Review of microRNA detection workflows from liquid biopsy for disease diagnostics

Published online by Cambridge University Press:  06 February 2025

Dulguunnaran Naranbat
Affiliation:
Center for Biomedical Engineering, School of Engineering, Brown University, Providence, RI, USA
Emilia Herdes
Affiliation:
Center for Biomedical Engineering, School of Engineering, Brown University, Providence, RI, USA
Nikos Tapinos
Affiliation:
Warren Alpert Medical School, Brown University, Providence, RI, USA Department of Neurosurgery, Rhode Island Hospital, Providence, RI, USA
Anubhav Tripathi*
Affiliation:
Center for Biomedical Engineering, School of Engineering, Brown University, Providence, RI, USA
*
Corresponding author: Anubhav Tripathi; Email: anubhav_tripathi@brown.edu
Rights & Permissions [Opens in a new window]

Abstract

MicroRNAs have emerged as effective biomarkers in disease diagnostics, particularly cancer, due to their role as regulatory sequences. More recently, microRNAs have been detected in liquid biopsies, which hold immense potential for early disease diagnostics. This review comprehensively analyses distinct liquid biopsy microRNA detection methods validated with clinical samples. Each step in the microRNA detection workflow, including sample collection, RNA isolation, processing, and detection of target microRNAs, has been thoroughly assessed. The review discusses the advantages and limitations of established and novel techniques in microRNA detection workflows, discussing their diagnostic capabilities and potential for future implementation at scale.

Type
Review
Creative Commons
Creative Common License - CCCreative Common License - BY
This is an Open Access article, distributed under the terms of the Creative Commons Attribution licence (http://creativecommons.org/licenses/by/4.0), which permits unrestricted re-use, distribution and reproduction, provided the original article is properly cited.
Copyright
© The Author(s), 2025. Published by Cambridge University Press

Introduction

In recent decades, a new frontier in disease diagnostics, particularly for cancer, has emerged with microRNA (miRNA) detection. Numerous researchers have underscored the pivotal role of specific miRNAs in the onset of tumorigenesis, suggesting their potential as diagnostic biomarkers. (Refs Reference Hejazi, Baghbani and Amini1, Reference Asl, Amini and Najafi2, Reference Zhang, Dahlberg and Tam3) miRNAs are short, non-coding RNA sequences, 19–25 nucleotides long, that play a crucial role in post-transcriptional gene regulation. (Refs Reference Paranjape, Slack and Weidhaas4, Reference Precazzini, Detassis and Imperatori5) The maturation of functional miRNAs begins with a long single-stranded RNA that forms a hairpin structure called primary miRNA (pri-miRNA). In the nucleus, the pri-miRNA is cleaved by an enzyme complex into precursor miRNA (pre-miRNA), which is then exported to the cytoplasm. An enzyme cleaves it into a double-stranded RNA, which unwinds to produce mature miRNA. Mature miRNAs bind to target messenger RNAs, modulating gene expression and contributing to disease progression. (Refs Reference Paranjape, Slack and Weidhaas4, Reference Bushati and Cohen6) Numerous target miRNAs have been identified to exhibit either overexpression or underexpression in association with specific diseases. (Ref. Reference Paranjape, Slack and Weidhaas4) Therefore, once the correlation between specific miRNAs and disease is established, the abundance of the target miRNA in affected individuals can be measured and compared to controls, serving as a diagnostic marker. (Refs Reference Paranjape, Slack and Weidhaas4, Reference Kim and Croce7)

Currently, cancer diagnosis predominantly relies on tissue biopsies. While effective, these biopsies are invasive, often difficult to access (such as in brain tumours), and not well-suited for continuous monitoring. (Refs Reference Gastman, Agarwal and Berger8, Reference Hodel, Laux and Farei-Campagna9, Reference Crowley, Di Nicolantonio and Loupakis10, Reference Robertson and Baxter11) As a result, miRNA detection through liquid biopsies - such as blood, urine, and saliva - has become a rapidly advancing area of research. This progress enhances the development of cost-effective and time-efficient point-of-care (POC) testing systems for miRNA detection, with significant potential to improve early disease screening and diagnosis. (Ref. Reference Wang, Sun and Zhang12)

While miRNAs in liquid biopsy show promise as disease biomarkers, their practical applicability is still under evaluation. Successful implementation necessitates a standardised workflow from sample collection to detection [Figure 1 ]. The absence of standardised protocols and guidelines for extracting, quantifying, and normalising miRNA levels has significantly hindered the reliability and validity of miRNA detection methods as they transition from basic research to clinical applications. (Refs Reference Felekkis, Pieri and Papaneophytou13, Reference Faraldi, Gomarasca and Sansoni14) Novel detection methods are continually advancing, but validation with clinical data is crucial to establish their feasibility. This review comprehensively analyses distinct miRNA detection methods from liquid biopsies validated with clinical samples. Here, we examine each step in the miRNA detection workflow, including sample collection, RNA isolation, processing and detection of target miRNAs [Figure 2]. A comprehensive list detailing each study’s workflow and the reported limit of detection (LOD) is provided in Table 1. The figures in this review were made with BioRender.com (Toronto, Canada).

Figure 1. General overview of miRNA detection workflow. The steps are as follows: A. Sample collection from the patient, B. sample treatment, C. small RNA isolation including miRNA, D. sample processing including methods of target amplification, library preparation, and surface hybridisation, and E. detection of signature miRNA.

Figure 2. Comparison of clinically validated miRNA detection workflows. Sample type, treatment, RNA isolation, processing, and detection methods are outlined.

Table 1. Overview of unique miRNA detection workflow for targeted disease diagnostics from liquid biopsies

Note: Normalised data not reported is abbreviated as NDNR.

Liquid biopsy samples and treatment

Numerous studies suggest a correlation between miRNA levels in whole blood and tissue samples, presenting an opportunity to develop a less invasive method for accurate target miRNA detection. (Refs Reference Chen, Wang and Cui15, Reference Srinivasan, Yeri and Cheah16, Reference Liu, da Cunha and Rezende17) This review focused on miRNA detection methods using blood, saliva, and urine samples.

Peripheral blood

Blood is the most commonly used liquid biopsy sample for detecting miRNA due to its higher concentration of detectable miRNAs than other liquid biopsies. (Refs Reference Kim and Croce7, Reference Grimaldi, Lapucci and Salvatore18, Reference Balakittnen, Ekanayake Weeramange and Wallace19, Reference El-Mogy, Lam and Haj-Ahmad20) Evidence suggests that miRNAs circulate in peripheral blood within cells, encapsulated in extracellular vesicles (EVs), or as cell-free molecules. Furthermore, miRNAs encapsulated in protein complexes or EVs gain stability, making them ideal biomarkers. (Ref. Reference Sanz-Rubio, Martin-Burriel and Gil21) Therefore, blood-based miRNA detection represents a promising frontier but necessitates standardised extraction methods to prevent procedural variations from affecting miRNA levels. (Ref. Reference Takizawa, Matsuzaki and Ochiya22) EV-derived miRNAs have demonstrated better diagnostic capabilities in some cases; for example, target miRNAs – miR-200c-3p and miR-21-5p have performed better in distinguishing patients with prostate cancer when analysed in plasma EVs compared to whole plasma with the potential to be a consistent source of RNA. (Ref. Reference Endzeliņš, Berger and Melne23) However, this may not be true depending on the specific target miRNA or pathological condition, as seen with miR-375, which showed better detection performance in whole plasma compared to EVs. (Ref. Reference Endzeliņš, Berger and Melne23)

Peripheral blood is typically collected in vacutainer tubes and processed differently based on the application. Plasma or serum is ideal for detecting miRNA as it contains miRNA encapsulating EVs and has fewer contaminants than whole blood [Figure 3] (Refs Reference Rieske, Kutcher and Audia24, Reference Ding, An and Zhao25, Reference Ban and Song26, Reference Sandau, Wiedrick and McFarland27). Plasma is obtained by adding an anticoagulant and centrifugation. In contrast, serum is obtained by clotting the blood and extracting the liquid layer. Both contain EVs and platelets, which influence miRNA concentration. Filtering out platelets after centrifugation improves the quality and reproducibility of EV-miRNA analysis. (Ref. Reference Bracht, Los and van Eijndhoven28)

Figure 3. Blood treatment for plasma and serum separation. Plasma is obtained by centrifugation and treatment with an anticoagulant, such as EDTA, sodium citrate, or heparin. The serum is obtained by extracting the remaining liquid after clotted blood.

Anticoagulants like EDTA and sodium citrate are essential for preserving blood components before plasma separation. (Ref. Reference Jin and Gopinath29) However, studies on miRNA detection in patients with cardiovascular diseases have found that administering heparin can affect miRNA concentrations in the blood, complicating the normalisation of miRNA levels. (Refs Reference Kaudewitz, Lee and Willeit30, Reference Mompeón, Ortega-Paz and Vidal-Gómez31, Reference Ma, Fu and Masula32) Furthermore, heparin can co-isolate with RNA, interfering with downstream applications like qPCR, and thus is not recommended in these cases. (Ref. Reference Boileau, Lino Cardenas and Lindsay33) Interestingly, there have been reports of the miRNA levels in serum remaining more stable over time compared to plasma, suggesting that samples without anticoagulants may be more suitable for miRNA detection. (Refs Reference Jaeschke, Harvey and Tsurkan34, Reference Wakabayashi, Marumo and Ekawa35) One significant concern with plasma and serum samples is hemolysis, a process where blood cells release their contents, including miRNAs, potentially affecting measured miRNA levels. (Refs Reference Ye, Feldman and Valentino36, Reference Myklebust, Rosenlund and Gjengstø37) The impact of hemolysis varies among different miRNAs, with some exhibiting greater stability in blood than others. For instance, miR-371-3p, a biomarker for testicular germ cell tumour detection, has been shown to be affected by hemolysis. (Refs Reference Ye, Feldman and Valentino36, Reference Myklebust, Rosenlund and Gjengstø37)

While target miRNA concentration levels may differ depending on overall pathological conditions, the concentration of miRNA in serum samples has been observed to be higher than in plasma samples. (Ref. Reference Wang, Yuan and Cho38) This difference may be linked to the coagulation process affecting extracellular miRNA in blood samples. (Ref. Reference Wang, Yuan and Cho38) Additionally, a case study of miRNA expression from patients with acute myocardial infarction showed that circulating miRNA concentrations in plasma and serum samples differ, indicating that research findings using distinct biological materials cannot be directly compared. (Ref. Reference Mompeón, Ortega-Paz and Vidal-Gómez31) Common reference genes such as RNU44, RNU48, and MammU6 were used to normalise target miRNA expression levels, but they also showed variation across different sample types, leading to inconsistency. (Ref. Reference Wang, Yuan and Cho38)

Plasma and serum have been the most common sample types for miRNA detection after blood extraction. However, Chen (Ref. Reference Chen, Chen and Kong117) demonstrated that erythrocytes could also be used to detect miRNA associated with β-thalassemia, a blood disorder that affects haemoglobin. (Ref. Reference Zhang, Cheng and Xu39) Despite lacking a nucleus and enclosed miRNAs, erythrocytes can carry cell-free miRNAs originating from various tissues circulating in the blood. (Ref. Reference Sun, Yu and Niu40) While erythrocytes are likely not a suitable sample for all target miRNAs, they are used effectively in conditions that directly impact erythrocytes.

Urine, Saliva and other Bodily Fluids

Although blood is the most extensively tested liquid biopsy sample for miRNA detection, successful detection from saliva and urine samples in clinical samples has also been achieved. While blood collection is less invasive than tissue biopsy, saliva and urine collection is non-invasive. However, the range of detected miRNAs in saliva and urine is generally less than half of those in the blood, limiting their use to specific diseases. (Ref. Reference El-Mogy, Lam and Haj-Ahmad20)

Free from protein contamination, urine samples have been effectively used to detect kidney, prostate, and bladder diseases through miRNA detection. (Refs Reference Zhang, Zhu and Chen41, Reference Yao, Liu and Lu42, Reference Shankar, Shetty and N.S43, Reference Zheng, Zhang and Johnson44) In this review, cases of detection methods, including thermal cycling amplification, isothermal amplification and nCounter, have been discussed in detecting miRNA from urine.

While El-Mogy (Ref. Reference El-Mogy, Lam and Haj-Ahmad20) found that saliva contains the lowest range of detectable miRNAs, saliva samples have been successfully used to detect polyomavirus, human papillomavirus, and head and neck squamous cell carcinomas in clinical samples, as these viruses and cancers affect the oral cavity, allowing discernible differences in salivary miRNAs to be detected. (Ref. Reference El-Mogy, Lam and Haj-Ahmad20), (Refs Reference Wan, Vagenas and Salazar45, Reference Mamana, Stincarelli and Sarmento46) Despite the potential use, limited clinical studies have used salivary miRNAs with novel processing methods due to the low range of miRNAs in saliva.

Other bodily fluids, such as cerebrospinal fluid (CSF) and pleural fluid, have also been effectively utilised as sources of miRNAs in patient samples. (Refs Reference Marqués, Pont and Hidalgo47, Reference Baraniskin, Kuhnhenn and Schlegel48) Unlike blood collection, obtaining CSF necessitates a lumbar puncture, while acquiring pleural fluid requires thoracentesis. (Refs Reference Jane and Wray49, Reference Farah, Sidar-Orlin and Sharabi-Nov50) Both procedures are significantly more invasive, demand specialised training, and carry higher risks and potential side effects compared to standard blood draws. While studies using other bodily fluids as liquid biopsies for clinical research are currently limited, they represent an emerging research direction in cancer detection.

It is important to note that standardisation and pre-analytical normalisation are crucial when comparing the capacity of detection methods from different liquid biopsy sample types. Multiple different procedures for establishing a consistent workflow have been suggested by ensuring standardised operating procedures (SOPs) for miRNA sample collection and detection. These procedures often improve reproducibility and provide quality control measures to help minimise variables during sample collection and testing. Variables such as collection method, turn-around time, anti-coagulation (such as avoiding heparin), storage conditions, volume, or formalin fixation options should all be considered. (Refs Reference Khan, Lieberman and Lockwood51, Reference Farina, Wood and Perrapato52) Given the sensitive nature of RNA, internal standardisation, such as freeze–thaw cycles, sample cold-chain systems, or qualitative assessments, such as hemolysis rate, are important factors when validating results. Depending on the sample type, different endogenous controls or spike-ins should be used for normalisation as reference genes here. (Ref. Reference MacLellan, MacAulay and Lam53)

RNA isolation methods

After sample collection and treatment, all small RNA, including miRNAs, are usually isolated through extraction and purification procedures [Figure 1]. Several different isolation methods exist as commercially available kits for this process. The studies examined in this review utilise various miRNA extraction techniques, including commercial kits or modified protocols.

Organic phase separation-based extraction

Organic phase separation-based extraction is a widely used RNA isolation method. (Ref. Reference Wright, de Silva and Purdie54) Here, the method often employs phenol-containing solutions to lyse cells and vesicles, separating RNA from other cellular components. During lysis, often freely existing amphiphilic detergents create spherical micelles that insert themselves between the lipid molecules of the vesicle. This weakens the interactions that hold the bilayer together, thereby releasing nucleic acids from the vesicle [Figure 4A] (Ref. Reference Holohan, Feely and Li55). Subsequently, the lysed samples are centrifuged to produce three distinct layers: 1) a lower organic phase, 2) an interphase containing unwanted protein and remaining DNA, and 3) an upper aqueous phase containing the RNA of interest. (Ref. Reference Tan and Yiap56) After isolating the aqueous phase, numerous centrifugation and washing steps are required to purify the small RNA. Adding glycogen and tRNA to the aqueous phase of plasma improves the extraction efficiency due to their roles as carrier molecules, which aid in forming precipitates that trap miRNA. (Refs Reference McAlexander, Phillips and Witwer57, Reference Ban, Chae and Yoo58) While effective, the organic-phase separation method is manually intensive, laborious, and contains caustic and corrosive materials that must be handled cautiously. (Ref. Reference Espy, Uhl and Sloan59) These factors increase the risk of errors and contamination, especially with larger sample sizes that demand more manual processing. (Ref. Reference Espy, Uhl and Sloan59)

Figure 4. Overview of the three common types of RNA isolation for miRNA detection. A. Organic phase separation-based extraction method where vesicles are lysed, and contaminants are separated by centrifugal force. B. Spin column-based extraction method with a similar lysis procedure but where RNA is bound and eluted through a silica filter and C. Magnetic bead-based extraction method where RNA is bound to silica-based magnetic nanoparticles through ionic bridges supported by cations. The RNA is eluted sequentially in a water-based solution. (Washing steps not shown).

Spin column-based extraction

The spin column-based extraction method provides a safer and simpler alternative to organic extraction for miRNA isolation. (Ref. Reference McAlexander, Phillips and Witwer57) Similar to organic extraction, a lysis buffer is also used to lyse membranes. However, instead of organic solvents, it employs a silica membrane column and centrifugal force to capture RNA while allowing impurities to flow through [Figure 4B]. Due to the centrifugation requirement, the method remains challenging to automate, costly, and time-consuming. (Ref. Reference Diefenbach, Lee and Kefford60) The miRNeasy Serum/Plasma Kit (Qiagen) has consistently demonstrated superior miRNA recovery in terms of quality and quantity compared to other commercial spin column-based kits available. (Refs Reference Sriram, Khanka and Kedia61, Reference Timmerman, Gillis and Mego62).

Magnetic bead-based extraction

Magnetic bead-based small RNA isolation employs a unique solid-phase reversible immobilisation (SPRI) bead technique. (Ref. Reference Oberacker, Stepper and Bond63) Silica-based carboxyl-coated magnetic beads capture small RNA molecules from lysates, aided by a binding buffer containing polyethylene glycol (PEG) or alcohol, salt, and RNase inhibitors. PEG and alcohol create a “crowding” effect, forcing nucleic acids to cluster and intensify their negative charge. (Ref. Reference Scott, Shaheen and McGuinness64) Additionally, the salts’ cations attract the negatively charged phosphates of the nucleic acid, creating an ionic bridge between carboxylated SPRI beads. (Refs Reference Harnett, Weir and Michieletto65, Reference Stortchevoi, Kamelamela and Levine66) Selection of small RNAs from the sample is often achieved by modifying the ratio of PEG and isopropanol during SPRI clean-up [Figure 4C] (Ref. Reference Fishman and T.Lamm67). Magnetic bead-based extraction is advantageous due to its relatively low cost and quicker processing time, yet is limited due to its comparatively lower nucleic acid recovery rates. (Ref. Reference Diefenbach, Lee and Kefford60) However, Timmerman (2021) reported that at low-volume inputs, the magnetic bead-based method outperformed the spin-column based miRNeasy Serum/Plasma Kit (Qiagen) (Ref. Reference Timmerman, Gillis and Mego62).

Other extraction methods

Preparative polyacrylamide gel electrophoresis (PAGE) is another RNA isolation method in which input samples are loaded onto a polyacrylamide gel, and an electric current separates the RNA by size. (Ref. Reference Petrov, Wu and Puglisi68) In PAGE, a denaturing condition in the gel disrupts the hydrogen bonds formed by the nucleotides of the RNA input, allowing the RNA to migrate evenly by the size-sieving effect of the gel matrix. PAGE has been successfully utilised by Xie (Ref. Reference Xie, Chen and Zhang69) to detect miRNA from serum. (Ref. Reference Xie, Chen and Zhang69)

Interestingly, some recent studies have explored methods that successfully bypass RNA isolation altogether. (Refs Reference Wang, Lu and Tang70, Reference Joshi, Deitz-McElyea and Liyanage71, Reference Dai, Wang and Tan72) These methods are discussed further in the Processing and Detection methods section. If miRNA detection workflows without extraction achieve accuracy and sensitivity comparable to those with extraction, this advancement could significantly advance the development of POC testing platforms by drastically simplifying the entire detection process.

Extraction of miRNA exclusively from EVs is sometimes favoured because the target miRNAs may be more specific to EVs. (Ref. Reference Yekula, Muralidharan and Kang73) Hence, several procedures, either prior to extraction or with combination extraction methods, have been utilised to extract EV-miRNA. Common techniques include ultracentrifugation, which is considered the gold standard for EV isolation, size-exclusion chromatography, and immunoaffinity capture processes, all of which have been reported to extract EV-miRNA efficiently. (Refs Reference Zhang, Jeppesen and Higginbotham74, Reference Park, Bae and Seong75, Reference Böing, van der Pol and Grootemaat76, Reference McNamara and Dittmer77) Multiple different commercial exosome isolation kits and exosome RNA isolation kits have also been made available for miRNA downstream processes, such as Total Exosome RNA & Protein Isolation Kit by Invitrogen (Waltham, MA) or RNeasy Plus Kits for RNA Isolation by Qiagen (Germantown, MD).

Processing and Detection methods

Thermal cycling amplification

The most common method of thermal cycling amplification is reverse transcription-quantitative polymerase chain reaction (RT-qPCR). RT-qPCR is considered the gold standard for detecting small RNAs, including miRNAs, due to its cost-effectiveness, sensitivity, specificity, and large dynamic input range. (Ref. Reference Chen, Tan and Wong78) Two versions of miRNA specific RT-qPCR methods are discussed here.

Stem-loop RT-qPCR

Stem-loop RT-qPCR employs a universal stem-loop RT primer binding at the 3′ end of the miRNA template. (Ref. Reference Chen, Ridzon and Broomer79) The RNA secondary structure is melted and elongated in the RT thermal cycler to generate complementary DNA (cDNA) strands. These are then amplified using miRNA-specific forward and reverse primers [Figure 5A]. PCR inhibition by reverse transcriptase is common and is due to low template concentration; thus, cDNA strands must be diluted before being added to the PCR mix. (Refs Reference Sellner, Coelen and Mackenzie80, Reference González, Kato and Horikoshi81) Dilution also reduces the concentration of competing RT primers with PCR primers. (Ref. Reference Xue, Wang and Jiang82) It is important to note that the RT and qPCR steps are separate and are not done in a single pot. The amplified products can be detected and quantified using SYBR green fluorescent molecules, which bind to any double-stranded DNA (dsDNA) [Figure 5Ai] or a fluorescent probe [Figure 5Aii].

Figure 5. Schematic illustration of thermal amplification methods for miRNA detection. A. Stem-Loop RT-qPCR (Refs Reference Zhang, Zhu and Chen41, Reference Mamana, Stincarelli and Sarmento46, Reference Chen, Ridzon and Broomer79, Reference Lin, Mao and Shu83, Reference Resnick, Alder and Hagan84) with (i) SYBR Green detection - universal and (ii) Probe-based detection - specific; B. Poly-A Tail RT-qPCR; (Refs Reference Wan, Vagenas and Salazar45, Reference Almohsen, Al-Rubaie and Habib86) C. Sensitive and Multiplexed One-Step RT-qPCR (SMOS-qPCR); (Ref. Reference Xue, Wang and Jiang82) and D. Droplet digital PCR (ddPCR). (Refs Reference Cirillo, Margiotti and Mesoraca88,Reference Torii, Suzuki and Fukuda92)

Lin (Ref. Reference Lin, Mao and Shu83) used stem-loop RT-qPCR with SYBR green fluorescence to detect three miRNAs in serum samples of stage IV lung cancer patients. (Ref. Reference Lin, Mao and Shu83) Similarly, Resnick (Ref. Reference Resnick, Alder and Hagan84) used a probe-based fluorescence detection method to detect six miRNAs in serum samples of ovarian cancer patients (Ref. Reference Resnick, Alder and Hagan84) Zhang (Ref. Reference Zhang, Zhu and Chen41) utilised stem-loop RT-qPCR to detect upregulated urinary extracellular vesicle miRNAs in early-stage renal cell carcinoma patients. (Ref. Reference Zhang, Zhu and Chen41) (Ref. Reference Mamana, Stincarelli and Sarmento46) reported detecting a polyomavirus-related miRNA in saliva and plasma from renal transplant patients using stem-loop RT-qPCR. (Ref. Reference Mamana, Stincarelli and Sarmento46) Currently, there are two commercially available RT-qPCR kits for detecting microRNAs in liquid biopsies: Applied Biosystems’ TaqMan primer-probe assay and Qiagen’s miRCURY locked nucleic acid (LNA) assay [Table 1].

Poly(A)-tailing RT-qPCR

Instead of using stem-loop primers for RT, in poly(A)-tailing RT-qPCR, the 3′ end of the target miRNA sequence is polyadenylated. Subsequently, a universal RT primer transcribes the cDNA template before PCR amplification [Figure 5B] (Ref. Reference Kang, Zhang and Liu85). Almohsen (Ref. Reference Almohsen, Al-Rubaie and Habib86) used poly(A) tailing with SYBR green fluorescence to detect significant differences in the expression of target miRNAs from platelet-poor plasma in patients with acute myeloid leukaemia. (Ref. Reference Almohsen, Al-Rubaie and Habib86) Wan (Ref. Reference Wan, Vagenas and Salazar45) have also successfully used poly(A)-tailing RT-qPCR to detect HPV-positive head and neck squamous cell carcinomas from saliva samples. (Ref. Reference Wan, Vagenas and Salazar45) It is worth noting that sequence-specific and universal probes can be utilised for detecting amplified products in poly(A)-tailing RT-qPCR (Ref. Reference Vautrot, Behm-Ansmant, Biassoni and Raso87).

Other thermal amplification methods

Xue (Ref. Reference Xue, Wang and Jiang82) reported a sensitive and multiplexed one-step RT-qPCR (SMOS-qPCR) technique for detecting target miRNAs, validated with serum samples of oesophagal cancer patients. (Ref. Reference Kang, Zhang and Liu85) SMOS-qPCR combines RT and PCR into a single step by utilising an extended primer for RT, which enables efficient binding and a higher reverse transcription rate [Figure 5C]. Compared to stem-loop RT-qPCR, which uses a lower primer concentration in the mix due to competition concerns during PCR, SMOS-qPCR uses a larger sequence that enhances reverse transcription efficiency. (Ref. Reference Xue, Wang and Jiang82) Among the novel detection methods discussed in this review, this system achieved the lowest reported LOD of 0.1 zM. (Ref. Reference Kang, Zhang and Liu85)

Droplet digital PCR (ddPCR) is an ultra-precise detection method used by Cirillo (Ref. Reference Cirillo, Margiotti and Mesoraca88) to detect signature miRNAs in serum samples of ovarian cancer patients. (Ref. Reference Cirillo, Margiotti and Mesoraca88) Unlike regular PCR, which amplifies nucleic acids in a single reaction, ddPCR partitions DNA samples into tens of thousands of water-in-oil nanodroplets, which all undergo PCR separately yet simultaneously in a single container [Figure 5D] (Ref. Reference Olmedillas-López, García-Arranz and García-Olmo89). Although this workflow requires a cumbersome setup, it effectively determines the absolute quantity of miRNA. (Refs Reference Quan, Sauzade and Brouzes90, Reference Ito, Kawashima and Fujikawa91) Similarly, Torii (Ref. Reference Torii, Suzuki and Fukuda92) successfully used ddPCR to detect miRNA from urine exosomes in infants with congenital cytomegalovirus. (Ref. Reference Torii, Suzuki and Fukuda92) When comparing the performance of ddPCR and RT-qPCR for detecting target miRNA, RT-qPCR shows greater sensitivity than ddPCR, especially at low concentrations. This further emphasises the need to optimise ddPCR workflows before their widespread implementation in clinical settings. (Refs Reference Limothai, Chuaypen and Poovorawan93, Reference Nuno, Lafin and Scarpini94)

Isothermal amplification

Isothermal amplification methods are continuously being developed, which, unlike PCR, occur at a constant temperature. The primary advantages of isothermal amplification include eliminating costly thermal cycling equipment and its suitability for POC applications. (Ref. Reference Oliveira, Veigas and Baptista95) However, in general, they are still limited by their relatively lower specificity levels, which can be attributed to their low amplification temperatures. (Ref. Reference Oliveira, Veigas and Baptista95)

Rolling circle amplification

One of the most common isothermal amplification techniques is rolling circle amplification (RCA). (Ref. Reference Ali, Li and Zhang96) Here, a circular DNA probe binds to the target miRNA, forming a template for continuous DNA synthesis and amplifying the miRNA sequence. (Ref. Reference Wang, Zhao and Chen97) Xie (Ref. Reference Xie, Chen and Zhang69) have reported the successful use of apyrimidinic endonuclease 1 (APE1) assisted RCA in dual-signal amplification to detect a target miRNA from serum samples in Alzheimer’s patients. (Ref. Reference Xie, Chen and Zhang69) The RCA + APE1 system used two probes: a dumbbell-shaped probe (DP) and a reporter probe (RP), where the sealed DP transforms into an active circular structure in the presence of target miRNA and phi29 DNA polymerase [Figure 6A]. Xu (Ref. Reference Xu, Wu and Liu98) have also successfully detected miRNA in clinical samples using a reconstructed conventional linear padlock probe to improve the performance of RCA. (Ref. Reference Xu, Wu and Liu98) By creating two stem-loops at the terminal ends of the padlock probe, the target recognition is enhanced by triggering a hyperbranched RCA (HRCA) reaction [Figure 6B].

Figure 6. Schematic illustration of isothermal amplification methods for detecting miRNA. A. Rolling circle amplification (RCA) + APE1 enzyme; (Ref. Reference Xie, Chen and Zhang69) B. Padlock-assisted hyperbranched rolling circle amplification (HRCA); (Ref. Reference Xu, Wu and Liu98) C. Primer-based loop-mediated isothermal amplification (PS-LAMP); (Ref. Reference Talap, AL-maskri and Shen100) D. Catalytic assembly of DNAzyme integrated with primer exchange reaction (CDiPER); (Ref. Reference Zhang, Li and Jiang101) and E. Multiplexed exponential amplification reaction (MEXPAR). (Ref. Reference Yao, Liu and Lu42)

Loop-mediated isothermal amplification

Loop-mediated isothermal amplification (LAMP) is another amplification technique in which four to six primers are used to recognise the target miRNA sequence. (Ref. Reference Notomi, Okayama and Masubuchi99) Talap (Ref. Reference Talap, AL-maskri and Shen100) have reported a modified ligation-initiated phosphorothioate primer-based LAMP (PS-LAMP) strategy for target miRNA detection from papillary thyroid cancer patients. (Ref. Reference Talap, AL-maskri and Shen100) This method involves ligating two linker probes, stem-loop linker (SLL) and phosphorylation stem-loop linker (PP-SLL), with the target miRNA to obtain an active amplicon. Strand-displacing Bst DNA polymerase initiates the amplification using two inner primers (PS-FIP/BIP) to form a loop structure. The remaining loop primers then initiate further cyclic amplification cycles. Both methods utilise probes with fluorescence and quencher to detect amplification rate [Figure 6C] (Ref. Reference Notomi, Okayama and Masubuchi99).

Other enzyme-based isothermal amplification methods

Other enzyme-based isothermal amplification methods were also explored for detecting miRNAs. For example, Zhang (Ref. Reference Zhang, Li and Jiang101) introduced a novel isothermal amplification method: catalytic DNAzyme integrated with primer exchange reaction (CDiPER) for miRNA detection in diabetes mellitus patients. (Ref. Reference Zhang, Li and Jiang101) In CDiPER, the target miRNA binds to hairpin probes (H/H2 probe), inducing conformational changes to form a DNAzyme. DNAzymes are single-stranded DNA molecules with catalytic function, and here, they cleave the double-loop probe (D probe), which activates the primer exchange reaction (PER), whereby primers are exchanged to amplify the fluorescence signals [Figure 6D].

Yao (Ref. Reference Yao, Liu and Lu42) reported a cascade-based multiplexed exponential isothermal amplification reaction (MEXPAR) to simultaneously detect multiple urological target miRNAs to diagnose prostate, bladder, and renal cell cancers. (Ref. Reference Yao, Liu and Lu42) The miRNAs undergo four parallel EXPARs, utilising Bst DNA polymerase and nicking endonucleases to release single-strand DNA (ssDNA), which is hybridised with microgels functionalised with fluorescence-labelled DNA probes [Figure 6E].

CRISPR/Cas-assisted isothermal amplification

Innovative miRNA detection systems integrating isothermal amplification and CRISPR technology have effectively distinguished various diseases from clinical samples. CRISPR, or clustered regularly interspaced short palindromic repeats, is a precise gene-editing tool that efficiently modifies targeted DNA sequences and has been successfully adapted for disease detection applications. (Ref. Reference Jinek, Chylinski and Fonfara102) These mechanisms integrate CRISPR-associated (Cas) enzymes, guided by synthetic RNA, to locate and cleave specific DNA sequences. (Ref. Reference Banakar, Schubert and Collingwood103) This enables targeted genetic modifications through insertions, deletions, or alterations. (Ref. Reference Banakar, Schubert and Collingwood103)

Liu (Ref. Reference Liu, Zhao and Yuan104) presented a combination of RCA and CRISPR/Cas9 for miRNA detection. (Ref. Reference Liu, Zhao and Yuan104) In this method, the target miRNA hybridises with the ends of a dumbbell probe, forming a cyclised padlock structure. This padlock then undergoes RCA, facilitated by phi29 DNA polymerase, to produce ssDNA with repeated hairpin structures. The ssDNA is hybridised with the CRISPR/Cas9 recognition site, which the CRISPR/Cas9 system cleaves, disassembling the RCA product into hairpin probes (H1 probes). The target miRNA then unfolds the H1 probes, initiating a catalytic hairpin assembly (CHA) process that generates a detectable fluorescence signal through binding with a fluorescence-labelled H2 probe [Figure 7A] (Ref. Reference Liu, Zhao and Yuan104).

Figure 7. Schematic illustration of isothermal amplification methods with CRISPR/Cas systems for detecting miRNA. A. Rolling circle amplification (RCA)-CRISPR/Cas9; (Ref. Reference Liu, Zhao and Yuan104) B. exponential amplification reaction (EXPAR)-CRISPR/Cas12a; (Ref. Reference Yang, Yang and Gong105) C. Strand displacement amplification (SDA)-CRISPR/Cas14a; (Ref. Reference Chi, Wu and Chen107) and D. Target-triggered SDA (ttSDA)-CRISPR/Cas12a + Pt@Au nanozymes. (Ref. Reference Luo, Zhou and Zhan108)

Similarly, Yang (Ref. Reference Yang, Yang and Gong105) used EXPAR and CRISPR/Cas12a to detect miRNAs. Yang (Ref. Reference Yang, Yang and Gong105) reported that CRISPR/Cas12a has a more sensitive detection profile when coupled with EXPAR, as opposed to using CRISPR/Cas12a as a standalone assay. (Ref. Reference Yang, Yang and Gong105) The target miRNA has a complementary template strand containing binding regions for an endonuclease-nicking enzyme and a CRISPR/Cas12a activator. With vent(exo-) polymerase, the EXPAR releases activators that bind to Cas12a-CRISPR RNA (crRNA). The volume of the activator that is repeatedly produced from the amplification reaction prompts Cas12a to trans-cleave a quenched fluorescence reporter probe, leading to an amplified fluorescent signal [Figure 7B] (Ref. Reference Yang, Yang and Gong105).

Strand displacement amplification (SDA) is an isothermal technique that uses four primers to initiate a cyclic reaction, in addition to restriction enzymes and endonucleases, to target recognition and cleavage sites. (Ref. Reference Walker, Fraiser and Schram106) DNA polymerase binds at the nicked site, allowing strand extension and displacement, leading to exponential amplification. (Ref. Reference Walker, Fraiser and Schram106) Chi (Ref. Reference Chi, Wu and Chen107) used this technique to detect circulating miRNA from serum by integrating it with CRISPR/Cas14a. (Ref. Reference Chi, Wu and Chen107) Here, a template DNA binds with the target miRNA at its recognition site, and an endonuclease nicks the Cas14a activation site of the template DNA, releasing activators that bind to the CRISPR/Cas14a-single guide RNA (sgRNA) and allow Cas14a to trans-cleave quenched fluorescence reporter probes, amplifying the fluorescent signal [Figure 7C] (Ref. Reference Chi, Wu and Chen107).

Luo (Ref. Reference Luo, Zhou and Zhan108) used a similar method of SDA and a nanozyme-mediated CRISPR-Cas12a system for miRNA detection in hepatocellular carcinoma patients. (Ref. Reference Luo, Zhou and Zhan108) This method employed toehold-triggered SDA, which bypasses the need for enzymes. Here, the target miRNA and assisting single-stranded DNA4 (S4) hybridise with exposed toehold regions on custom magnetic beads, triggering the SDA reaction and releasing DNA complementary to the crRNA, activating the CRISPR-Cas12a system. The active Cas12a cleaves ssDNA linkers, releasing Pt@Au nanozymes from magnetic beads. The released nanozymes, which possess peroxidase-like activity, catalyse 3,3′,5,5′-tetramethylbenzidine (TMB) oxidation by hydrogen peroxide (H2O2). This reaction causes a colour change that can be visually observed or measured using an ultraviolet–visible spectrophotometer, where the intensity of the colour change positively correlates with the concentration of the target miRNA [Figure 7D] (Ref. Reference Luo, Zhou and Zhan108).

Furthermore, alternative CRISPR-based systems for miRNA detection, particularly those utilising Cas13, have been demonstrated. Notably, the CRISPR/Cas13a-based workflows were capable of ultra-sensitive miRNA detection (4.5aM—2.6fM), exhibiting impressively low detection limits when tested with miRNA spiked in human liquid biopsy samples. (Refs Reference Cui, Fan and Yuan109, Reference Zhou, Huang and Lin110, Reference Shan, Zhou and Huang111) CRISPR/Cas amplification systems are also being explored, along with innovative light-activation mechanisms to enhance signal amplification. A recent study demonstrated a light-activated CRISPR/Cas12a amplification system that successfully detected miRNA21 in MCF-7 cancer cell lines (Ref. Reference Liu, Dong and Duan112).

Isothermal amplification methods using nanomaterials

Various miRNA detection systems that incorporate isothermal amplification methods with nanomaterials to modify or amplify the reaction have been developed. One example can be found in the research of Wei (Ref. Reference Wei, He and Mao113), whose research used hybridisation chain reaction (HCR) with molybdenum disulfide (MoS2) nanosheets in plasma-derived exosomes from lung cancer patients. (Ref. Reference Wei, He and Mao113) HCR uses specifically designed DNA hairpins to amplify target-specific signals without enzymes. (Refs Reference Choi, Schwarzkopf and Fornace114,Reference Xu, Lai and Wilson115) The process starts with the adsorption of fluorescently-labelled DNA probes (FAM-H1 and FAM-H2) onto the MoS2 nanosheets, resulting in fluorescence quenching. The presence of target miRNA initiates the HCR by binding to FAM-H1, triggering a cascade of hybridisation events between FAM-H1 and FAM-H2. This forms long dsDNA products that detach from the nanosheets, leading to amplified signals and fluorescence detection [Figure 8A] (Ref. Reference Wei, He and Mao113).

Figure 8. Schematic illustration of isothermal amplification methods with nanomaterials for detecting miRNA. A. Hybridisation chain reaction (HCR) + Molybdenum disulfide (MoS2) nanosheets; (Ref. Reference Wei, He and Mao113) B. HCR + magnetic covalent organic framework (COF) nanospheres; (Ref. Reference Liang, Zhang and Wang116) C. Catalytic hairpin assembly reaction (CHA) + HCR + Au-Ag nanoparticles; (Ref. Reference Chen, Chen and Kong117) and D. CHA + enzyme-assisted signal amplification (EASA) + Au–Ag nanoshuttles. (Ref. Reference Dai, Wang and Tan72)

Similarly, Liang (Ref. Reference Liang, Zhang and Wang116) used HCR with magnetic covalent organic framework (COF) nanospheres to detect miRNA in glioma patients. (Ref. Reference Liang, Zhang and Wang116) The nanospheres consist of Fe3O4 nanoparticles and high-crystalline COF shells. When target miRNA is absent, the two hairpin DNA probes, H1 and H2, are adsorbed onto the nanospheres, resulting in quenched fluorescence. However, when miRNA is present, the hairpin probes hybridise with the miRNA, creating non-adsorbable dsDNA, thus blocking the fluorescence quenching [Figure 8B]. Therefore, the change in fluorescence intensity reflects the target miRNA concentration. (Ref. Reference Liang, Zhang and Wang116)

Chen (Ref. Reference Chen, Chen and Kong117) developed a surface-enhanced Raman spectroscopy (SERS) biosensor to detect miR-210 in blood erythrocytes using two signal amplification strategies: CHA and HCR. The system employs four Raman reporter-labelled probes. (Ref. Reference Chen, Chen and Kong117) When the target miRNA binds to the C1’ probe, the C2’ probe binds and subsequently releases the miRNA. With the addition of Au@Ag nanoparticles and H1 & H2 probes, HCR is triggered, causing consecutive and continuous binding of H1/H2 probes onto the sticky ends of the C1’/C2’ double-strand. The Au@Ag nanoparticles cause aggregation of Raman reporters, generating a detectable SERS signal [Figure 8C] (Ref. Reference Chen, Chen and Kong117).

Dai (Ref. Reference Dai, Wang and Tan72) introduced a highly sensitive version of CHA combined with enzyme-assisted signal amplification (EASA). (Ref. Reference Dai, Wang and Tan72) In EASA, the target miRNA binds to the looped section of the hairpin DNA1 (hpDNA1) probe, which contains an endonuclease recognition site. The endonuclease cleaves the hairpins into DNA1 and DNA2, leaving the target miRNA intact for repeated EASA cycles. DNA2 binds to a hairpin DNA2–1 (hpDNA2–1) probe labelled with Au–Ag nanoshuttles, serving as the SERS nanoprobes. This structure further hybridises with a hairpin DNA3 (hpDNA3) probe labelled with magnetic beads, which act as capture nanoprobes. The magnets allow aggregation of the complexes, further amplifying the SERS signal [Figure 8D] (Ref. Reference Dai, Wang and Tan72).

Isothermal amplification methods using metal–organic frameworks

Metal–organic frameworks (MOFs) are increasingly integrated into miRNA detection systems to enhance isothermal amplification strategies. This novel class of self-assembling crystalline materials offers unique advantages. (Ref. Reference Mahata, Roy, Mellot-Draznieks, Xu, Gao and Chen118) Their popularity in miRNA detection stems from their adjustable porosity, expansive surface area, customisable chemistry, thermal resilience, and luminescent properties. (Ref. Reference Cui, Zhang and He119)

Sun (Ref. Reference Sun, Li and Tong120) developed a miRNA detection method using MOF-525, a self-fluorescent framework, and target-triggered RCA for colorectal cancer patients from serum exosomes. (Ref. Reference Sun, Li and Tong120) Here, the target miRNA hybridises with a template strand, initiating RCA via phi29 DNA polymerase. The amplified product then forms dsDNA with fluorescent reporter probes, preventing their adsorption and quenching by MOF-525. This results in a change in the fluorescence ratio that correlates positively with the target miRNA concentration, enabling sensitive and specific detection [Figure 9A] (Ref. Reference Sun, Li and Tong120).

Figure 9. Schematic illustration of isothermal amplification methods with metal–organic frameworks (MOFs) for detecting miRNA. A. Rolling circle amplification (RCA) + MOF-525; (Ref. Reference Sun, Li and Tong120) B. Primer exchange reaction (PER) + MOF@Pt@MOF nanozyme; (Ref. Reference Li, Li and Li121) C. Rolling hoop orbital amplification (RHOA) + Iron-Zirconium (Fe-Zr) MOF/G-quadruplex (G4) nanozyme; (Ref. Reference Gao, Wu and Li122) and D. λ-Exonuclease + ZIF-67 MOF/ Tungsten(VI) oxide (WO3) nanoflakes. (Ref. Reference Wang, Yang and Shi123)

Li (Ref. Reference Li, Li and Li121) successfully integrated isothermal amplification with MOFs and nanomaterials, developing a PER-combined MOF@Pt@MOF nanozyme. (Ref. Reference Li, Li and Li121) Nanozymes are nanoparticles with enzyme-like catalytic properties. (Ref. Reference Li, Li and Li121) The MOF nanozyme consists of Pt nanoparticles sandwiched between MIL-88 and coated with ssDNA. It utilises a gated hairpin that activates upon hybridisation with target miRNA, splitting into catalytic and protector A probe-miRNA complexes. The catalytic hairpin initiates PER with primers and KF DNA polymerase, releasing ssDNA. This ssDNA binds to a protector B probe on an Au electrode functionalised with 6-mercapto-1-hexanol (MCH), displacing multiple protector B probes. The nanozyme’s ssDNA then hybridises with the capture probe on the electrode, catalysing TMB oxidation by H2O2 and generating an amplified electrochemical signal [Figure 9B] (Ref. Reference Li, Li and Li121).

Another notable isothermal amplification technique is rolling hoop orbital amplification (RHOA) - a technique used by Gao (Ref. Reference Gao, Wu and Li122). Here, RHOA was combined with an iron-zirconium (Fe-Zr) MOF/G-quadruplex (G4) nanozyme with peroxidase activity. When the target miRNA binds to the RHOA design, the multiplexed nucleic acid enzyme (MNAzyme) is displaced by sequentially opening circular DNA probes (Ra and Rb). The process is initiated as the MNAzyme amplifies and cleaves the Ra probe in the presence of magnesium ions (Mg2+), which is repeated with the Rb probe. As a result, many Ra and Rb probes bind to the Fe-Zr MOF/G4 nanozyme, which is hybridised onto a glassy carbon electrode with copper (I) oxide (Cu2O)/Au nanocubes. After adding glucose to the system, oxidation by Cu2O/Au generates H2O2 and glucuronic acid. The MOF catalyses the decomposition of H2O2, which in turn reacts with luminol, producing a detectable signal via electrochemiluminescence [Figure 9C] (Ref. Reference Gao, Wu and Li122).

Platform-based isothermal amplification assisted by λ-exonuclease (λ-exo) has also been researched by Wang (Ref. Reference Wang, Yang and Shi123) for the detection of exosomal miRNA. (Ref. Reference Wang, Yang and Shi123) Here, mesoporous silica nanoparticles (MSNs) trap hemin nanocarriers, which quench photocurrent upon introducing the target miRNA. The release of hemin is prompted by the displacement of miRNA from the DNA hybrid on the MSN surface. A novel ZIF-67 MOF decorated with tungsten (VI) oxide (WO3) nanoflakes is utilised to enhance the photoresponse during the reaction by facilitating charge separation correlated with the abundance of target miRNA. With hemin, the electrons from the MOF are transferred to the WO3 nanoflakes, using ascorbic acid as the electron donor [Figure 9D] (Ref. Reference Wang, Yang and Shi123). The oxidised products can then be detected electrochemically.

Hybridisation

A significant advance in miRNA detection, specifically towards POC testing, has been the development of detection platforms that bypass signal amplification. Several different amplification-free hybridisation methods have been successfully validated with clinical patient samples.

Nanomaterial biosensors

Previously, nanomaterials were described in conjunction with isothermal amplification. However, these materials have also been applied successfully in amplification-free methods. For example, Wang (Ref. Reference Wang, Lu and Tang70) used gold-coated magnetic microbead nanoparticles (AuNP-MMBs) with diblock oligonucleotide (ODN)-modified AuNPs to detect serum-derived miRNAs in glioma patients. (Ref. Reference Wang, Lu and Tang70) Two target miRNAs were hybridised on two respective hairpin probes on the surface of AuNP-MMBs, allowing attachment of the diblock-ODN modified AuNPs to the end of the hairpin probe [Figure 10A]. The abundance of target miRNA was quantified by measuring the oxidation peak currents of methylene blue (MB) and Ferrocene (Fc) moieties present on the diblock-ODN modified AuNPs. (Ref. Reference Wang, Lu and Tang70)

Figure 10. Schematic illustration of electrochemical and electrical nanoparticle miRNA detection methods. A. Magnetic microbeads & diblock oligonucleotide-modified Au nanoparticles; (Ref. Reference Wang, Lu and Tang70) B. Peptide nucleic acid (PNA) functionalised nanochannel biosensor; (Ref. Reference Xiao, Wan and Liao124) C. DNA tetrahedral nanostructure-based biosensor; (Ref. Reference Zeng, Wang and Meng125) D. Carbon nanotube biosensor; (Ref. Reference Li, Liang and Li126) and E. Phosphorodiamidate morpholino oligomers (PMO) - graphene quantum dots (GQDs) - functionalised reduced graphene oxide (RGO) field effect transistor (FET) biosensor. (Ref. Reference Li, Tu and Zhang127)

Xiao (Ref. Reference Xiao, Wan and Liao124) used a unique biosensing technique to detect miRNAs associated with pancreatic cancer. (Ref. Reference Xiao, Wan and Liao124) They employed a peptide nucleic acid (PNA) functionalised nanochannel to detect exosomal miR-10b. The PNA was covalently bound to nanochannels, where extracted exosomal miRNAs were hybridised to the surface PNAs. Due to the negatively charged phosphate backbone of miRNA, the ionic current was measured and correlated to the abundance of miRNA in the sample [Figure 10B] (Ref. Reference Xiao, Wan and Liao124).

Zeng (Ref. Reference Zeng, Wang and Meng125) reported multiplexed simultaneous detection of target miRNAs associated with overexpression in pancreatic carcinoma patients using a nanostructure-based electrochemical biosensor. (Ref. Reference Zeng, Wang and Meng125) DNA tetrahedral nanostructures are bound with a capture probe on an Au electrode. Subsequently, the target miRNA hybridises with the capture probe on the nanostructure, and a biotin-tagged signal probe facilitates the binding of streptavidin poly-HRP40, a biotin-binding protein. This catalyses the reduction of H2O2 in the presence of the TMB substrate, generating a measurable amperometric signal [Figure 10C] (Ref. Reference Zeng, Wang and Meng125).

Li (Ref. Reference Li, Liang and Li126) have demonstrated the use of field-effect transistor (FET) biosensors based on a carbon nanotube (CNT) film for detecting exosomal miRNA in breast cancer patients with a highly sensitive detection limit of 0.87 aM. (Ref. Reference Li, Liang and Li126) Here, thiolated DNA probes (HS-DNA) are immobilised on the surface of AuNPs. When the DNA probe hybridises with the target miRNA, it introduces additional negative charges at the sensing interface, leading to p-type electrostatic doping, measured by the FET sensor [Figure 110D].

Li (Ref. Reference Li, Tu and Zhang127) also demonstrated a similar case of exosomal miRNA detection in breast cancer patients using surface probe hybridisation and electrical detection. (Ref. Reference Li, Tu and Zhang127) This technique used a phosphorodiamidate morpholino oligomers-graphene quantum dots (PMO-GQD) complex covalently bound to a reduced graphene oxide (RGO) FET [Figure 10E]. Here, the miRNA changes the net carrier density, which is similarly measured by the FET sensor.

Afzalinia and Mirzaee (Ref. Reference Afzalinia and Mirzaee128) used a sandwich-type hybridisation biosensor to detect lung and breast cancer from plasma samples. (Ref. Reference Afzalinia and Mirzaee128) This study used La(III)-MOF and Ag nanoparticles to hybridise with the target miRNA. Upon hybridisation, the La(III)-MOF and Ag nanoparticles function as energy donor-acceptor pairs in the fluorescence resonance energy transfer (FRET) process. The presence of the target miRNA induces FRET, leading to the quenching of the fluorescent signal [Figure 11A] (Ref. Reference Afzalinia and Mirzaee128).

Figure 11. Schematic illustration of fluorescent, plasmonic, and colourimetric nanoparticle miRNA detection methods. A. Lanthanum oxide (La(III))-metal–organic framework (MOF) & silver (Ag) nanoparticle biosensor; (Ref. Reference Afzalinia and Mirzaee128) B. Iron(II,III) oxide@Titanium oxide (Fe3O4@TiO2) nanoparticle biosensor; (Ref. Reference Jiang, Li and Wang129) C. Gold (Au) nanoprism biosensor;(Ref. Reference Joshi, Deitz-McElyea and Liyanage71) and D. Gold (Au) nanoparticle liposome-based microfluidic chip biosensor. (Ref. Reference Zeng, Wang and Liu130)

Jiang (Ref. Reference Jiang, Li and Wang129) utilised a locked nucleic acid (LNA)-modified Au@DTNB (Raman reporter molecule) SERS tags for epithelial ovarian cancer detection. (Ref. Reference Jiang, Li and Wang129) The LNA provides increased stability during high temperatures and consistent affinity for the target miRNA. The Au@DTNB SERS tags enter the exosomes and hybridise with the target miRNA. Subsequently, magnetic Fe3O4@TiO2 nanoparticles capture the SERS tags, enriching the exosomes and further amplifying the SERS signal [Figure 11B] (Ref. Reference Jiang, Li and Wang129).

Joshi (Ref. Reference Joshi, Deitz-McElyea and Liyanage71) demonstrated a label-free detection of miRNA by utilising an ultrasensitive localised surface plasmon resonance (LSPR)-based sensor validated with pancreatic ductal adenocarcinoma and chronic pancreatitis patient plasma samples. (Ref. Reference Joshi, Deitz-McElyea and Liyanage71) Target miRNA complementary ssDNA probes were immobilised on Au nanoprisms. Here, the hybridisation of target miRNA with the ssDNA probes increases the local refractive index around the nanoprisms, resulting in a shift of the LSPR peak wavelength. The LSPR-based sensor was highly specific, distinguishing miR-10b from its single nucleotide variant, miR-10a [Figure 11C] (Ref. Reference Joshi, Deitz-McElyea and Liyanage71).

Zeng (Ref. Reference Zeng, Wang and Liu130) tested a liposome-based microfluidic platform to detect miRNA differences in type II diabetic patients. (Ref. Reference Zeng, Wang and Liu130) This biosensor immobilised Au nanoparticles on a glass surface, which hybridised with a DNA probe modified with thiol groups and biotin. The DNA probe then immobilises biotin-modified liposomes encapsulating glucose oxidase. Following hybridisation with the target miRNA, a duplex-specific nuclease (DSN) re-releases the liposomes. The contents of the liposome are released, causing a colour change due to H2O2 decomposition and TMB oxidation, catalysed by G4/hemin [Figure 11D] (Ref. Reference Zeng, Wang and Liu130).

Microarray

A microarray is a unique multiplexing technique for miRNA detection. However, it has a considerably lower throughput than the RNA sequencing detection method but a higher throughput than the typical qPCR method. (Ref. Reference Novogene131) In microarrays, the target miRNA is transcribed to cDNA and labelled with fluorescent markers. These signature cDNA strands hybridise onto probes on the microarray surface, creating a cluster of hybridised samples that fluoresce when scanned, leading to a distinct and detectable array [Figure 12A]. Researchers Zhang and Hu (Ref. Reference Zhang and Hu132) used a microarray slide that can target an extensive list of miRNAs: a 3D-gene human miRNA oligo chip. (Ref. Reference Zhang and Hu132) This microarray was used for early detection of various cancers, targeting 50 different signature miRNAs. The study reported detection sensitivity >99% in >1,000 serum samples from patients with stage 1 lung, biliary tract, bladder, colorectal, oesophagal, gastric, glioma, liver, pancreatic, sarcoma, breast and prostate cancers. (Ref. Reference Zhang and Hu132)

Figure 12. Schematic illustration of various miRNA detection methods. A. Microarray, (Ref. Reference Zhang and Hu132) B. nCounter, (Refs Reference Shankar, Shetty and N.S43, Reference Niemira, Erol and Bielska134) C. RNA sequencing, (Ref. Reference Geng, Tsou and Stass137) and D. Capillary electrophoresis. (Ref. Reference Ban, Lim and Kwon145)

nCounter

The nCounter (Nanostring) system can directly detect miRNA from the isolation step in a multiplex fashion. Like microarrays, it immobilises target miRNAs on a treated surface to identify multiple targets simultaneously with unique codes. However, unique to nCounter is the ligation of the target miRNA to miRtags and a bridge oligonucleotide sequence. The resulting complex hybridises with a ligand (biotin) for immobilisation and a target-specific barcode that generates a unique sequence of fluorophores. The complex is immobilised on the surface and quantified for each signature miRNA complex using Nanostring’s Digital Analyser [Figure 12B] (Ref. Reference Malkov, Serikawa and Balantac133). Niemira (Ref. Reference Niemira, Erol and Bielska134) successfully used nCounter to detect two miRNAs in serum from high-grade serous ovarian cancer patients. (Ref. Reference Niemira, Erol and Bielska134) Additionally, Shankar (Ref. Reference Shankar, Shetty and N.S43) reported successful miRNA detection in urinary exosomal miRNA expression levels in patients with IgA nephropathy with 100% sensitivity and specificity. (Ref. Reference Shankar, Shetty and N.S43)

RNA Sequencing

Next-generation sequencing (NGS) is widely used for large-scale nucleotide sequencing and diagnostics. RNA sequencing identifies and quantifies RNA molecules in a sample, offering a comprehensive view of the transcriptome and enabling the detection and measurement of various RNA forms. (Ref. Reference Marguerat and Bähler135) NGS has benefits such as low cost at pooled samples, multiplexed testing, and higher throughput. Still, it has notable drawbacks, including extensive sample preparation requirements compared to other nucleic acid detection methods and unsuitability for POC testing. (Ref. Reference Precazzini, Detassis and Imperatori5,Reference Khodakov, Wang and Zhang136)

Geng (Ref. Reference Geng, Tsou and Stass137) used the MiSeqDx (Illumina) NGS instrument to detect multiple miRNAs simultaneously in adenocarcinoma and squamous cell carcinoma patients from plasma samples. (Ref. Reference Geng, Tsou and Stass137) In this method, small RNA molecules, including miRNAs, are isolated from a sample. These RNAs are ligated with 3′ and 5′ adapters and converted into cDNA. The cDNA is then amplified via PCR and purified, after which the cDNA libraries are sequenced. This allows for the identification and quantification of miRNAs [Figure 12C] (Ref. Reference Geng, Tsou and Stass137).

Capillary Electrophoresis

Capillary electrophoresis (CE) is a common technique that separates ions by their movement under an applied voltage in tiny capillaries of micro- and nano-fluidic channels. (Ref. Reference Li138) CE is often utilised to separate and analyse proteins, peptides, and nucleic acids. (Refs Reference Claire139Reference Anazawa, Takahashi and Kambara144) Ban (Ref. Reference Ban, Lim and Kwon145) developed a novel method combining magnetic bead-based CE with laser-induced fluorescence (CE-LIF) to detect miRNAs in lung cancer patients’ plasma samples successfully. (Ref. Reference Ban, Lim and Kwon145) This method used streptavidin-coated magnetic beads that hybridise with biotinylated fluorescent DNA probes (6-FAM-biotin-labelled probes) complementary to the target miRNA. An RNase inhibitor was added to reduce miRNA degradation by RNase in the plasma sample. After hybridisation, the miRNA-probe complexes are eluted by heating. The eluted complexes are then directly injected into the CE-LIF system for analysis [Figure 12D].

Conclusion and Future Perspectives

The field of miRNA detection using liquid biopsies is rapidly evolving. Employing liquid biopsies minimises sample collection’s invasiveness and accelerates progress toward POC miRNA testing systems. These advancements are highly significant given the crucial role miRNAs play in gene regulation and disease progression. This analysis reviewed numerous miRNA workflows, detailing each step from sample collection and RNA isolation to amplification, processing, and detection methods. This review found that miRNAs can be effectively detected in clinical samples from blood, urine, and saliva - with blood being the most commonly used sample type. While some RNA isolation methods perform exceptionally well, they are unsuitable for high-throughput or POC testing, which are important considerations for advancing miRNA testing systems. Importantly, three studies successfully bypassed RNA isolation, achieving detection limits of 4.26 aM to 0.21fM, demonstrating the potential for future development and use. (Refs Reference Wang, Lu and Tang70, Reference Joshi, Deitz-McElyea and Liyanage71, Reference Dai, Wang and Tan72) In miRNA sample processing, many innovative techniques have been developed and successfully implemented. One example is SMOS-qPCR, which achieved the lowest reported LOD of 0.1 zM. (Ref. Reference Kang, Zhang and Liu85) However, as PCR methods are unsuitable for POC testing, ongoing advancements in alternative miRNA processing techniques are crucial. Thus, the wide range of isothermal amplification methods were detailed, either as standalone approaches or in combination with CRISPR/Cas systems, nanoparticles, or MOFs. Notably, processing methods that bypass amplification entirely - despite the challenge posed by the low abundance of miRNA in samples - further advance POC testing systems. Several amplification-free, hybridisation-based techniques demonstrated exceptionally low detection limits, with the lowest being 0.87 aM, as reported by Li (Ref. Reference Li, Liang and Li126), using a carbon-nanotube FET. (Ref. Reference Li, Liang and Li126) The emerging development of POC-compatible testing systems also holds significant potential for broader applications, including acute infectious disease monitoring and chronic disease management. (Ref. Reference Larkins and Thombare146)

In addition to miRNA, several other blood-based biomarkers are being explored for cancer detection. For instance, protein biomarkers have well-established assays but often lack the sensitivity and specificity needed for precise detection. (Ref. Reference Duffy, McDermott and Crown147) Additionally, circulating tumour DNA (ctDNA) enables mutation-specific monitoring and resistance profiling, though it requires timely sampling, is labour-intensive, and is costly. (Refs Reference Duffy, McDermott and Crown147,Reference Das, Dey and Devireddy148) Lastly, circulating tumour cells (CTC) show promise in early diagnosis and monitoring response to therapy, yet are also expensive and labour-intensive, and results can be impacted by CTC undergoing epithelial-mesenchymal transition. (Refs Reference Duffy, McDermott and Crown147,Reference Das, Dey and Devireddy148) Compared to other methods, in the landscape of liquid biopsy for cancer detection, miRNA has shown considerable promise. Its stability in blood, role as a gene regulator, and established detection and processing methods have contributed to its advancement in clinical applications. miRNA’s unique characteristics position it as a potentially valuable tool for early diagnosis and monitoring, though further research and validation are still needed to understand its clinical potential fully.

Given the role of miRNA dysregulation in cancer development, miRNAs are also being investigated for therapeutic applications. For example, instances of anti-microRNA formulations designed to block pathways associated with dysregulated miRNAs are also being explored by Yan et al (2023). (Ref. Reference Yan, Su and Hsia149) Although current methods are limited by challenges like off-target effects and low cellular uptake, innovative strategies are still being explored to overcome these limitations. (Ref. Reference Pagoni, Cava and Sideris150)

This review discusses and highlights novel workflows for miRNA detection from liquid biopsies. The reviewed papers have validated their designs using clinical samples. However, many clinical studies were limited to small, controlled sample sizes, necessitating future validation in diverse, large patient cohorts.

A significant advancement in the field worth mentioning is the initiation of multiple clinical trials to detect miR-371a-3p in peripheral blood to diagnose stage 1 testicular germ cell cancer. For instance, one trial has progressed to enrolling patients for an observational cohort study across a spectrum of risk levels at various North American hospitals. (Refs Reference Conduit, Grimison and Lewin151153) The detection of miR-371 in patient samples was successful both with RT-qPCR and ddPCR, although RT-qPCR detected the target miRNA with higher sensitivity than ddPCR. (Ref. Reference Nuno, Lafin and Scarpini94)

The integration of miRNA-based diagnostic workflows into clinical practices demands the establishment of standardised protocols with normalisation factors. Often, these protocols should employ stable expression of reference miRNA across sample types and abide by strict benchmarks for target miRNA quantification. Hence, robust normalisation processes for endogenous genes or spike-in controls are necessary for comparing results and providing an accurate diagnosis of the pathophysiological state of the patient from miRNA for precise clinical intervention. The interpretation of the results, therefore, should also incorporate a grey zone, representing values that fall outside expected levels, to account for minor abnormalities in miRNA levels that do not require immediate clinical intervention intervention. (Refs Reference Faraldi, Gomarasca and Sansoni14,Reference Peltier and Latham154)

Another challenge in developing miRNA detection systems is the significant cost variation of custom or off-the-shelf consumables, reagents, and instrumentation, highlighting the need for more cost-effective solutions to enhance the accessibility and adaptability of these technologies. (Ref. Reference Ye, Xu and Tian155) Ongoing advancements to address challenges like high costs, limited sensitivity, and insufficient detection limits in novel miRNA detection methods are expected to pave the way for more reliable and accessible POC platforms. These developments are anticipated to enhance early disease diagnostics significantly.

Acknowledgements

All figures were created with BioRender.com.

Author Contribution

D.N.: investigation and writing. E.H.: investigation and writing. N.T.: review and editing. A.T.: review, editing, and project administration.

Funding statement

This work was supported by Brown University and Revvity Inc., Waltham, MA, USA.

Competing interest

Dulguunnaran Naranbat, Emilia Herdes, Nikos Tapinos, and Anubhav Tripathi declare none.

Footnotes

*

D.N. and E.H. contributed equally to this work.

References

Hejazi, M, Baghbani, E, Amini, M, et al. (2020) MicroRNA-193a and taxol combination: A new strategy for treatment of colorectal cancer. Journal of Cellular Biochemistry 121, 13881399.Google Scholar
Asl, ER, Amini, M, Najafi, S, et al. (2021) Interplay between MAPK/ERK signaling pathway and MicroRNAs: A crucial mechanism regulating cancer cell metabolism and tumor progression. Life Sciences 278, 119499.Google Scholar
Zhang, W, Dahlberg, JE and Tam, W (2007) MicroRNAs in tumorigenesis. The American Journal of Pathology 171, 728738.Google Scholar
Paranjape, T, Slack, FJ and Weidhaas, JB (2009) MicroRNAs: tools for cancer diagnostics. Gut 58, 15461554.Google Scholar
Precazzini, F, Detassis, S, Imperatori, AS, et al. (2021) Measurements methods for the development of microRNA-based tests for cancer diagnosis. International Journal of Molecular Sciences 22, 1176.Google Scholar
Bushati, N and Cohen, SM (2007) MicroRNA functions. Annual Review of Cell and Developmental Biology 23, 175205.Google Scholar
Kim, T and Croce, CM (2023) MicroRNA: trends in clinical trials of cancer diagnosis and therapy strategies. Experimental & Molecular Medicine 55, 13141321.Google Scholar
Gastman, B, Agarwal, PK, Berger, A, et al. (2020) Defining best practices for tissue procurement in immuno-oncology clinical trials: consensus statement from the society for immunotherapy of cancer surgery committee. Journal for Immunotherapy of Cancer 8, e001583.Google Scholar
Hodel, S, Laux, C, Farei-Campagna, J, et al. (2018) The impact of biopsy sampling errors and the quality of surgical margins on local recurrence and survival in chondrosarcoma. Cancer Management and Research 10, 37653771.Google Scholar
Crowley, E, Di Nicolantonio, F, Loupakis, F, et al. (2013) Liquid biopsy: monitoring cancer-genetics in the blood. Nature Reviews Clinical Oncology 10, 472484.Google Scholar
Robertson, EG and Baxter, G (2011) Tumour seeding following percutaneous needle biopsy: The real story! Clinical Radiology 66, 10071014.Google Scholar
Wang, Z-Y, Sun, M-H, Zhang, Q, et al. (2023) Advances in point-of-care testing of microRNAs based on portable instruments and visual detection. Biosensors 13, 747.Google Scholar
Felekkis, K, Pieri, M and Papaneophytou, C (2023) Exploring the feasibility of circulating miRNAs as diagnostic and prognostic biomarkers in osteoarthritis: Challenges and opportunities. International Journal of Molecular Sciences 24, 13144.Google Scholar
Faraldi, M, Gomarasca, M, Sansoni, V, et al. (2019) Normalization strategies differently affect circulating miRNA profile associated with the training status. Scientific Reports 9, 1584.Google Scholar
Chen, G, Wang, J and Cui, Q (2013) Could circulating miRNAs contribute to cancer therapy? Trends in Molecular Medicine 19, 7173.Google Scholar
Srinivasan, S, Yeri, A, Cheah, PS, et al. (2019) Small RNA sequencing across diverse biofluids identifies optimal methods for exRNA isolation. Cell 177, 446462.e16.Google Scholar
Liu, S, da Cunha, AP, Rezende, RM, et al. (2016) The host shapes the gut microbiota via fecal microRNA. Cell host & microbe 19, 3243.Google Scholar
Grimaldi, AM, Lapucci, C, Salvatore, M, et al. (2022) Urinary miRNAs as a diagnostic tool for bladder cancer: A systematic review. Biomedicines 10, 2766.Google Scholar
Balakittnen, J, Ekanayake Weeramange, C, Wallace, DF, et al. (2024) A novel saliva-based miRNA profile to diagnose and predict oral cancer. International Journal of Oral Science 16, 113.Google Scholar
El-Mogy, M, Lam, B, Haj-Ahmad, TA, et al. (2018) Diversity and signature of small RNA in different bodily fluids using next generation sequencing. BMC Genomics 19, 408.Google Scholar
Sanz-Rubio, D, Martin-Burriel, I, Gil, A, et al. (2018) Stability of circulating exosomal miRNAs in healthy subjects. Scientific Reports 8, 10306.Google Scholar
Takizawa, S, Matsuzaki, J and Ochiya, T Circulating microRNAs: Challenges with their use as liquid biopsy biomarkers. Cancer Biomarkers 35, 19.Google Scholar
Endzeliņš, E, Berger, A, Melne, V, et al. (2017) Detection of circulating miRNAs: comparative analysis of extracellular vesicle-incorporated miRNAs and cell-free miRNAs in whole plasma of prostate cancer patients. BMC Cancer 17, 730.Google Scholar
Rieske, RR, Kutcher, ME, Audia, JP, et al. (2020) Analysis of plasma products for cellular contaminants: Comparing standard preparation methods. Journal of the American College of Surgeons 230, 596602.Google Scholar
Ding, H, An, Y, Zhao, T, et al. (2019) Large-scale rapid detection of circulating microRNAs in plasma for diagnosis and screening of specific diseases. Nanoscale 11, 1687916885.Google Scholar
Ban, E and Song, EJ (2022) Considerations and suggestions for the reliable analysis of miRNA in plasma using qRT-PCR. Genes 13, 328.Google Scholar
Sandau, US, Wiedrick, JT, McFarland, TJ, et al. (2024) Analysis of the longitudinal stability of human plasma miRNAs and implications for disease biomarkers. Scientific Reports 14, 2148.Google Scholar
Bracht, JWP, Los, M, van Eijndhoven, MAJ, et al. (2023) Platelet removal from human blood plasma improves detection of extracellular vesicle-associated miRNA. Journal of Extracellular Vesicles 12, 12302.Google Scholar
Jin, NZ and Gopinath, SCB (2016) Potential blood clotting factors and anticoagulants. Biomedicine & Pharmacotherapy 84, 356365.Google Scholar
Kaudewitz, D, Lee, R, Willeit, P, et al. (2017) Impact of intravenous heparin on quantification of circulating microRNAs in patients with coronary artery disease. Thrombosis and Haemostasis 110, 609615.Google Scholar
Mompeón, A, Ortega-Paz, L, Vidal-Gómez, X, et al. (2020) Disparate miRNA expression in serum and plasma of patients with acute myocardial infarction: a systematic and paired comparative analysis. Scientific Reports 10, 5373.Google Scholar
Ma, H-P, Fu, M, Masula, M, et al. (2022) miR-3064-5p and miR-4745-5p affect heparin sensitivity in patients undergoing cardiac surgery by regulating AT-III and factor X mRNA levels. Frontiers in Physiology 13, 914333.Google Scholar
Boileau, A, Lino Cardenas, CL, Lindsay, ME, et al. (2018) Endogenous heparin interferes with quantification of microRNAs by RT-qPCR. Clinical Chemistry 64, 863865.Google Scholar
Jaeschke, A, Harvey, NR, Tsurkan, M, et al. (2021) Techniques for RNA extraction from cells cultured in starPEG–heparin hydrogels. Open Biology 11, 200388.Google Scholar
Wakabayashi, I, Marumo, M, Ekawa, K, et al. (2024) Differences in serum and plasma levels of microRNAs and their time-course changes after blood collection. Practical Laboratory Medicine 39, e00376.Google Scholar
Ye, F, Feldman, DR, Valentino, A, et al. (2022) Analytical validation and performance characteristics of molecular serum biomarkers, miR-371a-3p and miR-372-3p, for male germ cell tumors, in a clinical laboratory setting. The Journal of Molecular Diagnostics 24, 867877.Google Scholar
Myklebust, MP, Rosenlund, B, Gjengstø, P, et al. (2019) Quantitative PCR measurement of miR-371a-3p and miR-372-p is influenced by hemolysis. Frontiers in Genetics 10, 463.Google Scholar
Wang, K, Yuan, Y, Cho, J-H, et al. (2012) Comparing the microRNA spectrum between serum and plasma. PLoS ONE 7, e41561.Google Scholar
Zhang, Z-W, Cheng, J, Xu, F, et al. (2011) Red blood cell extrudes nucleus and mitochondria against oxidative stress. IUBMB life 63, 560565.Google Scholar
Sun, L, Yu, Y, Niu, B, et al. (2020) Red blood cells as potential repositories of microRNAs in the circulatory system. Frontiers in Genetics 11.Google Scholar
Zhang, Y, Zhu, Y-Y, Chen, Y, et al. (2024) Urinary-derived extracellular vesicle microRNAs as non-invasive diagnostic biomarkers for early-stage renal cell carcinoma. Clinica Chimica Acta 552, 117672.Google Scholar
Yao, C, Liu, X, Lu, X, et al. (2024) Smartphone-based fluorescent profiling of quaternary microRNAs in urine for rapid diagnosis of urological cancers using a multiplexed isothermal exponential amplification reaction. Analytical Chemistry 96, 419426.Google Scholar
Shankar, M, Shetty, A, N.S, M, et al. (2023) Urinary exosomal miRNA signature of IgA nephropathy: a case–control study. Scientific Reports 13, 21400.Google Scholar
Zheng, J, Zhang, L, Johnson, M, et al. (2020) Comprehensive targeted metabolomic assay for urine analysis. Analytical Chemistry 92, 1062710634.Google Scholar
Wan, Y, Vagenas, D, Salazar, C, et al. (2017) Salivary miRNA panel to detect HPV-positive and HPV-negative head and neck cancer patients. Oncotarget 8, 99990100001.Google Scholar
Mamana, AC, Stincarelli, M, Sarmento, DJDS, et al. Detection of polyomavirus microRNA-5p expression in saliva shortly after kidney transplantation. Journal of Oral Microbiology 13, 1898838.Google Scholar
Marqués, M, Pont, M, Hidalgo, I, et al. (2023) MicroRNAs present in malignant pleural fluid increase the migration of normal mesothelial cells in vitro and may help discriminate between benign and malignant effusions. International Journal of Molecular Sciences 24, 14022.Google Scholar
Baraniskin, A, Kuhnhenn, J, Schlegel, U, et al. (2011) Identification of microRNAs in the cerebrospinal fluid as biomarker for the diagnosis of glioma. Neuro-Oncology 14, 29.Google Scholar
Jane, LA and Wray, AA (2024) Lumbar Puncture. StatPearls, StatPearls Publishing, Treasure Island (FL).Google Scholar
Farah, R, Sidar-Orlin, K and Sharabi-Nov, A (2024) Characterization of markers in blood tests of patients with pleural effusion and their correlation to different etiologies. Journal of Community Hospital Internal Medicine Perspectives 14, 30.Google Scholar
Khan, J, Lieberman, JA and Lockwood, CM (2017) Variability in, variability out: best practice recommendations to standardize pre-analytical variables in the detection of circulating and tissue microRNAs. Clinical Chemistry and Laboratory Medicine (CCLM) 55, 608621.Google Scholar
Farina, NH, Wood, ME, Perrapato, SD, et al. (2014) Standardizing analysis of circulating microRNA: Clinical and biological relevance. Journal of cellular biochemistry 115, 805811.Google Scholar
MacLellan, SA, MacAulay, C, Lam, S, et al. (2014) Pre-profiling factors influencing serum microRNA levels. BMC Clinical Pathology 14, 27.Google Scholar
Wright, K, de Silva, K, Purdie, AC, et al. (2020) Comparison of methods for miRNA isolation and quantification from ovine plasma. Scientific Reports 10, 825.Google Scholar
Holohan, C, Feely, N, Li, P, et al. (2022) Role of detergents and nuclease inhibitors in the extraction of RNA from eukaryotic cells in complex matrices. Nanoscale 14, 1215312161.Google Scholar
Tan, SC and Yiap, BC (2009) DNA, RNA, and protein extraction: The past and the present. Journal of Biomedicine and Biotechnology 2009, 574398.Google Scholar
McAlexander, MA, Phillips, MJ and Witwer, KW (2013) Comparison of methods for miRNA extraction from plasma and quantitative recovery of RNA from cerebrospinal fluid. Frontiers in Genetics 4.Google Scholar
Ban, E, Chae, D-K, Yoo, YS, et al. (2017) An improvement of miRNA extraction efficiency in human plasma. Analytical and Bioanalytical Chemistry 409, 63976404.Google Scholar
Espy, MJ, Uhl, JR, Sloan, LM, et al. (2006) Real-time PCR in clinical microbiology: Applications for routine laboratory testing. Clinical Microbiology Reviews 19, 165256.Google Scholar
Diefenbach, RJ, Lee, JH, Kefford, RF, et al. (2018) Evaluation of commercial kits for purification of circulating free DNA. Cancer Genetics 228–229, 2127.Google Scholar
Sriram, H, Khanka, T, Kedia, S, et al. (2021) Improved protocol for plasma microRNA extraction and comparison of commercial kits. Biochemia Medica 31, 030705.Google Scholar
Timmerman, DM, Gillis, AJM, Mego, M, et al. (2021) Comparative analyses of liquid-biopsy microRNA371a-3p isolation protocols for serum and plasma. Cancers 13, 4260.Google Scholar
Oberacker, P, Stepper, P, Bond, DM, et al. (2018) Bio-On-Magnetic-Beads (BOMB): Open platform for high-throughput nucleic acid extraction and manipulation. Bio-On-Magnetic-Beads (BOMB): Open platform for high-throughput nucleic acid extraction and manipulation 2018, 414516.Google Scholar
Scott, S, Shaheen, C, McGuinness, B, et al. (2019) Single-molecule visualization of the effects of ionic strength and crowding on structure-mediated interactions in supercoiled DNA molecules. Nucleic Acids Research 47, 63606368.Google Scholar
Harnett, J, Weir, S and Michieletto, D (2024) Effects of monovalent and divalent cations on the rheology of entangled DNA. Soft Matter 20, 39803986.Google Scholar
Stortchevoi, A, Kamelamela, N and Levine, SS (2020) SPRI beads-based size selection in the range of 2-10kb. Journal of Biomolecular Techniques : JBT 31, 710.Google Scholar
Fishman, A and T.Lamm, A (2019) QsRNA-seq: A protocol for generating libraries for high-throughput sequencing of small RNAs. Bio-protocol 9, e3179.Google Scholar
Petrov, A, Wu, T, Puglisi, EV, et al. (2013) RNA purification by preparative polyacrylamide gel electrophoresis. Methods in Enzymology 530, 315330.Google Scholar
Xie, J, Chen, J, Zhang, Y, et al. (2024) A dual-signal amplification strategy based on rolling circle amplification and APE1-assisted amplification for highly sensitive and specific miRNA analysis for early diagnosis of alzheimer’s disease. Talanta 272, 125747.Google Scholar
Wang, J, Lu, Z, Tang, H, et al. (2017) Multiplexed electrochemical detection of MiRNAs from sera of glioma patients at different stages via the novel conjugates of conducting magnetic microbeads and diblock oligonucleotide-modified gold nanoparticles. Analytical Chemistry 89, 1083410840.Google Scholar
Joshi, GK, Deitz-McElyea, S, Liyanage, T, et al. (2015) Label-free nanoplasmonic-based short noncoding RNA sensing at attomolar concentrations allows for quantitative and highly specific assay of MicroRNA-10b in biological fluids and circulating exosomes. ACS Nano 9, 1107511089.Google Scholar
Dai, C, Wang, K, Tan, M, et al. (2023) A LoC-SERS platform based on triple signal amplification for highly sensitive detection of colorectal cancer miRNAs. Analytical Methods 15, 41944203.Google Scholar
Yekula, A, Muralidharan, K, Kang, K, et al. (2020) From laboratory to clinic: Translation of extracellular vesicle based cancer biomarkers. Methods (San Diego, Calif.) 177, 5866.Google Scholar
Zhang, Q, Jeppesen, DK, Higginbotham, JN, et al. (2023) Comprehensive isolation of extracellular vesicles and nanoparticles. Nature Protocols 18, 14621487.Google Scholar
Park, J, Bae, M, Seong, H, et al. (2023) An innovative charge-based extracellular vesicle isolation method for highly efficient extraction of EV-miRNAs from liquid samples: miRQuick. Journal of Extracellular Biology 2, e126.Google Scholar
Böing, AN, van der Pol, E, Grootemaat, AE, et al. (2014) Single-step isolation of extracellular vesicles by size-exclusion chromatography. Journal of Extracellular Vesicles 3, 23430.Google Scholar
McNamara, RP and Dittmer, DP (2020) Modern techniques for the isolation of extracellular vesicles and viruses. Journal of Neuroimmune Pharmacology 15, 459472.Google Scholar
Chen, C, Tan, R, Wong, L, et al. (2011) Quantitation of microRNAs by real-time RT-qPCR. Methods in Molecular Biology (Clifton, N.J.) 687, 113134.Google Scholar
Chen, C, Ridzon, DA, Broomer, AJ, et al. (2005) Real-time quantification of microRNAs by stem–loop RT–PCR. Nucleic Acids Research 33, e179.Google Scholar
Sellner, LN, Coelen, RJ and Mackenzie, JS (1992) Reverse transcriptase inhibits taq polymerase activity. Nucleic Acids Research 20, 14871490.Google Scholar
González, JM, Kato, C and Horikoshi, K (1995) Thermococcus peptonophilus sp. nov., a fast-growing, extremely thermophilic archaebacterium isolated from deep-sea hydrothermal vents. Archives of Microbiology 164, 159164.Google Scholar
Xue, Y, Wang, K, Jiang, Y, et al. (2024) An ultrasensitive and multiplexed miRNA one-step real time RT-qPCR detection system and its application in esophageal cancer serum. Biosensors and Bioelectronics 247, 115927.Google Scholar
Lin, Q, Mao, W, Shu, Y, et al. (2012) A cluster of specified microRNAs in peripheral blood as biomarkers for metastatic non-small-cell lung cancer by stem-loop RT-PCR. Journal of Cancer Research and Clinical Oncology 138, 8593.Google Scholar
Resnick, KE, Alder, H, Hagan, JP, et al. (2009) The detection of differentially expressed microRNAs from the serum of ovarian cancer patients using a novel real-time PCR platform. Gynecologic Oncology 112, 5559.Google Scholar
Kang, K, Zhang, X, Liu, H, et al. (2012) A novel real-time PCR assay of microRNAs using S-Poly(T), a specific oligo(dT) reverse transcription primer with excellent sensitivity and specificity. PLoS ONE 7, e48536.Google Scholar
Almohsen, F, Al-Rubaie, HA, Habib, MA, et al. (2022) Circulating miR-126-3p and miR-423-5p expression in de novo adult acute myeloid leukemia: Correlations with response to induction therapy and the 2-Year overall survival. Journal of Blood Medicine 13, 8392.Google Scholar
Vautrot, V and Behm-Ansmant, I (2020) Enhanced probe-based RT-qPCR quantification of microRNAs using poly(A) tailing and 5′ adaptor ligation. In Biassoni, R., Raso, A. (eds.), Quantitative Real-Time PCR: Methods and Protocols, Springer, New York, NY, pp. 3954.Google Scholar
Cirillo, PDR, Margiotti, K, Mesoraca, A, et al. (2020) Quantification of circulating microRNAs by droplet digital PCR for cancer detection. BMC Research Notes 13, 351.Google Scholar
Olmedillas-López, S, García-Arranz, M and García-Olmo, D (2017) Current and emerging applications of droplet digital PCR in oncology. Molecular Diagnosis & Therapy 21, 493510.Google Scholar
Quan, P-L, Sauzade, M and Brouzes, E (2018) dPCR: A technology review. Sensors (Basel, Switzerland) 18, 1271.Google Scholar
Ito, T, Kawashima, Y, Fujikawa, T, et al. (2019) Rapid screening of copy number variations in STRC by droplet digital PCR in patients with mild-to-moderate hearing loss. Human Genome Variation 6, 16.Google Scholar
Torii, Y, Suzuki, T, Fukuda, Y, et al. (2024) MicroRNA expression profiling of urine exosomes in children with congenital cytomegalovirus infection. Scientific Reports 14, 5475.Google Scholar
Limothai, U, Chuaypen, N, Poovorawan, K, et al. (2020) Reverse transcriptase droplet digital PCR vs reverse transcriptase quantitative real-time PCR for serum HBV RNA quantification. Journal of Medical Virology 92, 33653372.Google Scholar
Nuno, M, Lafin, JT, Scarpini, C, et al. (2023) Comparison of serum miR-371a-3p assay performance by digital droplet PCR and reverse transcriptase quantitative PCR in patients with malignant germ cell tumor. Journal of Clinical Oncology 41, 426426.Google Scholar
Oliveira, BB, Veigas, B and Baptista, PV (2021) Isothermal amplification of nucleic acids: The race for the next “gold standard.” Frontiers in Sensors 2.Google Scholar
Ali, MM, Li, F, Zhang, Z, et al. (2014) Rolling circle amplification: a versatile tool for chemical biology, materials science and medicine. Chemical Society Reviews 43, 33243341.Google Scholar
Wang, R, Zhao, X, Chen, X, et al. (2020) Rolling circular amplification (RCA)-assisted CRISPR/Cas9 cleavage (RACE) for highly specific detection of multiple extracellular vesicle microRNAs. Analytical Chemistry 92, 21762185.Google Scholar
Xu, H, Wu, X, Liu, Q, et al. (2023) A universal strategy for enhancing the circulating miRNAs’ detection performance of rolling circle amplification by using a dual-terminal stem-loop padlock. ACS Nano 18, 436450.Google Scholar
Notomi, T, Okayama, H, Masubuchi, H, et al. (2000) Loop-mediated isothermal amplification of DNA. Nucleic Acids Research 28, e63.Google Scholar
Talap, J, AL-maskri, AAA, Shen, M, et al. (2023) Ultrasensitive detection of serum miRNA biomarkers related to papillary thyroid cancer using ligation-initiated phosphorothioated primer-based loop-mediated isothermal amplification. Sensors and Actuators B: Chemical 374, 132785.Google Scholar
Zhang, K, Li, Y, Jiang, S, et al. (2024) Catalytic assembly of DNAzyme integrates with primer exchange reaction (CDiPER) for highly sensitive detection of microRNA. ACS Omega 9, 1089710903.Google Scholar
Jinek, M, Chylinski, K, Fonfara, I, et al. (2012) A programmable dual-RNA–guided DNA endonuclease in adaptive bacterial immunity. Science 337, 816821.Google Scholar
Banakar, R, Schubert, M, Collingwood, M, et al. (2020) Comparison of CRISPR-Cas9/Cas12a ribonucleoprotein complexes for genome editing efficiency in the rice phytoene desaturase (OsPDS) gene. Rice 13, 4.Google Scholar
Liu, X, Zhao, X, Yuan, Y, et al. (2022) Accurate detection of lung cancer-related microRNA through CRISPR/Cas9-assisted garland rolling circle amplification. Journal of Thoracic Disease 14, 44274434.Google Scholar
Yang, Y, Yang, J, Gong, F, et al. (2022) Combining CRISPR/Cas12a with isothermal exponential amplification as an ultrasensitive sensing platform for microRNA detection. Sensors and Actuators B: Chemical 367, 132158.Google Scholar
Walker, GT, Fraiser, MS, Schram, JL, et al. (1992) Strand displacement amplification--an isothermal, in vitro DNA amplification technique. Nucleic Acids Research 20, 16911696.Google Scholar
Chi, Z, Wu, Y, Chen, L, et al. (2022) CRISPR-Cas14a-integrated strand displacement amplification for rapid and isothermal detection of cholangiocarcinoma associated circulating microRNAs. Analytica Chimica Acta 1205, 339763.Google Scholar
Luo, B, Zhou, J, Zhan, X, et al. (2024) Visual and colorimetric detection of microRNA in clinical samples based on strand displacement amplification and nanozyme-mediated CRISPR-Cas12a system. Talanta 277, 126310.Google Scholar
Cui, Y, Fan, S, Yuan, Z, et al. (2021) Ultrasensitive electrochemical assay for microRNA-21 based on CRISPR/Cas13a-assisted catalytic hairpin assembly. Talanta 224, 121878.Google Scholar
Zhou, T, Huang, M, Lin, J, et al. (2021) High-Fidelity CRISPR/Cas13a trans-cleavage-triggered rolling circle amplified DNAzyme for visual profiling of microRNA. Analytical Chemistry 93, 20382044.Google Scholar
Shan, Y, Zhou, X, Huang, R, et al. (2019) High-fidelity and rapid quantification of miRNA combining crRNA programmability and CRISPR/Cas13a trans-cleavage activity. Analytical Chemistry 91, 52785285.Google Scholar
Liu, H, Dong, J, Duan, Z, et al. (2024) Light-activated CRISPR-Cas12a for amplified imaging of microRNA in cell cycle phases at single-cell levels. Science Advances 10, eadp6166.Google Scholar
Wei, J, He, S, Mao, Y, et al. (2021) A simple “signal off-on” fluorescence nanoplatform for the label-free quantification of exosome-derived microRNA-21 in lung cancer plasma. Mikrochimica Acta 188, 397.Google Scholar
Choi, HMT, Schwarzkopf, M, Fornace, ME, et al. (2018) Third-generation in situ hybridization chain reaction: multiplexed, quantitative, sensitive, versatile, robust. Development (Cambridge, England) 145, dev165753.Google Scholar
Xu, G, Lai, M, Wilson, R, et al. (2019) Branched hybridization chain reaction—using highly dimensional DNA nanostructures for label-free, reagent-less, multiplexed molecular diagnostics. Microsystems & Nanoengineering 5, 17.Google Scholar
Liang, D, Zhang, X, Wang, Y, et al. (2022) Magnetic covalent organic framework nanospheres-based miRNA biosensor for sensitive glioma detection. Bioactive Materials 14, 145151.Google Scholar
Chen, Q, Chen, H, Kong, H, et al. (2024) Enzyme-free sensitive SERS biosensor for the detection of thalassemia-associated microRNA-210 using a cascade dual-signal amplification strategy. Analytica Chimica Acta 1292, 342255.Google Scholar
Mahata, P, Roy, P, Mellot-Draznieks, C, et al. (2007) Inter-relationship between the structures of metal-organic frameworks: is there a role for an intermediate? In Xu, R., Gao, Z., Chen, J., et al. (eds.), Studies in Surface Science and Catalysis, From Zeolites to Porous MOF Materials - The 40th Anniversary of International Zeolite Conference, Elsevier, Vol. 170, pp. 732738.Google Scholar
Cui, Y, Zhang, J, He, H, et al. (2018) Photonic functional metal–organic frameworks. Chemical Society Reviews 47, 57405785.Google Scholar
Sun, Z, Li, J, Tong, Y, et al. (2022) A ratiometric fluorescent biosensor based on self-fluorescent MOF and target-triggered rolling circle amplification for sensitive detection of exosome-derived miRNA. Analytica Chimica Acta 1221, 340136.Google Scholar
Li, X, Li, X, Li, D, et al. (2020) Electrochemical biosensor for ultrasensitive exosomal miRNA analysis by cascade primer exchange reaction and MOF@Pt@MOF nanozyme. Biosensors and Bioelectronics 168, 112554.Google Scholar
Gao, X, Wu, H, Li, Y, et al. (2022) Dancing in local space: rolling hoop orbital amplification combined with local cascade nanozyme catalytic system to achieve ultra-sensitive detection of exosomal miRNA. Journal of Nanobiotechnology 20, 357.Google Scholar
Wang, Y, Yang, M, Shi, H, et al. (2022) Photoelectrochemical detection of exosomal miRNAs by combining target-programmed controllable signal quenching engineering. Analytical Chemistry 94, 30823090.Google Scholar
Xiao, P-P, Wan, Q-Q, Liao, T, et al. (2021) Peptide nucleic acid-functionalized nanochannel biosensor for the highly sensitive detection of tumor exosomal microRNA. Analytical Chemistry 93, 1096610973.Google Scholar
Zeng, D, Wang, Z, Meng, Z, et al. (2017) DNA tetrahedral nanostructure-based electrochemical miRNA biosensor for simultaneous detection of multiple miRNAs in pancreatic carcinoma. ACS Applied Materials & Interfaces 9, 2411824125.Google Scholar
Li, T, Liang, Y, Li, J, et al. (2021) Carbon nanotube field-effect transistor biosensor for ultrasensitive and label-free detection of breast cancer exosomal miRNA21. Analytical Chemistry 93, 1550115507.Google Scholar
Li, K, Tu, J, Zhang, Y, et al. (2022) Ultrasensitive detection of exosomal miRNA with PMO-graphene quantum dots-functionalized field-effect transistor biosensor. iScience 25, 104522.Google Scholar
Afzalinia, A and Mirzaee, M (2020) Ultrasensitive fluorescent miRNA biosensor based on a “sandwich” oligonucleotide hybridization and fluorescence resonance energy transfer process using an Ln(III)-MOF and Ag nanoparticles for early cancer diagnosis: Application of central composite design. ACS Applied Materials & Interfaces 12, 1607616087.Google Scholar
Jiang, S, Li, Q, Wang, C, et al. (2021) In situ exosomal microRNA determination by target-triggered SERS and Fe3O4@TiO2-based exosome accumulation. ACS Sensors 6, 852862.Google Scholar
Zeng, X, Wang, L, Liu, C, et al. (2024) An integrated liposome-based microfluidic strategy for rapid colorimetric analysis: A case study of microRNA-21 detection. Talanta 272, 125838.Google Scholar
Novogene, (2024) Next-Generation Sequencing versus qPCR and Microarrays | RNA-Seq Blog. Next-Generation Sequencing versus qPCR and Microarrays | RNA-Seq Blog. https://www.rna-seqblog.com/next-generation-sequencing-versus-qpcr-and-microarrays/ Accessed June 25, 2024.Google Scholar
Zhang, A and Hu, H (2022) A novel blood-based microRNA diagnostic model with high accuracy for multi-cancer early detection. Cancers 14, 1450.Google Scholar
Malkov, VA, Serikawa, KA, Balantac, N, et al. (2009) Multiplexed measurements of gene signatures in different analytes using the nanostring nCounter™ assay system. BMC Research Notes 2, 80.Google Scholar
Niemira, M, Erol, A, Bielska, A, et al. (2023) Identification of serum miR-1246 and miR-150-5p as novel diagnostic biomarkers for high-grade serous ovarian cancer. Scientific Reports 13, 19287.Google Scholar
Marguerat, S and Bähler, J (2010) RNA-seq: from technology to biology. Cellular and Molecular Life Sciences 67, 569579.Google Scholar
Khodakov, D, Wang, C and Zhang, DY (2016) Diagnostics based on nucleic acid sequence variant profiling: PCR, hybridization, and NGS approaches. Advanced Drug Delivery Reviews 105, 319.Google Scholar
Geng, X, Tsou, J-H, Stass, SA, et al. (2023) Utilizing MiSeq sequencing to detect circulating microRNAs in plasma for improved lung cancer diagnosis. International Journal of Molecular Sciences 24, 10277.Google Scholar
Li, SFY (1992) Capillary electrophoresis: Principles, practice and applications. Capillary Electrophoresis: Principles, Practice and Applications ; Elsevier, 1992.Google Scholar
St. Claire, RL (1996) Capillary electrophoresis. Analytical Chemistry 68, 569586.Google Scholar
Benedek, K and Thiede, S (1994) High-performance capillary electrophoresis of proteins using sodium dodecyl sulfate-poly(ethylene oxide). Journal of Chromatography A 676, 209217.Google Scholar
Guttman, A, Shieh, P, Lindahl, J, et al. (1994) Capillary sodium dodecyl sulfate gel electrophoresis of proteins II. On the ferguson method in polyethylene oxide gels. Journal of Chromatography A 676, 227231.Google Scholar
Guttman, A, Shieh, P, Hoang, D, et al. (1994) Effect of operational variables on the separation of proteins by capillary sodium dodecyl sulfate-gel electrophoresis. ELECTROPHORESIS 15, 221224.Google Scholar
Krawczak, M and Lubjuhn, T (1995) An informativity index for multilocus DNA fingerprints. ELECTROPHORESIS 16, 1621.Google Scholar
Anazawa, T, Takahashi, S and Kambara, H (1996) A capillary array gel electrophoresis system using multiple laser focusing for DNA sequencing. Analytical Chemistry 68, 26992704.Google Scholar
Ban, E, Lim, HJ, Kwon, H, et al. (2023) Practical magnetic bead-based capillary electrophoresis with laser-induced fluorescence for detecting endogenous miRNA in plasma. Heliyon 9.Google Scholar
Larkins, MC and Thombare, A (2024) Point-of-Care Testing. StatPearls, StatPearls Publishing, Treasure Island (FL).Google Scholar
Duffy, MJ, McDermott, EW and Crown, J (2018) Blood-based biomarkers in breast cancer: From proteins to circulating tumor cells to circulating tumor DNA. Tumour Biology: The Journal of the International Society for Oncodevelopmental Biology and Medicine 40, 1010428318776169.Google Scholar
Das, S, Dey, MK, Devireddy, R, et al. (2023) Biomarkers in cancer detection, diagnosis, and prognosis. Sensors (Basel, Switzerland) 24, 37.Google Scholar
Yan, L, Su, Y, Hsia, I, et al. (2023) Delivery of anti-microRNA-21 by lung-targeted liposomes for pulmonary fibrosis treatment. Molecular Therapy Nucleic Acids 32, 3647.Google Scholar
Pagoni, M, Cava, C, Sideris, DC, et al. (2023) miRNA-based technologies in cancer therapy. Journal of Personalized Medicine 13, 1586.Google Scholar
Conduit, C, Grimison, PS, Lewin, JH, et al. (2023) CLIMATE: Assessing the clinical utility of miR-371a-3p as a marker of residual disease in stage 1 testicular germ cell tumour (TGCT) following orchiectomy: ANZUP 1906. Journal of Clinical Oncology 41, TPS5100–TPS5100.Google Scholar
Morin Coulombe, A, Ozgun, G, Johnson, R, et al. (2023) SWOG S1823/CCTG GCC1: Translational observational investigational study of the liquid biomarker microRNA 371a-3p in newly diagnosed germ cell tumours—Real-world trial design, rapid accrual, and robust secondary use of data opportunities. Journal of Clinical Oncology 41, TPS5103–TPS5103.Google Scholar
S1823 | SWOG. https://www.swog.org/clinical-trials/s1823 Accessed November 30, 2024.Google Scholar
Peltier, HJ and Latham, GJ (2008) Normalization of microRNA expression levels in quantitative RT-PCR assays: Identification of suitable reference RNA targets in normal and cancerous human solid tissues. RNA 14, 844852.Google Scholar
Ye, J, Xu, M, Tian, X, et al. (2019) Research advances in the detection of miRNA. Journal of Pharmaceutical Analysis 9, 217226.Google Scholar
Figure 0

Figure 1. General overview of miRNA detection workflow. The steps are as follows: A. Sample collection from the patient, B. sample treatment, C. small RNA isolation including miRNA, D. sample processing including methods of target amplification, library preparation, and surface hybridisation, and E. detection of signature miRNA.

Figure 1

Figure 2. Comparison of clinically validated miRNA detection workflows. Sample type, treatment, RNA isolation, processing, and detection methods are outlined.

Figure 2

Table 1. Overview of unique miRNA detection workflow for targeted disease diagnostics from liquid biopsies

Figure 3

Figure 3. Blood treatment for plasma and serum separation. Plasma is obtained by centrifugation and treatment with an anticoagulant, such as EDTA, sodium citrate, or heparin. The serum is obtained by extracting the remaining liquid after clotted blood.

Figure 4

Figure 4. Overview of the three common types of RNA isolation for miRNA detection. A. Organic phase separation-based extraction method where vesicles are lysed, and contaminants are separated by centrifugal force. B. Spin column-based extraction method with a similar lysis procedure but where RNA is bound and eluted through a silica filter and C. Magnetic bead-based extraction method where RNA is bound to silica-based magnetic nanoparticles through ionic bridges supported by cations. The RNA is eluted sequentially in a water-based solution. (Washing steps not shown).

Figure 5

Figure 5. Schematic illustration of thermal amplification methods for miRNA detection. A. Stem-Loop RT-qPCR (Refs 41, 46, 79, 83, 84) with (i) SYBR Green detection - universal and (ii) Probe-based detection - specific; B. Poly-A Tail RT-qPCR; (Refs 45, 86) C. Sensitive and Multiplexed One-Step RT-qPCR (SMOS-qPCR); (Ref. 82) and D. Droplet digital PCR (ddPCR). (Refs 88,92)

Figure 6

Figure 6. Schematic illustration of isothermal amplification methods for detecting miRNA. A. Rolling circle amplification (RCA) + APE1 enzyme; (Ref. 69) B. Padlock-assisted hyperbranched rolling circle amplification (HRCA); (Ref. 98) C. Primer-based loop-mediated isothermal amplification (PS-LAMP); (Ref. 100) D. Catalytic assembly of DNAzyme integrated with primer exchange reaction (CDiPER); (Ref. 101) and E. Multiplexed exponential amplification reaction (MEXPAR). (Ref. 42)

Figure 7

Figure 7. Schematic illustration of isothermal amplification methods with CRISPR/Cas systems for detecting miRNA. A. Rolling circle amplification (RCA)-CRISPR/Cas9; (Ref. 104) B. exponential amplification reaction (EXPAR)-CRISPR/Cas12a; (Ref. 105) C. Strand displacement amplification (SDA)-CRISPR/Cas14a; (Ref. 107) and D. Target-triggered SDA (ttSDA)-CRISPR/Cas12a + Pt@Au nanozymes. (Ref. 108)

Figure 8

Figure 8. Schematic illustration of isothermal amplification methods with nanomaterials for detecting miRNA. A. Hybridisation chain reaction (HCR) + Molybdenum disulfide (MoS2) nanosheets; (Ref. 113) B. HCR + magnetic covalent organic framework (COF) nanospheres; (Ref. 116) C. Catalytic hairpin assembly reaction (CHA) + HCR + Au-Ag nanoparticles; (Ref. 117) and D. CHA + enzyme-assisted signal amplification (EASA) + Au–Ag nanoshuttles. (Ref. 72)

Figure 9

Figure 9. Schematic illustration of isothermal amplification methods with metal–organic frameworks (MOFs) for detecting miRNA. A. Rolling circle amplification (RCA) + MOF-525; (Ref. 120) B. Primer exchange reaction (PER) + MOF@Pt@MOF nanozyme; (Ref. 121) C. Rolling hoop orbital amplification (RHOA) + Iron-Zirconium (Fe-Zr) MOF/G-quadruplex (G4) nanozyme; (Ref. 122) and D. λ-Exonuclease + ZIF-67 MOF/ Tungsten(VI) oxide (WO3) nanoflakes. (Ref. 123)

Figure 10

Figure 10. Schematic illustration of electrochemical and electrical nanoparticle miRNA detection methods. A. Magnetic microbeads & diblock oligonucleotide-modified Au nanoparticles; (Ref. 70) B. Peptide nucleic acid (PNA) functionalised nanochannel biosensor; (Ref. 124) C. DNA tetrahedral nanostructure-based biosensor; (Ref. 125) D. Carbon nanotube biosensor; (Ref. 126) and E. Phosphorodiamidate morpholino oligomers (PMO) - graphene quantum dots (GQDs) - functionalised reduced graphene oxide (RGO) field effect transistor (FET) biosensor. (Ref. 127)

Figure 11

Figure 11. Schematic illustration of fluorescent, plasmonic, and colourimetric nanoparticle miRNA detection methods. A. Lanthanum oxide (La(III))-metal–organic framework (MOF) & silver (Ag) nanoparticle biosensor; (Ref. 128) B. Iron(II,III) oxide@Titanium oxide (Fe3O4@TiO2) nanoparticle biosensor; (Ref. 129) C. Gold (Au) nanoprism biosensor;(Ref. 71) and D. Gold (Au) nanoparticle liposome-based microfluidic chip biosensor. (Ref. 130)

Figure 12

Figure 12. Schematic illustration of various miRNA detection methods. A. Microarray, (Ref. 132) B. nCounter, (Refs 43, 134) C. RNA sequencing, (Ref. 137) and D. Capillary electrophoresis. (Ref. 145)