 232.Google Scholar
Ekstrand, J, Hägglund, M and Waldén, M (2011) Epidemiology of muscle injuries in professional football (soccer). The American Journal of Sports Medicine 39(6), 12261232. https://doi.org/10.1177/0363546510395879CrossRefGoogle Scholar
Freckleton, G and Pizzari, T (2013) Risk factors for hamstring muscle strain injury in sport: A systematic review and meta-analysis. British Journal of Sports Medicine 47(6), 351358. https://doi.org/10.1136/bjsports-2011-090664CrossRefGoogle ScholarPubMed
Gaidhani, A, Moon, KS, Ozturk, Y, Lee, SQ and Youm, W (2017) Extraction and analysis of respiratory motion using wearable inertial sensor system during trunk motion. Sensors 17(12), 2932.CrossRefGoogle ScholarPubMed
Guan, F, Peng, L, Perneel, L and Timmerman, M (2016) Open source FreeRTOS as a case study in real-time operating system evolution. Journal of Systems and Software 118, 1935. https://doi.org/10.1016/j.jss.2016.04.063CrossRefGoogle Scholar
Hu, B, Ding, T, Peng, Y, Liu, L and Wen, X (2020) Flexible and attachable inertial measurement unit (IMU)-based motion capture instrumentation for the characterization of hand kinematics: A pilot study. Instrumentation Science & Technology 49, 125145. https://doi.org/10.1080/10739149.2020.1789657CrossRefGoogle Scholar
Kelly, DM, Gregson, W, Reilly, T and Drust, B (2013) The development of a soccer-specific training drill for elite-level players. The Journal of Strength & Conditioning Research 27(4), 938943. https://doi.org/10.1519/JSC.0b013e3182610b7dCrossRefGoogle ScholarPubMed
Kolo, JG, Shanmugam, SA, Lim, DWG and Ang, L-M (2015) Fast and efficient lossless adaptive compression scheme for wireless sensor networks. Computers & Electrical Engineering 41, 275287. https://doi.org/10.1016/j.compeleceng.2014.06.008CrossRefGoogle Scholar
Komolafe, A, Torah, R, Wei, Y, Nunes-Matos, H, Li, M, Hardy, D, Dias, T, Tudor, M and Beeby, S (2019) Integrating flexible filament circuits for E-textile applications. Advanced Materials Technologies 4(7), 1900176. https://doi.org/10.1002/admt.201900176CrossRefGoogle Scholar
Kos, A, Milutinović, V and Umek, A (2019) Challenges in wireless communication for connected sensors and wearable devices used in sport biofeedback applications. Future Generation Computer Systems 92, 582592. https://doi.org/10.1016/j.future.2018.03.032CrossRefGoogle Scholar
Kurusingal, A, Dhamdhere, A and Sivaraman, V (2010) Modeling signal strength of body-worn devices. Paper Presented at the IEEE Local Computer Network Conference, 10–14 October 2010.CrossRefGoogle Scholar
López-Valenciano, A, Ruiz-Pérez, I, Garcia-Gómez, A, Vera-Garcia, FJ, De Ste Croix, M, Myer, GD and Ayala, F (2020) Epidemiology of injuries in professional football: A systematic review and meta-analysis. British Journal of Sports Medicine 54(12), 711. https://doi.org/10.1136/bjsports-2018-099577CrossRefGoogle ScholarPubMed
Luinge, HJ, Veltink, PH and Baten, CTM (2007) Ambulatory measurement of arm orientation. Journal of Biomechanics 40(1), 7885. https://doi.org/10.1016/j.jbiomech.2005.11.011CrossRefGoogle ScholarPubMed
Madgwick, SOH, Harrison, AJL and Vaidyanathan, R (2011) Estimation of IMU and MARG orientation using a gradient descent algorithm. Paper Presented at the 2011 IEEE International Conference on Rehabilitation Robotics, 29 June–1 July 2011.CrossRefGoogle Scholar
Malfait, B, Dingenen, B, Smeets, A, Staes, F, Pataky, T, Robinson, MA, Vanrenterghem, J and Verschueren, S (2016) Knee and hip joint kinematics predict quadriceps and hamstrings neuromuscular activation patterns in drop jump landings. PLoS One 11(4), e0153737.CrossRefGoogle ScholarPubMed
Nüesch, C, Roos, E, Pagenstert, G and Mündermann, A (2017) Measuring joint kinematics of treadmill walking and running: Comparison between an inertial sensor based system and a camera-based system. Journal of Biomechanics 57, 3238. https://doi.org/10.1016/j.jbiomech.2017.03.015CrossRefGoogle Scholar
Owoeye, OBA, VanderWey, MJ and Pike, I (2020) Reducing injuries in soccer (football): An umbrella review of best evidence across the epidemiological framework for prevention. Sports Medicine - Open 6(1), 46. https://doi.org/10.1186/s40798-020-00274-7CrossRefGoogle ScholarPubMed
Rawashdeh, SA, Rafeldt, DA and Uhl, TL (2016) Wearable IMU for shoulder injury prevention in overhead sports. Sensors 16(11), 1847.CrossRefGoogle ScholarPubMed
Roetenberg, D, Slycke, PJ and Veltink, PH (2007) Ambulatory position and orientation tracking fusing magnetic and inertial sensing. IEEE Transactions on Biomedical Engineering 54(5), 883890. https://doi.org/10.1109/TBME.2006.889184CrossRefGoogle ScholarPubMed
Schache, AG, Dorn, TW, Blanch, PD, Brown, NA and Pandy, MG (2012) Mechanics of the human hamstring muscles during sprinting. Medicine & Science in Sports & Exercise 44(4), 647658.CrossRefGoogle ScholarPubMed
Sivaraman, V, Grover, S, Kurusingal, A, Dhamdhere, A and Burdett, A (2010) Experimental study of mobility in the soccer field with application to real-time athlete monitoring. Paper Presented at the 2010 IEEE 6th International Conference on Wireless and Mobile Computing, Networking and Communications, 11–13 October 2010.CrossRefGoogle Scholar
Steijlen, ASM, Bastemeijer, J, Plaude, L, French, PJ, Bossche, A and Jansen, KMB (2020) Development of sensor tights with integrated inertial measurement units for injury prevention in football. Paper presented at the Proceedings of the 6th International Conference on Design4Health, Amsterdam.Google Scholar
Stiefmeier, T, Roggen, D, Ogris, G, Lukowicz, P and Tröster, G (2008) Wearable activity tracking in car manufacturing. IEEE Pervasive Computing 7(2), 4250. https://doi.org/10.1109/MPRV.2008.40CrossRefGoogle Scholar
Tadano, S, Takeda, R and Miyagawa, H (2013) Three dimensional gait analysis using wearable acceleration and gyro sensors based on quaternion calculations. Sensors 13(7), 93219343. https://doi.org/10.3390/s130709321CrossRefGoogle ScholarPubMed
Teague, CN, Heller, JA, Nevius, BN, Carek, AM, Mabrouk, S, Garcia-Vicente, F, Inan, OT and Etemadi, M (2020) A wearable, multimodal sensing system to monitor knee joint health. IEEE Sensors Journal 20(18), 1032310334. https://doi.org/10.1109/JSEN.2020.2994552CrossRefGoogle Scholar
Varga, M (2017) Electronics integration. In Schneegass, S and Amft, O (eds), Smart Textiles: Fundamentals, Design, and Interaction. Cham: Springer International Publishing, pp. 161184.CrossRefGoogle Scholar
Wallace, JL and Norton, KI (2014) Evolution of world cup soccer final games 1966–2010: Game structure, speed and play patterns. Journal of Science and Medicine in Sport 17(2), 223228. https://doi.org/10.1016/j.jsams.2013.03.016CrossRefGoogle ScholarPubMed
Wang, Q, Chen, W, Timmermans, AAA, Karachristos, C, Martens, JB and Markopoulos, P (2015) Smart rehabilitation garment for posture monitoring. Paper Presented at the 2015 37th Annual International Conference of the IEEE Engineering in Medicine and Biology Society (EMBC), 25–29 August 2015.CrossRefGoogle Scholar
Wicaksono, I, Tucker, CI, Sun, T, Guerrero, CA, Liu, C, Woo, WM, Pence, EJ and Dagdeviren, C (2020) A tailored, electronic textile conformable suit for large-scale spatiotemporal physiological sensing in vivo. NPJ Flexible Electronics 4(1), 5. https://doi.org/10.1038/s41528-020-0068-yCrossRefGoogle Scholar
Wilmes, E, de Ruiter, CJ, Bastiaansen, BJC, Zon, JFJA v, Vegter, RJK, Brink, MS, Goedhart, EA, KAPM, Lemmink and GJP, Savelsbergh (2020) Inertial sensor-based motion tracking in football with movement intensity quantification. Sensors 20(9), 2527.CrossRefGoogle ScholarPubMed
Figure 0

Figure 1. System overview and location of the electronics.

Figure 1

Table 1. Scheduling scheme for microprocessor

Figure 2

Figure 2. (a) Side view sensor tights, (b) Back view sensor tights, (c) Pocket with central unit PCB and battery, (d) Battery and central unit with user interface, (e) Pocket of a sensor node, (f) A sensor node, (g) Interlaced wiring that allows for 100% stretching, and (h) Central unit during assembling.

Figure 3

Figure 3. Joint angles and angular velocities of the right leg during a kick. The vertical gray line indicates the moment of ball contact.

Figure 4

Figure 4. Joint angles and angular velocities of the right leg during a jump. The first vertical gray line indicates the time of push-off and the second indicates the landing.

Figure 5

Figure 5. (a) Results from the left lower leg accelerometer (x-axis) of participant 1, during walking (1), jogging (2), running (3), sprinting (4), and running out (5). The spacings between each movement type indicate that the player is standing still or turning. (b) A recording of a shot. Results from the right lower leg accelerometer and gyroscope in a single direction. (c) Power spectrum analysis of accelerometer recordings (left lower leg, x-axis) at different running intensities (walking, jogging, running, and sprinting). For each running intensity, a recording of 1,000 samples (4 s) is used. A zoomed-in view of the range between 80 and 125 Hz is shown at the top right. The training session measurements showed that the accelerometers and gyroscopes worked properly and data transfer using the SPI protocol through the laced wiring appeared to be no problem. However, the magnetometers occasionally showed double readings. Most likely, this is caused by a not-updated register of the sensor itself. In future versions, the magnetometer can be replaced with a newer version, which may improve reliability of the readings.

Figure 6

Table 2. Results of the user experience tests. Positive feedback is indicated with a (+) and negative feedback with a (−)

Figure 7

Figure A1. The setup for the validation experiment from posterior (left) and anterior (right) view: the IMUs are highlighted in blue. The Vicon markers (white dots) are placed at the following anatomical locations: At the posterior and anterior superior iliac spines, Halfway the length of hip and knee on the lateral and posterior sides of the thighs, at the medial and lateral femoral epicondyles, halfway the length of knee and ankle on the lateral and anterior sides of the shanks, and at the medial and lateral malleoli.

Figure 8

Table B1. Root mean square differences between the optoelectronic measurement system and the prototype for the different movement types at different intensity levels

Figure 9

Table B2. Coefficients of multiple correlation between the optoelectronic measurement system and the prototype for the different movement types at different intensity levels