Estimation of body composition is a vital element of nutritional assessment as fat and fat-free compartments of body mass have different health implications. Fat mass (FM) is closely linked with metabolic complications of obesity because the adipose tissue functions as an endocrine organ that releases bioactive substances having pro-inflammatory properties( Reference Ouchi, Parker and Lugus 1 ). In contrast, fat-free mass (FFM), especially muscle mass, plays a protective role against the risk of chronic diseases including diabetes and osteoporosis( Reference Wolfe 2 ). Ethnic differences in the relationship between BMI and disease risk have been associated with differences in body composition( Reference Dulloo, Jacquet and Solinas 3 , Reference Nightingale, Rudnicka and Owen 4 ).
A number of techniques are available for the assessment of body composition, and the choice of technique usually depends on precision, accuracy, ease of application as well as the cost. DXA is increasingly used for body composition assessment because of its high precision and low dose of radiation. A number of studies have validated other, less precise, techniques such as anthropometry and bioelectrical impedance analysis against DXA as a reference method( Reference Loveday, Thompson and Mitchell 5 – Reference Bousbiat, Jaffrin and Dongmo 7 ). However, DXA is not without limitations. Although studies have shown that estimates of body composition by DXA are highly correlated with those derived using more accurate methods, variations have been reported between the estimates( Reference Visser, Fuerst and Lang 8 , Reference Schoeller, Tylavsky and Baer 9 ).
With increasing recognition of the association between the high prevalence of the metabolic syndrome and ‘thin-fat’ phenotype in Indians, there is enhanced interest in the assessment of body composition( Reference Yajnik, Fall and Coyaji 10 , Reference Misra, Vikram and Arya 11 ). A number of studies in India have reported the body composition of different population groups using different techniques including DXA( Reference Bhat, Yajnik and Sayyad 12 – Reference Kulkarni, Shatrugna and Nagalla 15 ). However, different studies that have compared the estimates of body composition using different methods of body composition measurement need to consider the variation in estimates associated with these methods. Moreover, studies comparing different methods of body composition measurement tend to be population-specific due to ethnic variations in body composition( Reference Haroun, Taylor and Viner 16 ). Studies comparing the estimates of body composition using DXA with those measured by other reference methods have not so far been reported in India. Therefore, the aim of the present study was to compare the estimates of body composition by DXA with those using the isotope dilution technique.
Participants and methods
The present study was conducted according to the guidelines laid down in the Declaration of Helsinki, and all procedures involving human participants were approved by the ethics committees of the National Institute of Nutrition, Hyderabad, India, the London School of Hygiene & Tropical Medicine, UK and Queensland University of Technology, Australia. Written informed consent was obtained from all participants.
Participants
Healthy volunteers aged 19–70 years were enrolled in the present study from two pre-established cohorts (Andhra Pradesh Children and Parents Study (APCAPS), n 58 and Indian Migration Study (IMS), n 94) living around the city of Hyderabad, India. The APCAPS cohort was established to assess the long-term impact of early nutrition supplementation provided to pregnant women and young children( Reference Kinra, Radha Krishna and Kuper 17 ), whereas the IMS cohort was established to examine the association between rural to urban migration and cardiometabolic risk( Reference Ebrahim, Kinra and Bowen 18 ). To obtain a representative sample, participants were chosen on the basis of pre-defined age, sex, cohort, intervention group (in the case of the APCAPS) or rural/urban migrants (in the case of the IMS), and BMI categories (see online supplementary Tables S1 and S2). The target enrolment was 160 participants.
Demographic and anthropometric data
Demographic information was collected from all study participants using an interviewer-administered questionnaire. Weight was measured to the nearest 0·1 kg in light clothing without footwear, using a digital Seca scale (www.seca.com). Standing height was measured to the nearest 1 mm using a portable stadiometer (Leicester Height Measure; Chasmors Limited). Anthropometric measurements were taken twice, and the average of the two values for each measurement was used in the analysis. BMI was calculated as weight (kg)/height (m2).
Body composition of each participant was assessed by DXA and isotope dilution technique on the same day.
Isotope dilution technique
Participants arrived at the National Institute of Nutrition in the morning after an overnight fast. A baseline urine sample was collected on arrival for the measurement of background isotopic enrichment, followed by the administration of an oral dose of 18O (0·2 g/kg body weight) to each participant at about 09.00 hours. The bottle containing the dose was rinsed with 50 ml deionised water before its consumption by the participants. A light breakfast was provided with 50 ml water at about 10.00 hours. Any subsequent oral intake was avoided. Follow-up urine samples were collected 4 and 5 h after the intake of dose to allow complete equilibration of the isotope within the body water compartments( Reference Colley, Byrne and Hills 19 ). Aliquots of all the urine samples were stored in screw-capped glass containers at − 20°C until analysis. Isotopic enrichment in the pre- and post-dose urine samples, the dose given and the local tap water was measured using isotope ratio mass spectrometry (Hydra 20-20; SerCon) at St John's Research Institute, Bangalore, India. The CV calculated using repeated analysis for the natural background samples as well as for the enriched samples was less than 0·01 %. Each sample was analysed in duplicate, and the mean was used for the analysis. Total body water was calculated, allowing a correction by 0·7 % for in vivo exchange( Reference Racette, Schoeller and Luke 20 ). FFM was calculated from total body water using a hydration constant of 0·73. FM was calculated by subtracting FFM from body weight.
Dual-energy X-ray absorptiometry scans
Body composition was assessed by a whole-body DXA scan using a fan-beam DXA machine (Hologic Discovery A model, software version 12.5; www.hologic.com). The scanner was calibrated periodically with a phantom, and its performance was monitored according to the manufacturer's quality assurance protocol. During the scan, the participants were asked to lie supine on the scanning bed with their arms at their sides. Standard software options were used to calculate the total FFM and FM. FFM was the sum of lean soft tissue mass and bone mineral content. Precision estimates (CV%) of body composition by DXA based on repeat measurements in thirty participants were 0·7 and 1·4 % for FFM and FM, respectively.
Statistical analyses
All analyses were conducted using Stata (version 11.2; StataCorp). As FFM and FM showed a skewed distribution, these variables were log-transformed before analysis, and, therefore, the mean differences between the two are expressed as ratios. Other continuous variables were used in the original scale. Differences between the body composition estimates (FFM, FM and BF%) by DXA and isotope dilution technique were assessed using paired t tests. The Bland–Altman method was used to assess the agreement between the estimates of body composition determined by the two techniques( Reference Martin Bland and Altman 21 ). The mean difference in the estimates by the two techniques (bias) and their 95 % limits of agreement (2 sd of the mean difference) were calculated. As the bias and limits of agreement for FFM and FM were on a logarithmic scale, these values are presented as ratios. Correlation coefficients were calculated to examine the association between the average values of body composition measurements by the two methods and the difference between these methods, which indicates the proportional bias. All analyses were conducted for the whole sample and additionally stratified by sex.
Results
A total of seventy-three men and seventy-nine women participated in the study. Their characteristics are presented in Table 1. The participants were chosen to represent a wide range of BMI varying from 13·8 to 39·7 kg/m2. The total mass value measured by DXA showed a strong correlation with weight measured by the scale (0·99, P< 0·01). Although there was a strong correlation between the estimates of body composition measured by DXA and isotope dilution technique (FFM: r 0·95, FM: r 0·95, BF%: r 0·89 all P< 0·01), the estimates of FFM obtained by DXA were higher than those obtained by the isotope dilution technique in the whole sample as well as in the subgroups stratified by sex (Table 2). The estimates of FM and BF% obtained by DXA were lower than those measured by the isotope dilution technique. On average, DXA overestimated the FFM values by about 7 (95 % CI 6, 9 %) % compared with the isotope dilution technique (Table 3; Fig. 1(a)). However, the limits of agreement showed that 95 % of the estimates of FFM measured by DXA were expected to be between 9 % lower and 26 % higher than the values measured by the isotope dilution technique. For FM, the bias was greater, and, on average, the estimates by DXA were about 21 % lower than those by the isotope dilution technique (Table 3; Fig. 1(b)). The limits of agreement for FM were much larger ( − 54 to 17 %) than those for FFM between the two methods. There was no correlation between the bias and the average values of the estimates measured by the two methods for both FFM and FM, indicating that the bias in the estimates of FFM and FM did not change with the amount of FFM and FM, respectively. On average, the estimates of BF% measured by DXA were lower than those measured by the isotope dilution technique by 7·4 (95 % CI − 8·2, − 6·6 %) % (Table 3; Fig. 1(c)). The bias in the estimates of BF% was negatively correlated with the average values of BF%, indicating that the difference between the two methods was higher for the participants with lower values of BF% (Table 3). The estimates of FFM, FM and BF% measured by DXA explained about 89, 85 and 78 % of the variation in the respective estimates measured by the isotope dilution technique.
TM, total mass; DXA, dual-energy X-ray absorptiometry.
* P value was obtained from the paired t test of the difference.
* Mean bias and 95 % CI for fat-free mass and fat mass are expressed as the ratio of DXA:isotope dilution technique values. Bias is the difference (DXA − isotope dilution) between the log-transformed values of fat-free mass and fat mass estimated from the two techniques. The values of body fat percentage are given on the original scale.
† 95 % Limits of agreement (2 sd of the mean difference) expressed as the ratio of DXA:isotope dilution values of fat-free mass and fat mass. The values of body fat percentage are given on the original scale.
‡ r is Pearson's correlation coefficient between the difference between DXA and isotope dilution technique and the average of DXA and isotope measures of fat-free mass, fat mass and body fat.
§ Significance of the correlation coefficient.
Discussion
The present study compared the estimates of body composition measured by two precise techniques – DXA and isotope dilution technique – in apparently healthy, weight-stable Indian men and women with a wide range of BMI. In this sample of participants, the estimates of FFM were higher whereas those of FM and BF% were lower using DXA than using the isotope dilution technique. The agreement between the two methods was not good as indicated by the significant bias between these methods and wide limits of agreement, especially for the estimates of FM and BF%. The bias in the estimates of BF% measured by the two methods was higher for individuals with lower values of BF%. The present study indicates that these two methods cannot be used interchangeably as systematic differences exist between the estimates of body composition.
Previous studies that have compared the estimates of body composition by DXA and isotope dilution technique have reported inconsistent results. In general, studies that used older equipment (e.g. Hologic QDR 2000, Hologic QDR 1000W, Lunar DPX-L) with scans done in a pencil-beam mode have shown underestimation of FFM and overestimation of FM and BF% by DXA compared with the isotope dilution technique( Reference Fuller, Jebb and Laskey 22 , Reference He, Li and Kung 23 ). In contrast, studies that used newer equipment (e.g. Hologic QDR 4500W, QDR 4500A) have shown overestimation of FFM by DXA compared with the isotope dilution technique( Reference Schoeller, Tylavsky and Baer 9 , Reference Deurenberg-Yap and Deurenberg 24 ). For example, a study in Chinese women in 1999 has shown that DXA (Hologic QDR 2000) underestimated FFM by 0·5 kg and overestimated BF% by 0·8 %( Reference He, Li and Kung 23 ). Similarly, a study from the UK (n 28) in 1992 has also found that DXA (Hologic QDR 1000W) underestimated FFM by 0·2 kg compared with the isotope dilution technique( Reference Fuller, Jebb and Laskey 22 ). However, a later study by Deurenberg-Yap & Deurenberg( Reference Deurenberg-Yap and Deurenberg 24 ) in Chinese, Malays and Indians living in Singapore has shown overestimation of FFM and underestimation of BF% by DXA (Hologic QDR 4500W) compared with the 2H dilution technique. Similarly, a study by Schoeller et al. ( Reference Schoeller, Tylavsky and Baer 9 ) from the USA that compared body composition by DXA with other reference techniques in 1195 men and women (DXA compared with the isotope dilution technique in 395 participants) has also shown that DXA overestimated FFM by 1·8 to 4·7 kg and underestimated FM by about 1·3 to 5·1 kg. The findings of the present study that used a newer model of DXA (Hologic Discovery) are consistent with relatively recent studies that have shown overestimation of FFM by DXA compared with the isotope dilution technique. However, the magnitude of bias in the estimates of FFM (approximately 3 kg) and FM (approximately 4·5 kg) in the present study is larger than the bias reported in other studies.
A number of studies (Table 4) comparing the estimates of body composition by DXA with those by multi-component criterion methods have also reported inconsistent results( Reference Fuller, Jebb and Laskey 22 , Reference Arngrimsson, Evans and Saunders 25 – Reference Withers, LaForgia and Pillans 29 ). Although the majority of these studies reported underestimation of BF% by DXA, similar to the present study, a few studies have reported a bias in the opposite direction. For example, a study by Williams et al. ( Reference Williams, Wells and Wilson 30 ) compared DXA with a four-compartment (4C) model and reported the overestimation of FM and BF% by DXA in non-obese adults. In contrast, a few studies did not detect significant difference in BF% by DXA compared with the 4C model( Reference van Marken Lichtenbelt, Hartgens and Vollaard 31 – Reference LaForgia, Dollman and Dale 33 ).
F, female; M, male.
* Statistically significant difference between DXA and the 4C model (P< 0·05).
† Includes weight-stable, healthy adults.
Differences in the results of studies comparing the estimates of body composition by DXA with those by other techniques could be related to a number of factors such as the use of DXA machines by different manufacturers and differences in the scan mode or software used for analyses. Machines developed by different manufacturers as well as different models by the same manufacturer, although based on the same physical principles, differ in various aspects such as the generation of high- and low-energy X-ray beams, X-ray detectors, calibration methodology, algorithms used for selective tissue imaging, edge detection, region-of-interest definition, system calibration, etc.( Reference Genton, Hans and Kyle 34 ). Variations in the estimates of body composition with the machines developed by different manufacturers and even with different models by the same manufacturer have been reported( Reference Shepherd, Fan and Lu 35 – Reference Pearson, Horton and Green 38 ). In addition, isotope dilution technique has a number of limitations as the estimates of body composition are based on a number of assumptions including the equal distribution of a tracer in body water and constant hydration of FFM( Reference Krumbiegel 39 ). Both these techniques are thus error prone, and a lack of agreement between the methods for the estimation of body composition could be related to a number of factors that can lead to inaccuracies in the estimates.
However, limits of agreement between the two methods observed in the present study were wider (FFM: − 9,+26 %; FM: − 46,+17 %; BF%: − 17·3, 2·6 %) than those reported by other studies, the majority of which have reported the limits of agreement between ± 10 % of the mean( Reference Toombs, Ducher and Shepherd 40 ). In contrast, a few other studies have reported that DXA could under- or overestimate the FM of an individual by almost 28 %( Reference Fuller, Jebb and Laskey 22 ). One of the reasons for the narrow limits of agreement reported by other studies could be the exclusion of extreme values of the differences. For example, Schoeller et al. ( Reference Schoeller, Tylavsky and Baer 9 ) excluded observations in which the difference in the estimates of FFM measured by DXA and isotope dilution technique was >6 kg. The present study did not exclude observations with larger differences between the measurements, which may have contributed to a larger bias between the measurements reported herein.
An interesting finding of the present study is that the bias in the estimates of BF% by the two methods was higher at lower values of BF% (r − 0·345, P <0·001; Table 3). A previous study comparing the estimates of abdominal fat by DXA with those using MRI in this sample has also shown that overestimation of abdominal fat by DXA was greater in individuals with less abdominal fat( Reference Taylor, Kuper and Varma 41 ). It is possible that the algorithms used for the estimation of body composition by DXA produce a larger error at very low levels of body fat. A number of studies from other centres have shown that the bias in the estimates of body composition by DXA varied according to a number of factors including age, body size, body fat, sex, health status, type of the instrument, etc.( Reference Williams, Wells and Wilson 30 ).
An important strength of the present study includes enrolment of a large sample representing a broad range of age and BMI. In addition, the present study used 18O as the isotope tracer that may provide a more accurate estimate of total body water than a more commonly used 2H2O as 18O exchanges to a smaller degree with non-aqueous molecules( Reference Krumbiegel 39 ). As differences in body composition in relation to ethnicity are well known, population-specific validation studies comparing DXA with other precise methods are required. Therefore, the present study provides much-needed evidence on the comparability of DXA with the isotope dilution technique in a population group that is known to have a high percentage of body fat at a given BMI compared with other ethnic groups( Reference Yajnik, Fall and Coyaji 10 , Reference Kulkarni, Shatrugna and Nagalla 15 ). A limitation of the present study is the use of the isotope dilution technique for validating DXA measurements of body composition instead of a multi-component criterion technique. However, estimates of body composition using the isotope dilution technique are highly correlated with those using the criterion technique of the 4C model( Reference Deurenberg-Yap, Schmidt and van Staveren 27 ). A study comparing the estimates of body composition by densitometry, DXA and isotope dilution technique with those by the 4C model in Asian adults has shown that estimates of BF% by the isotope dilution technique had the lowest bias while those by DXA had the highest bias compared with the 4C model, suggesting that the isotope dilution technique may be the best two-compartment model for measuring body fat( Reference Deurenberg-Yap and Deurenberg 24 ).
In conclusion, the present study shows that estimates of body composition by two commonly used reference methods such as DXA and isotope dilution technique may be considerably different at the individual level, with particularly larger differences in the estimates of BF%. The two methods are therefore not directly interchangeable. However, the differences in the absolute values at the individual level may not affect the results of the studies exploring the relationship of body composition using either of these methods with health outcomes, as these values were highly correlated. Additional studies are required to develop correction factors that could be used to calibrate DXA in order to alleviate the differences in these two methods.
Supplementary material
To view supplementary material for this article, please visit http://dx.doi.org/10.1017/S0007114514001718
Acknowledgements
The present study was funded by the Wellcome Trust, UK (grant no. WT083707AIA). The study sponsor had no role in the study design, in the collection, analysis and interpretation of the data, in the writing of the report, or in the decision to submit the paper for publication.
The authors thank Dr B. Sesikeran, Director, National Institution of Nutrition, Hyderabad for providing the facilities and support system to carry out the study. They thank Mrs K. Usha Rani for conducting and analysing the DXA scans and Pete Shiarly for data management. They also thank Ms Sarita, St Johns Research Institute, Bangalore, India for carrying out the isotope analyses. The authors are extremely grateful to the committed and diligent fieldwork team led by Ms Santhi Bhogadi.
The authors' contributions are as follows: B. K., H. K. and A. V. K. conceived and carried out the study; J. C. W. conceived the study and helped in the interpretation of the results; B. K. analysed the data and wrote the manuscript. All authors were involved in the interpretation of the results and writing of the manuscript, and approved the final draft of the manuscript.
The authors declare that they have no competing interests.