To minimize waste, improve process safety, and reduce costs, modifications were implemented to a method for quantifying gallium in plutonium metal using wavelength dispersive X-ray fluorescence. These changes included reducing sample sizes, reducing ion exchange process volumes, using cheaper reagent grade acids, eliminating the use of HF acid, and using more robust containment films for sample analysis. Relative precision and accuracy achieved from analyzing multiple aliquots from a single parent sample were approximately 0.2 and 0.1%, respectively. The same precision was obtained from analyzing a total of four parent materials, and the average relative accuracy from all the samples was 0.4%, which is within programmatic uncertainty requirements.