Flexible endoscopes are frequently used for diagnostic and therapeutic interventions. They are semicritical devices because they encounter mucous membranes and are reprocessed using high-level disinfection destroying all microorganisms except small numbers of bacterial spores.Reference Kovaleva, Peters, van der Mei and Degener 1 Due to their complex design with several narrow and long lumens, flexible endoscopes are difficult to clean and disinfect. The estimated incidence of infections associated with gastrointestinal endoscopy is low (1 in 1.8 million procedures).Reference Kovaleva, Peters, van der Mei and Degener 1 , Reference Cowen 2 Nevertheless, contaminated endoscopes are among the medical devices most frequently linked to healthcare-associated outbreaks.Reference Gillespie, Kotsanas and Stuart 3 Moreover, because most reported outbreaks involve multidrug-resistant organisms, it is likely that most outbreaks are being missed.Reference Buss, Been and Borgers 4
Pathogen transmission is most often related to failure to comply with established cleaning and disinfection guidelines or with the use of defective equipment.Reference Kovaleva, Peters, van der Mei and Degener 1 Manual cleaning and drying are critical steps in reprocessing flexible endoscopes. Manual cleaning reduces the initial bioburden, enabling high-level disinfection to adequately decontaminate the endoscopes.Reference Kovaleva, Peters, van der Mei and Degener 1 Endoscope drying reduces the risk of bacterial proliferation during endoscope storage.Reference Saliou, Le Bars and Payan 5 , Reference Kovaleva, Degener and van der Mei 6 Another potential risk is biofilm growth inside endoscope channels,Reference Pajkos, Vickery and Cossart 7 , Reference Roberts 8 which compromises disinfection and facilitates microbial transmission.Reference Kovaleva, Peters, van der Mei and Degener 1 , Reference Kovaleva, Degener and van der Mei 6 – Reference Roberts 8
Possibly, early detection of endoscope contamination using microbiological surveillance could prevent cross-transmission and infection of patients.Reference Kovaleva, Peters, van der Mei and Degener 1 , Reference Kovaleva, Degener and van der Mei 6 Most European guidelines recommend routine surveillance of flexible endoscopes using the culture method. In the United States, there are currently no guidelines for routine monitoring,Reference Komanduri, Abu Dayyeh and Bhat 9 and agreement is lacking among guidelines regarding acceptance criteria, testing frequency, sampling technique, culture medium, and incubation conditions (Table 1).Reference Saliou, Le Bars and Payan 5
NOTE. MO, microorganisms; CFU, colony-forming units; … , not mentioned.
Because the sensitivity of different sampling strategies may vary, we aimed to compare different techniques of sampling flexible endoscopes. We compared 4 techniques reflecting current guidelines: flushing with sterile physiological saline (PHYS), flushing with neutralizing pharmacopeia diluent (NPD), and 2 flush-brush-flush techniques using PHYS in combination with the Olympus single-use, dual-ended cleaning brush or the PULL THRU brush.
METHODS
Endoscope Model
Endoscope model and sampling
For the endoscope model, we used polytetrafluoroethylene (PTFE) tubes with a 2.4-mm internal diameter and a 20-cm length (volume, 0.91 mL). Next, 20 PTFE tubes were each flushed with 1 mL from a positive hemoculture containing Klebsiella pneumoniae or Escherichia coli and were kept at room temperature for 24 hours (ie, non–biofilm-coated PTFE tubes). In addition, 2×20 PTFE tubes coated with biofilm (2 batches) were produced according to ISO 15883-5 Annex F and HTM 2030 standards that describe a model for growing biofilms representative of contamination inside an endoscope channel (ie, biofilm-coated PTFE tubes). Compared to the ISO standard, thinner PTFE tubes, closer to the actual size of endoscope channels, were used. Moreover, in addition to Pseudomonas aeruginosa (CIP A22), 2 relevant bacterial species (Klebsiella pneumoniae ATCC600703 and Staphylococcus epidermidis ATCC35984) were added to the biofilm to increase robustness.
We performed 4 sampling techniques 5 times on these PTFE tubes: (1) flushing with 10 mL PHYS (ie, 10PHYS), (2) flushing with 10 mL NPD (ie, 10NPD), (3) flush-brush-flush using 10 mL PHYS and a standard cleaning brush (Olympus, Hamburg, Germany) (ie, 10PHYS+SB), or (4) a PULL THRU brush (Medivators, Minneapolis, MN) (ie, 10PHYS+PT). Fluids (and brush tips) were collected in sterile containers. Moreover, 2 non–biofilm-coated PTFE tubes and 1 biofilm-coated PTFE tube were used as positive controls; they were cut into small pieces that were collected in sterile containers filled with 10 mL reverse osmosis water. These containers were then vortexed for 30 seconds, sonicated for 5 minutes, and vortexed again for 30 seconds. All samples and positive controls were processed for adenosine triphosphate (ATP) measurement and culture at the microbiology laboratory within 1 hour.
ATP measurement and microbial culture
ATP measurement was performed in duplicate using the Aquasnap Total test (Hygiena, Watford, UK) according to the manufacturer’s instructions with the SystemSURE Plus luminometer, except for NPD (due to interference with ATP quantification). Additionally, samples were diluted (1:10,000), and 100 µL was plated on trypticase soy agar (TSA), which was incubated for 7 days at 30°C. The total number of colony-forming units (CFU) was recorded.
Endoscopes
Ghent University Hospital hosts 42 endoscopes and 5 automated endoscope reprocessors (AER; ETD3, Olympus, Hamburg, Germany). The reprocessing cycle consists of (1) bedside precleaning, (2) manual leak testing, (3) cleaning in the cleaning facility using the Olympus standard cleaning brush, (4) mechanical leak testing, and (5) high-level disinfection using glutaraldehyde in the AER. With the exception of gastroscopes, which are stored in storage cabinets, all flexible endoscopes are stored in endoscope drying cabinets.
Samples
After distinct reprocessing procedures, we compared 2 sampling techniques on a subset of 40 endoscopes each: 10 gastroscopes, 10 coloscopes, 5 endoscopic retrograde cholangiopancreatography (ERCP) scopes, 5 echo-endoscopes, and 10 bronchoscopes. The commissioning date of each endoscope was recorded. Flushing of 100 mL PHYS (ie, 100PHYS) was performed because most guidelines recommend this technique. A 10-fold higher volume was used than with the endoscope model because of the greater length of the endoscope tubes.
A flush-brush-flush technique using PHYS and PULL THRU brush (ie, 100PHYS+PT) was selected as second technique based on the endoscope model results. The 100PHYS+PT method consists of flushing endoscope channels with 50 mL PHYS, brushing the biopsy channel using a PULL THRU brush, and flushing again with 50 mL PHYS. In addition, a special brush (MyBrush, Olympus, Hamburg, Germany) was used to sample the forceps elevator recess of ERCP scopes. Fluid was collected in sterile containers with the brush tip(s). To include all channels, a sterile connector (MAJ-621, Olympus, Hamburg, Germany) was used to flush the endoscopes (except for bronchoscopes having only 1 channel). Because ATP tests and PULL THRU brushes are not sterile, they were cultured 10-fold as a negative control.
ATP measurement and microbial culture
An Aquasnap Total test was performed on all samples. The remaining sample was filtered through a 0.45-µm membrane using an EZ-Stream pump (Merck Millipore, Molsheim, France). This membrane was put on TSA agar, which was incubated for 7 days at 30°C. The CFU count was recorded daily (except weekends), and matrix-assisted laser desorption/ionization time-of-flight mass spectrometry (MALDI-TOF MS; Microflex LT, Bruker Daltonics, Bremen, Germany) was used to identify indicator microorganisms: Enterobacteriaceae, Pseudomonas aeruginosa and other Pseudomonas spp., Stenotrophomonas malthophilia, Acinetobacter spp., Staphylococcus aureus, and Candida spp.Reference Saliou, Le Bars and Payan 5 , 12
Statistical Analysis
Culture results obtained from non–biofilm- and biofilm-coated PTFE tubes were expressed as recovery rate compared to positive controls. Mean and 95% confidence intervals (CI) of ATP and culture results were calculated for each sampling technique. Results were displayed in box-and-whisker plots. Statistical differences between sampling techniques were evaluated using the Kruskall-Wallis test (>2 groups) and the Mann-Whitney test (2 groups). P<.05 was considered statistically significant. Data from the endoscopes were also compared with French National Technical Committee on Nosocomial Infection (FNTCNI) criteria: (1) <25 CFU per endoscope and (2) absence of indicator microorganisms. 12 The χ2 test was used to compare the proportion of unacceptable samples. P<.05 was considered statistically significant.
A scatterplot and the Spearman correlation coefficient were used to check for a linear relationship between ATP and culture results. P<.01 was considered statistically significant. ATP values of endoscope samples with acceptable versus unacceptable culture results (based on FNTCI criteria) were compared, both for the entire group and for the 100PHYS and 100PHYS+PT subgroups. Receiver operator curve (ROC) analysis was used to determine the optimal cutoff ATP value. All statistical analyses were conducted using SPSS version 23 statistical software (IBM, Armonk, NY).
RESULTS
Endoscope Model
Mean ATP values obtained using different sampling techniques on non–biofilm-coated PTFE tubes were comparable: 5,574 for 10PHYS, 4,454 for 10PHYS+SB, and 5,014 for 10PHYS+PT (P=.37). Conversely, differences in ATP results using biofilm-coated tubes were significant (P=.045) (Table 2). In a pairwise comparison, only the difference between the 10PHYS and 10PHYS+PT subgroups was retained as statistically significant (P=.047).
NOTE. 10PHYS, flushing with 10 mL sterile physiological saline; 10NPD, flushing with 10 mL NPD; 10PHYS+SB, flush-brush-flush using 10 mL sterile physiological saline and a standard cleaning brush; 10PHYS+PT, flush-brush-flush using 10 mL sterile physiological saline and a PULL THRU brush; ATP, adenosine triphosphate; RLUs, relative light units; CI, confidence interval; CFU, colony-forming units; PTFE, polytetrafluoroethylene; …, experiment not performed because of interference of yellow-colored NPD solutions with measurement of ATP.
a Percentage recovery of a certain technique compared to the positive controls.
b Statistically significant difference between mean ATP value of 10PHYS and 10PHYS+PT sampling methods on biofilm-coated PTFE tubes (P=.047).
Culture results are presented as percentage recovery compared to the positive control (Table 2). The mean number of CFUs using the 4 different sampling techniques did not differ statistically in either non–biofilm- or biofilm-coated tubes (P=.53 and P=.27, respectively). However, the 10PHYS and 10PHYS+PT techniques had the highest mean yields for biofilm-coated PTFE tubes, while the 10PHYS and 10NPD methods produced the highest mean yields for non–biofilm-coated tubes.
There was no correlation between ATP measurements and culture results (rS=−0.08; P=.56).
Endoscopes
ATP and culture results varied widely with only a weak correlation (rS=0.38; P=.001). However, ATP values of samples classified as unacceptable based on FNTCNI criteria were higher compared to those classified as acceptable (P=.002). Subgroup analyses revealed that this finding was true only for samples obtained with the PHYS+PULL THRU brush method (P=.001) and not for the PHYS method alone (P=.9). An ATP cutoff value of >2 RLU on 100PHYS+PT samples was predictive for classification as unacceptable, with sensitivity and specificity of 87.5% and 71%, respectively.
Culture results showed important differences (Table 3). The culture yield using 100PHYS+PT sampling (mean, 43 CFU; range, 1–400 CFU) was significantly higher than for 100PHYS sampling (mean, 17 CFU; range, 0–500 CFU; P<.001). Subgroup analysis showed that addition of a PULL THRU brush to the sampling procedure resulted in higher culture results for all endoscope types, except for bronchoscopes. The CFU counts of negative controls for ATP tests (mean, 0.5 CFU; 95% CI, 0.1–0.9 CFU) and PULL THRU brushes (mean, 0.9 CFU;95% CI, 0.4–1.8 CFU) were negligible (Figure 1).
NOTE. 100PHYS, flushing with 100 mL sterile physiological saline; 100PHYS+PT, flush-brush-flush using 100 mL sterile physiological saline and a PULL THRU brush; RLUs relative light units; CFU, colony-forming units; MO, microorganisms; CI, confidence interval.
a Culture results were obtained from TSA agars with filter.
b In 1/10 bronchoscopes Aspergillus fumigatus was found.
c Statistically significant difference between mean culture yield using 100PHYS and 100PHYS+PT sampling (P<.001).
d Statistically significant difference between the number of samples classified as (un)acceptable using 100PHYS vs 100PHYS+PT (P=.03).
Indicator microorganisms were detected in 5 samples from different endoscopes: 2 Pseudomonas species, 2 Pseudomonas putida, and 1 Acinetobacter iwoffii. Overall, 3 indicator microorganisms were obtained from 100PHYS samples, and 2 were obtained from 100PHYS+PT samples. In a single bronchoscope sample obtained using 100PHYS+PT, Aspergillus fumigatus was detected. Results (ie, total CFU, ATP and identified microorganisms) of all endoscope samples are shown in Online Supplementary Table 1. Identified microorganisms were mainly skin commensals (eg, coagulase negative staphylococci and Micrococcus luteus) and environmental bacteria (eg, Bacillus spp.).
Using French acceptance criteria, the number of samples classified as unacceptable was significantly higher using the 100PHYS+PT technique (ie, 16 of 40) compared with the 100PHYS only method (ie, 7 of 40; P=.03). The age of the endoscopes was comparable between both groups (ie, 4.6 years for 100PHYS vs 4.8 years for 100PHYS+PT); no correlation was detected between endoscope age and culture results (rs=−0.07; P=.6).
In 16 of 80 samples, CFU counts were not recorded at 48 hours because plates were not read during weekends. From the remaining 64 samples, only 4 (6%) developed growth after 48 hours with 1–5 CFU and no indicator microorganisms. The other 60 (94%) samples either showed no growth (10 samples) or growth already developed at 48 hours (50 samples). In 35 of those 50 samples, there was no additional growth after 48 hours. The remaining 15 samples showed minor increases in CFU counts after 48 hours, but all indicator organisms grew within this time frame. Overall, all samples were classified correctly as (un)acceptable at 48 hours of incubation.
DISCUSSION
Microbiological surveillance of endoscopes is influenced by culture method and sampling technique, especially its recovery rate. However, guidelines show major differences with respect to recommended technique. To discriminate performance of sampling techniques, 4 were selected: (1) flushing with PHYS, (2) flushing with NPD, (3) flush-brush-flush using PHYS and standard cleaning brush, and (4) flush-brush-flush using the PHYS+ PULL THRU brush method. Retrograde sampling (from distal to proximal end)Reference Buss, Been and Borgers 4 was not included because it is not recommended in guidelines and is impractical. Samples were cultured on TSA agar after filtration. We processed our data using French acceptance criteria, which appear to have been based on expert opinion rather than on clinical evidence. 12
ATP measurement was performed in addition to culture. It is quick (<1 minute) and simple, but it fails to detect small quantities of microorganisms,Reference Batailler, Saviuc, Picot-Gueraud, Bosson and Mallaret 21 so it could be considered an indicator of endoscope cleanliness, notably to audit manual cleaning adequacy.Reference Fushimi, Takashina and Yoshikawa 19 , Reference Shin and Kim 20 Recommended maximum RLU values for samples taken at the end of reprocessing (during storage or just before reuse) are not available.
In vitro experiments revealed that, for biofilm-coated PTFE tubes, ATP values of 10PHYS+PT samples were significantly higher than those of 10PHYS samples. ATP values of non–biofilm-coated tubes were comparable among the 4 sampling techniques. Culture results showed that mean yield from biofilm-coated PTFE tubes was highest for 10PHYS and 10PHYS+PT techniques, whereas for non–biofilm-coated tubes 10PHYS and 10NPD produced the highest mean yield. However, differences in mean CFU count did not reach statistical significance. Taken together, because biofilm-coated PTFE tubes likely resemble the real-life situation more closely than non–biofilm-coated tubes, the PHYS+PULL THRU brush method was selected for comparison with PHYS alone, which is recommended for use on endoscopes by most guidelines because it is inexpensive and simple.
In our study, there was no correlation between ATP and culture results in in vitro experiments or in endoscope samples. This result corresponds to the findings of Batailler et al,Reference Batailler, Saviuc, Picot-Gueraud, Bosson and Mallaret 21 who concluded that ATP cannot be used as an alternative to microbiological tests for monitoring endoscope reprocessing. However, according to our data, ATP seems to be able to distinguish samples classified as acceptable from samples classified as unacceptable. Subgroup analysis showed that this is only true for 100PHYS+PT samples, not for 100PHYS samples. Using an ATP cutoff value of >2 RLU for 100PHYS+PT samples, sensitivity and specificity were 87.5% and 71%, respectively. Applying this cutoff to our results, 31 of 40 samples would have been immediately classified correctly: 17 acceptable and 14 unacceptable. There were 7 false-positive results and 2 false-negative results; both had >25 CFU per endoscope, and 1 sample also grew indicator microorganisms. Due to the intrinsic inability of ATP to detect small numbers of microorganisms and based on our limited data, microbiological culture remains necessary and should not be omitted. The value of ATP in this setting and the ATP threshold to discriminate acceptable from unacceptable endoscopes needs to be validated in larger studies.
On endoscopes, the 100PHYS+PT method yielded significantly higher culture results than the 100PHYS only method. Mechanical action seems to facilitate the release of organic matter and microorganisms. Also, the number of endoscope samples classified as unacceptable using French acceptance criteria was significantly higher using the 100PHYS+PT method: 40% for 100PHYS+PT versus 17.5% for 100PHYS. Notably, these differences are not influenced by endoscope age. Analysis of negative controls shows that differences cannot be explained solely by the use of nonsterile brushes. Moreover, subgroup analysis revealed that adding a PULL THRU brush to the sampling procedure resulted in higher culture results for all types of endoscopes, except for bronchoscopes. The simpler design of bronchoscopes (1 channel only), compared to more complex gastrointestinal endoscopes, may account for this difference.
Physical removal of soil by complete surface contact between the circular rubber discs of the PULL THRU brush and the lumen wall probably explains the superiority of the PULL THRU brush over the standard cleaning brush. Based on our findings, it could be argued to replace standard cleaning brushes with PULL THRU brushes for manual endoscope cleaning. Because current evidence is limited, future research on the efficacies of different brush types for manual cleaning of flexible endoscopes is warranted.Reference Charlton 22
In our study, the final results were obtained at 48 hours of incubation because almost all positive endoscopes (50 of 54) developed growth within this time frame. These results contrast with other studies in which 30%–45% of endoscope samples became positive after >2 days of incubation.Reference Saliou, Le Bars and Payan 5 , Reference Ofstead, Wetzler, Heymann, Johnson, Eiland and Shaw 23 Different sampling and culture protocols impede direct comparison of results. In a study compiling the results of >1,000 samplings on gastrointestinal endoscopes, only 55.5% of all contaminated endoscopes were positive at 48 hours of incubation. The risk of contamination was significantly reduced when endoscopes were kept in storage cabinets (as in our setting).Reference Saliou, Le Bars and Payan 5 Despite the fact that culture methods used by Saliou et al are comparable with those used in our study, sampling methods and reprocessing methods were different. Notably, we did not use neutralizers, which are known to improve microbial recovery. Therefore, it is possible that slow-growing microorganisms, causing a change in endoscope classification after 2 days, were unable to survive in physiological saline between sampling and culture.Reference Saliou, Le Bars and Payan 5 Overall, the reduced incubation period of 48 hours might have an important impact on logistical issues and workload, but this aspect needs further validation prior to inclusion in a surveillance protocol.
To the best of our knowledge, only 1 other study compared efficacies of several sampling techniques for microbial surveillance of endoscopes. Aumeran et alReference Aumeran, Thibert, Chapelle, Hennequin, Lesens and Traore 24 used an experimental model of biofilm grown on endoscope internal tubing and performed an in-use evaluation sampling endoscopes during routine clinical practice with 2 different sampling solutions. They concluded that the use of tensioactive sampling fluid was significantly more efficient. However, brushing was not included in this study; thus, direct comparison of the results is difficult.
Our study has several limitations. Endoscopes were sampled after distinct reprocessing cycles. Although endoscope conditions differed between samplings, consecutive sampling on the same endoscope would induce a greater sampling bias. As mentioned above, we did not use any substance to neutralize remaining high-level disinfectant (glutaraldehyde in our case). It is also possible that other culture conditions, such as incubation temperature (eg, 35°C instead of 30°C) or different agar plates (eg, blood agar), would generate a higher yield or allow growth of different microorganisms. However, because the focus of this study was the evaluation of various sampling techniques, comparison of different culture methods could be the object of a separate study. Finally, our study was conducted in a single center on a limited number of endoscopes. It remains to be demonstrated whether our findings can be extrapolated to other settings, where, for example, peracetic acid instead of glutaraldehyde is being used.
In conclusion, sampling methods influence recovery rate and thus results and interpretation of microbial surveillance cultures of flexible endoscopes. The association of brushing using a PULL THRU brush to the endoscope sampling procedure increased the yield of microbial surveillance culture. However, generally accepted criteria for endoscope culture need to be defined, ideally based on clinical data regarding the risk of nosocomial transmission. Moreover, thresholds may need to be adjusted depending on the sensitivity of the sampling technique. The added value of ATP in the surveillance of endoscopes needs to be confirmed in future studies. In our study, all endoscopes were classified correctly as acceptable or unacceptable at 48 hours of incubation. However, given the concern about slow-growing microorganisms, it seems prudent to extend the incubation period to 7 days.