Hostname: page-component-cd9895bd7-fscjk Total loading time: 0 Render date: 2024-12-26T05:45:50.008Z Has data issue: false hasContentIssue false

An epicardial pacing safety net: an alternative technique for pacing in the young

Published online by Cambridge University Press:  10 March 2009

Scott R. Ceresnak*
Affiliation:
Department of Pediatrics, Division of Pediatric Cardiology, Morgan Stanley Children’s Hospital, The Children’s Hospital of New York, New York Presbyterian Hospital – Columbia University, New York, New York, United States of America
Leonardo Liberman
Affiliation:
Department of Pediatrics, Division of Pediatric Cardiology, Morgan Stanley Children’s Hospital, The Children’s Hospital of New York, New York Presbyterian Hospital – Columbia University, New York, New York, United States of America
Jonathan M. Chen
Affiliation:
Department of Cardiothoracic Surgery, Morgan Stanley Children’s Hospital, The Children’s Hospital of New York, New York Presbyterian Hospital – Weill Cornell Medical Center, New York, United States of America
Allan J. Hordof
Affiliation:
Department of Pediatrics, Division of Pediatric Cardiology, Morgan Stanley Children’s Hospital, The Children’s Hospital of New York, New York Presbyterian Hospital – Columbia University, New York, New York, United States of America
John J. Lamberti
Affiliation:
Department of Cardiothoracic Surgery, Rady Children’s Hospital, The Children’s Hospital of San Diego, San Diego, California, United States of America
William J. Bonney
Affiliation:
Department of Pediatrics, Division of Pediatric Cardiology, Morgan Stanley Children’s Hospital, The Children’s Hospital of New York, New York Presbyterian Hospital – Columbia University, New York, New York, United States of America
Robert H. Pass
Affiliation:
Department of Pediatrics, Division of Pediatric Cardiology, Children’s Hospital at Montefiore, University Hospital of Albert Einstein College of Medicine, Bronx, New York, United States of America
*
Correspondence to: Scott R. Ceresnak, MD, Department of Pediatrics, Division of Pediatric Cardiology, Lucille Packard Children’s Hospital, Stanford University, 750 Welch Road, Suite #305, Palo Alto, CA 94304USA. Tel: (732) 233 – 8968; Fax: (866) 4593443; E-mail: ceresnak@yahoo.com

Abstract

Epicardial pacing is the standard approach for permanent pacing in small children and patients with functionally univentricular physiology. The longevity of epicardial leads, however, is compromised by increased occurrences of exit block and lead fractures. We report our experience with a technique of placing a second ventricular lead, and attaching it to the atrial port of a dual chamber pacemaker to prevent the need for early re-operation in the event of failure of the primary epicardial lead. A retrospective review showed that, over the period from 2001 through 2007, epicardial ventricular pacemakers had been placed in 88 patients. In 6 of these, we had placed 2 ventricular leads, their median weight being 8.0 kilograms, with a range from 4.2 to 31.8 kilograms. Fracture of a lead occurred in 1 of the patients (17%) 8 months after placement, requiring reprogramming to pace from the atrial port. This possibility avoided the need for repeated emergent surgery. At a median follow-up of 1.5 years, with a range from 0.3 to 4.4 years, there have been no complications. During the same time period, overall failure of epicardial leads at our institution was 13%. Placement of a second ventricular epicardial pacing lead, attached to the atrial port of a dual chamber pacemaker, therefore, may provide a safe and effective means of ventricular pacing in the setting of epicardial lead failure, and may obviate the need for repeat, potentially urgent, pacemaker surgery.

Type
Original Article
Copyright
Copyright © Cambridge University Press 2009

Access options

Get access to the full version of this content by using one of the access options below. (Log in options will check for institutional or personal access. Content may require purchase if you do not have access.)

References

1. Walsh, CA, McAlister, HF, Andrews, CA, Steeg, CN, Eisenberg, R, Furman, S. Pacemaker implantation in children: a 21-year experience. Pacing Clin Electophysiol 1988; 11: 19401944.CrossRefGoogle ScholarPubMed
2. Warner, KG, Halpin, DP, Berul, CI, Payne, DD. Placement of permanent epicardial pacemaker in children using a subcostal approach. Ann Thorac Surg 1999; 68: 173175.CrossRefGoogle ScholarPubMed
3. Balmer, C, Fasnacht, , Rahn, M, Molinari, L, Bauersfeld, U. Long-term follow-up of children with congenital complete atrioventricular block and the impact of pacemaker therapy. Europace 2002; 4: 345349.CrossRefGoogle ScholarPubMed
4. Noiseux, N, Khairy, P, Fournier, A, Vobecky, SJ. Thirty years of experience with epicardial pacing in children. Cardiol Young 2004; 14: 512519.CrossRefGoogle ScholarPubMed
5. Cohen, MI, Vetter, VL, Wernovsky, G, et al. Epicardial pacemaker implantation and follow-up in patients with a single ventricle after the Fontan operation. J Thorac Cardiovasc Surg 2001; 121: 804811.CrossRefGoogle ScholarPubMed
6. Cohen, MI, Bush, DM, Vetter, VL, et al. Permanent epicardial pacing in pediatric patients: seventeen years of experience and 1200 outpatient visits. Circulation 2001; 103: 25852590.CrossRefGoogle ScholarPubMed
7. Kratz, JM, Gillette, PC, Crawford, FA, Sade, RM, Zeigler, VL. Atrioventricular pacing in congenital heart disease. Ann Thorac Surg 1992; 54: 485489.CrossRefGoogle ScholarPubMed
8. Molina, JE, Dunnigan, AC, Crosson, JE. Implantation of transvenous pacemakers in infants and small children. Ann Thorac Surg 1995; 59: 689694.CrossRefGoogle ScholarPubMed
9. Kiviniemi, MS, Pirnes, MA, Eranen, HJ, Kettunen, RV, Hartikainen, JE. Complications related to permanent pacemaker therapy. Pacing Clin Electrophysiol 1999; 22: 711720.CrossRefGoogle ScholarPubMed
10. Fortescue, EB, Berul, CI, Cecchin, F, Walsh, EP, Triedman, JK, Alexander, ME. Comparison of modern steroid-eluting epicardial and thin transvenous pacemaker leads in pediatric and congenital heart disease patients. J Interv Card Electrophyisol 2005; 14: 2736.CrossRefGoogle ScholarPubMed
11. Hamilton, R, Gow, R, Bahaoric, B, Griffiths, J, Freedom, R, Williams, W. Steroid-eluting epicardial leads in pediatrics: improved epicardial thresholds in the first year. Pacing Clin Electrophysiol 1991; 14: 20662072.CrossRefGoogle Scholar
12. Karpawich, PP, Hakimi, M, Arciniegas, E, Cavitt, DL. Improved chronic epicardial pacing in children: steroid contribuation to porous platinized electrodes. Pacing Clin Electrophysiol 1992; 15: 11511157.CrossRefGoogle Scholar
13. Sachweh, JS, Vazquez-Jimenz, JF, Schondube, FA, et al. Twenty years experience with pediatric pacing: epicardial and transvenous stimulation. Eur J Cardiothorac Surg 2000; 17: 455461.CrossRefGoogle ScholarPubMed
14. Thomson, JD, Blackburn, ME, Van Doorn, C, Nicholls, A, Watterson, KG. Pacing activity, patient and lead survival over 20 years of permanent epicardial pacing in children. Ann Throrac Surg 2004; 77: 13661370.CrossRefGoogle ScholarPubMed
15. Fortescue, EB, Berul, CI, Cecchin, F, Walsh, EP, Triedman, JK, Alexander, ME. Patient, procedural, and hardware factors associated with pacemaker lead failures in pediatrics and congenital heart disease. Heart Rhythm 2004; 1: 150159.CrossRefGoogle Scholar
16. Silvetti, MS, De Santis, A, Grutter, G, Di Ciommo, V, Drago, F. Ventricular capture management in pediatric pacing: efficacy and safety. Ital Heart J 2005; 6: 751756.Google ScholarPubMed
17. Silvetti, MS, Drago, F, Grutter, G, De Santis, A, Di Ciommo, V, Rava, L. Twenty years of paediatric cardiac pacing: 515 pacemakers and 480 leads implanted in 292 patients. Europace 2006; 8: 530536.CrossRefGoogle ScholarPubMed
18. Aellig, NC, Balmer, C, Dodge-Khatami, A, Rahn, M, Pretre, R, Bauersfeld, U. Long-term follow-up after pacemaker implantation in neonates and infants. Ann Thorac Surg 2007; 83: 14201423.CrossRefGoogle ScholarPubMed
19. Johns, JA, Fish, FA, Burger, JD, Hammon, JW Jr. Steroid-eluting epicardial pacing leads in pediatric patients: encouraging early results. J Am Coll Cardiol 1992; 20: 395401.CrossRefGoogle ScholarPubMed
20. Cutler, NG, Karpawich, PP, Cavitt, D, Hakimi, M, Walters, HL. Steroid-eluting epicardial pacing electrodes: six year experience of pacing thresholds in a growing pediatric population. Pacing Clin Electrophysiol 1997; 20: 29432948.CrossRefGoogle Scholar
21. Horenstein, MS, Hakimi, M, Walters, H 3rd, Karpawich, PP. Chronic performance of steroid-eluting epicardial leads in a growing pediatric population: a 10-year comparison. Pacing Clin Electrophysiol 2003; 26: 14671471.CrossRefGoogle Scholar
22. Silka, MJ, Bar-Cohen, Y. Pacemakers and implantable cardioverter-difibrillators in pediatric patients. Heart Rhythm 2006; 11: 13601366.CrossRefGoogle Scholar
23. Beaufort-Krol, GM, Mulder, H, Nagelkerke, D, Waterbolk, TW, Bink-Boelkens, MT. Comparison of longevity, pacing, and sensing characteristics of steroid-eluting epicardial versus conventional endocardial pacing leads in children. J Thorac Cardiovasc Surg 1999; 117: 523528.CrossRefGoogle ScholarPubMed