Extended applications of the Amplatzer vascular plug IV in infants
Published online by Cambridge University Press: 08 December 2010
Abstract
A variety of devices are available for transcatheter closure of unwanted shunts. We describe our experience with the use of the Amplatzer vascular plug IV in a consecutive series of infants.
A total of eight consecutive infants – all born preterm at gestational ages ranging from 24 to 35 weeks – undergoing transcatheter closure of unwanted shunts – persistently patent arterial duct in five patients, an aorta to right atrium fistula in one, multiple aortopulmonary collateral vessels in one, and an azygos vein to left atrium connection in one – are described. Their age, from birth, ranged between 3 and 11 months, and weight between 2.6 and 11.3 kilograms. All devices were delivered using percutaneous arterial or venous vascular access via a large lumen (0.038 inch) 4-French delivery catheter.
All lesions could be successfully occluded using one or more devices. Device diameters ranged between 4 and 8 millimetres, and exceeded the minimum diameter of the target vessel by 1 to 2 millimetres. Successful occlusion was confirmed either directly at angiography or on follow-up echocardiography. Of the infants who were mechanically ventilated prior to the procedure, three could be successfully weaned following closure of the shunt. There were no procedure-related complications.
The new vascular plug IV is cheap and efficacious in closing a variety of shunts in young infants, and warrants further extended clinical application.