We use cookies to distinguish you from other users and to provide you with a better experience on our websites. Close this message to accept cookies or find out how to manage your cookie settings.
To save content items to your account,
please confirm that you agree to abide by our usage policies.
If this is the first time you use this feature, you will be asked to authorise Cambridge Core to connect with your account.
Find out more about saving content to .
To save content items to your Kindle, first ensure no-reply@cambridge.org
is added to your Approved Personal Document E-mail List under your Personal Document Settings
on the Manage Your Content and Devices page of your Amazon account. Then enter the ‘name’ part
of your Kindle email address below.
Find out more about saving to your Kindle.
Note you can select to save to either the @free.kindle.com or @kindle.com variations.
‘@free.kindle.com’ emails are free but can only be saved to your device when it is connected to wi-fi.
‘@kindle.com’ emails can be delivered even when you are not connected to wi-fi, but note that service fees apply.
Monte Carlo (MC) simulations have been used extensively for benchmarking photon dose calculations in modern radiotherapy using linear accelerators (linacs). Moreover, a major barrier to widespread clinical implementation of MC dose calculation is the difficulty in characterising the radiation source using data reported from manufacturers.
Purpose
This work aims to develop a generalised full MC histogram source model of an Elekta Precise SL-25 linac (electron exit window, target, flattening filter, monitor chambers and collimators) for 6 MV photon beams used in standard therapies. The inclusion of many different probability processes such as scatter, nuclear reactions, decay, capture cross-sections and more led to more realistic dose calculations in treatment planning and quality assurance.
Materials and methods
Two different codes, MCNPX 2·6 and EGSr-BEAM, were used for the calculation of particle transport, first in the geometry of the internal/external accelerator source, and then followed by tracking the transport and energy deposition in phantom-equivalent tissues. A full phase space file was scored directly above the upper multilayer collimator’s jaws to derive the beam characteristics such as planar fluence, angular distribution and energy spectrum. To check the quality of the generated photon beam, its depth dose curves and cross-beam profiles were calculated and compared with measured data.
Results
In-field dose distributions calculated using the accelerator models were tuned to match measurement data with preliminary calculations performed using the accelerator information provided by the manufacturer. Field sizes of 3×3, 5×5, 10×10, 15×15 and 20×20 cm2 were analysed. Local differences between calculated and measured curve doses beneath 2% were obtained for all the studied field sizes. Higher discrepancies were obtained in the air–water interface, where measurements of dose distributions with the ionisation chamber need to be shifted for the effective point of measurement.
Conclusion
The agreements between MC-calculated and measured dose distributions were excellent for both codes, showing the strength and stability of the proposed model. Beam reconstruction methods as direct input to dose-calculation codes using the recorded histograms can be implemented for more accurate patient dose estimation.